Accuracy Specifications:
- Standard accuracy ranges from ±0.02% to ±0.1% depending on the range percentage and temperature. Special accuracy options (S5 and S2) offer improved precision.

Operating Temperature:
- The device operates effectively between -10 to 50°C (14 to 122°F).

Display Features:
- The gauge features a 5.5-digit display with options for single or dual-line displays. It supports a display rate of up to 8 readings per second in PSV mode.

Power and Safety:
- The XP2i is intrinsically safe when powered by specific battery types, with certifications for hazardous locations.

Data and Communication:
- Includes RS232 digital interface and datalogging capabilities with a capacity of 32,000 data points.

Certifications:
- Complies with various international safety and compatibility standards, including ATEX and IECEx.

Enclosure and Sensor:
- The device is housed in diecast aluminum with IP66 and IP67 ratings, and features stainless steel wetted materials.

Special Features:
- Configurable via software for user-defined units, PSV mode, averaging, and more.

Standard Delivery and Accessories:
- Includes the gauge, manual, calibration certificate, batteries, and optional accessories like protective boots and carrying cases.

Pump Systems Overview:
- Various pump systems are available, offering different pressure ranges and case options for pneumatic and hydraulic applications.
See more

Catalog excerpts

XP2i Digital Pressure Gauge kPa Data Sheet-1

PANEL MOUNT FLANGE MAX PANEL THICKNESS 11.7 (0.46) Standard Accuracy Version (-10 to 50° C) 0 to 20% of Range: ± (0.02% of Full Scale*) 20% to 110% of Range: ± (0.1% of Reading) Vacuum: ± (0.25% of Full Scale**) -S5 Accuracy Option (-10 to 50° C) Vacuum to 110% of Range: ± (0.05% of Full Scale*) -S2 Accuracy Option Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Not recommended for continuous use below -100.0 kPa. Refer to XP2i-DP data sheet for gauges that are intended for continuous high vacuum use. **Full Scale = -100.0 kPa SENSOR DIAPHRAGM SURFACE OPERATING TEMPERATURE DISPLAY Single Line Display: 16.9 mm (0.67") high Dual Line Display: Main Display: 13.9 mm (0.55") Secondary Display: 5.3 mm (0.21") Display Rate: 4 readings/second (standard) 8 readings/second (PSV mode) *Full Scale is 100% of the range Vacuum to 110% of Range: ± (0.02% of Full Scale*) 15 to 25° C ± (0.05% of Full Scale*) -10 to 15° C, 25 to 50° C, typical PORT POSITION FOR -F4 AND -RP OPTIONS All models indicate vacuum, but vacuum specification applies to 100 kPa, 200 kPa, 700 kPa, and 2000 kPa models only. Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification. LCD readable in sunlight with bright backlight. PSV mode is intended for Pressure Safety Valve testing, and requires software to enable. (ConfigXP is a free download from ametekcalibration.com. ) 1/4 INCH MBSP ADAPTER ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS

 Open the catalog to page 1
XP2i Digital Pressure Gauge kPa Data Sheet-2

| POWER The XP2i is Intrinsically Safe only if powered by one of the following battery types. Approved Battery Type IECITe&x Ex ia IIC T4/T3 Ga IECEx FTZU 12.0009X Approved Battery Type Exia Intrinsically Safe and Non-incendive for Hazardous Locations: Class I, Division 1, Groups A, B, C and D, Temperature Code T4 or T3C. For hazardous location product warnings, refer to the operation manual. XP2i complies with the Electromagnetic Compatibility and the Pressure Equipment Directives. Ultra Low Power: > 1 year, typical** Uses 3 alkaline AA (LR6) batteries. Use of backlight and/or RS232 reduces...

 Open the catalog to page 2
XP2i Digital Pressure Gauge kPa Data Sheet-3

XP2i Pressure Gauge kPa |ENCLOSURE Weight includes batteries. Nickel plating over low copper, marine grade aluminum. LCD protected from impact damage by 1.5 mm (0.06") thick polycarbonate lens under polyester window. Skydrol® compatible. For hazardous location product warnings, refer to the operation manual. Wetted Materials: (wrench tight) 316 stainless steel (finger tight) 316 stainless steel and Viton® (internal o-ring) Diaphragm Seal Fluid: Silicone Oil Connection: Crystal CPF* Female Temperature Range: -40 to 75° C (-40 to 167° F) Batteries should be removed if stored for more than one month....

 Open the catalog to page 3
XP2i Digital Pressure Gauge kPa Data Sheet-4

XP2i Pressure Gauge kPa| RANGE & RESOLUTION TABLE Display Resolution SAMPLE PART NUMBERS and a waterproof carrying case. ► Ordering a Pump System Only Any pump system, carrying case, and connection fittings for an XP2i pressure gauge may be ordered separately from the gauge. Enter XP2I-NONE followed by the Pump System part number and the Carrying Case option code. SAMPLE PART NUMBER KPA versions available in USA direct from factory only. Page 4 of 5 sensors, test & calibration

 Open the catalog to page 4
XP2i Digital Pressure Gauge kPa Data Sheet-5

Crystal Wpressure #1%| PUMP SYSTEMS OVERVIEW XP2i Pressure Gauge kPa Case Options © 2020 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701 Aluminum Waterproof (Pelican Case)

 Open the catalog to page 5

All Ametek Calibration catalogs and technical brochures

  1. FastCalXP

    9  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.