CS2
8Pages

{{requestButtons}}

Catalog excerpts

CS2 - 1

Advanced Force Test System 225 lbf (1kN) & 1100 lbf (5kN)

Open the catalog to page 1
CS2 - 2

Force Measurement Has Never Been Easier T he Chatillon® CS2 digital force testers are revolutionary systems A 2-in-1 Laptop Console with Built-in plastics, packaging, textiles, rubber and electronics to name a few. No Flexibility for Production and Lab Environments personal computer is required; no proprietary software is required; no The CS2 uses a 2-in-1 Laptop Console with an 11 inch touch screen for easy and continuous operation in a knowledge of programming languages is required. This easy, flexible and production environment. The touch screen is also easy to operate, making it the...

Open the catalog to page 2
CS2 - 3

Tests Features & Functionality Limit and Break Testing Multi-Stage & Cycle Testing Setup a test based on a load, displacement, height, or time The multi-stage setup allows users to configure virtually any limit in both tensile and compression directions. Similarly, a type of test setup imaginable. Each part of a test is defined break or rupture can be defined as a sharp break or percent- as a “stage”. Each stage can be set up with its own indepen- age drop from the peak load measured. dent load or displacement characteristic and speed. The touch screen interface with stage specific icons...

Open the catalog to page 3
CS2 - 4

Tests Features & Functionality Key Features Test Results Test results are displayed in various views, either as a graph (single result)  Force measurements made easy Accurate results, easy programming, maximum throughput and minimal operator training.  One-click operation Simple and intuitive; with one-click access to favorite tests as well as test results. or in tabular format showing multiple test results. Measured results are available for all test types. Simply select the results (up to 10) you want displayed, Switch between load and distance view in graph mode. Add notes to tests....

Open the catalog to page 4
CS2 - 5

ChatillonW force measurement #1% I Tests Features & FunctionalityEffortless Communication Flexible Mounting The T-slot table contains multiple through holes for mounting various types of threaded fixtures and eye-end adapters. The fixture mounting block can be easily positioned to ensure proper alignment with the load measuring sensor. You can choose from a wide range of standard grips and fixtures as well as custom-designed versions for use with the CS2. Some examples are: Wedge action grips that self-tighten • Pneumatic action grips for faster sample clamping • Vice action grips for film...

Open the catalog to page 5
CS2 - 6

Test Setups ► Load limit ► Displacement limit ► Height limit ► Time limit ► Break or rupture ► Cycle and loop ► Multi-stage (define test as a combination of the above) ► Tensile ► Compression ► Peel Load Measurement System ► CLC Series load sensors ► Auto-recognition, "plug & test" ► Sensor overload history log ► On-site field calibration using traceable master load cells in combination with deadweights. Load measuring system meets or exceeds the following: ASTM E4, BS 1610, DIN 51221, ISO 7500-1, EN 10002-2, AFNOR A03-501. (AMETEK recommends system verification on-site at the time of...

Open the catalog to page 6
CS2 - 7

ChatillonW force measurement #1% ■ Specifications Throat Depth Crosshead Speed Range (at full load) Maximum Return Speed Speed Accuracy (unloaded) ± 1% of selected speed 2% to 100% of maximum speed. Force Capacity Maximum Crosshead Travel (between the eye ends) Technical SpecificationsDimensionsCrosshead Displacement Accuracy (unloaded) ± 0.25% of reading 2% to 100% of maximum travel. Displayed Displacement Resolution 0.001 in / 0.01 mm Load Cell Force Accuracy (non-dedicated) ± 0.25% full scale of load cell capacity Load Cell Force Accuracy (dedicated) ± 0.10 % full scale of load cell...

Open the catalog to page 7
CS2 - 8

Order Number—CS Machines Order Number—Load Cells and Resolutions All values are maximum force rating and smallest resolution. As standard your CLC load cell is calibrated to 0.25% of FS from the factory. A dedicated load cell can be supplied that is calibrated to 0.1% of FS. Please indicate this option with a DED notation after the ordering number, i.e. CLC-0010-DED. For detailed specifications go to the Data Sheet at ametektest.com ChatillonWforce measurement # I m No part of this document may be reproduced or modified in any form or by any means, electronic or mechanical, without express...