XP2i Digital Pressure Gauge bar Data Sheet-1

PANEL MOUNT FLANGE MAX PANEL THICKNESS 11.7 (0.46) Standard Accuracy Version (-10 to 50° C) 0 to 20% of Range: ± (0.02% of Full Scale*) 20% to 110% of Range: ± (0.1% of Reading) Vacuum: ± (0.25% of Full Scale**) -S5 Accuracy Option (-10 to 50° C) Vacuum to 110% of Range: ± (0.05% of Full Scale*) -S2 Accuracy Option Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Not recommended for continuous use below -1.0 bar. Refer to XP2i-DP data sheet for gauges that are intended for continuous high vacuum use. **Full Scale = -1.0 bar SENSOR DIAPHRAGM SURFACE OPERATING TEMPERATURE DISPLAY Single Line Display: 16.9 mm (0.67") high Dual Line Display: Main Display: 13.9 mm (0.55") Secondary Display: 5.3 mm (0.21") Display Rate: 4 readings/second (standard) 8 readings/second (PSV mode) *Full Scale is 100% of the range Vacuum to 110% of Range: ± (0.02% of Full Scale*) 15 to 25° C ± (0.05% of Full Scale*) -10 to 15° C, 25 to 50° C, typical PORT POSITION FOR -F4 AND -RP OPTIONS All models indicate vacuum, but vacuum specification applies to 1 bar, 2 bar, 7 bar, and 20 bar models only. Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification. LCD readable in sunlight with bright backlight. PSV mode is intended for Pressure Safety Valve testing, and requires software to enable. (ConfigXP is a free download from ametekcalibration.com. ) 1/4 INCH MBSP ADAPTER ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS

 Open the catalog to page 1
XP2i Digital Pressure Gauge bar Data Sheet-2

| POWER The XP2i is Intrinsically Safe only if powered by one of the following battery types. Approved Battery Type IECITe&x Ex ia IIC T4/T3 Ga IECEx FTZU 12.0009X Approved Battery Type Exia Intrinsically Safe and Non-incendive for Hazardous Locations: Class I, Division 1, Groups A, B, C and D, Temperature Code T4 or T3C. For hazardous location product warnings, refer to the operation manual. XP2i complies with the Electromagnetic Compatibility and the Pressure Equipment Directives. Ultra Low Power: > 1 year, typical** Uses 3 alkaline AA (LR6) batteries. Use of backlight and/or RS232 reduces...

 Open the catalog to page 2
XP2i Digital Pressure Gauge bar Data Sheet-3

XP2i Pressure Gauge bar |ENCLOSURE Weight includes batteries. Nickel plating over low copper, marine grade aluminum. LCD protected from impact damage by 1.5 mm (0.06") thick polycarbonate lens under polyester window. Skydrol® compatible. For hazardous location product warnings, refer to the operation manual. Wetted Materials: (wrench tight) 316 stainless steel (finger tight) 316 stainless steel and Viton® (internal o-ring) Diaphragm Seal Fluid: Silicone Oil Connection: Crystal CPF* Female Temperature Range: -40 to 75° C (-40 to 167° F) Batteries should be removed if stored for more than one month....

 Open the catalog to page 3
XP2i Digital Pressure Gauge bar Data Sheet-4

generation includes everything from small pneumatic hand pumps to a precision, hydraulic pressure comparator capable of generating up to 15 000 psi / 1000 bar / 100 MPa. All of our pumps may be ordered as part of a Pump System, complete with an XP2i and delivered in a sturdy carrying case with custom insert. * Refer to the following page for a more detailed description of each pump system.

 Open the catalog to page 4
XP2i Digital Pressure Gauge bar Data Sheet-5

Crystal Wpressure #1%| PUMP SYSTEMS OVERVIEW XP2i Pressure Gauge bar Case Options Pneumatic Hydraulic Hand Pump Bench Top Included Pump Aluminum Waterproof (Pelican Case) © 2020 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701