SENSORS, TEST & CALIBRATION Reference manual Reference Temperature Calibrator Jofra RTC-156/157/158/159/168/187/250/700
Open the catalog to page 1Reference Manual Reference Temperature Calibrator JOFRA RTC-156/157/158/159/168/187/250/700 A/B/C Copyright 2012 AMETEK Denmark A/S
Open the catalog to page 2About this manual…. • The structure of the manual This manual is divided into 11 sections. These describe how to set up, operate, service and maintain the calibrator. The technical specifications are described and accessories may be ordered from the list of accessories. Along with the calibrator, you should have received a user manual, which sets out the operating instructions for the instrument. It is designed to provide a quick reference guide for use in the field. Safety symbols This manual contains a number of safety symbols designed to draw your attention to instructions that must be followed...
Open the catalog to page 3Congratulations on your new AMETEK JOFRA RTC Calibrator! With this AMETEK JOFRA calibrator, you have chosen an extremely effective instrument, which we hope will live up to all your expectations. This RTC calibrator is a fast, timesaving, and reliable true industrial temperature calibrator designed for on-site use. Over the past many years, we have acquired extensive knowledge of industrial temperature calibration. This expertise is reflected in our products, which are all designed for daily use in an industrial environment. Please note that we are very interested in hearing from you, if you...
Open the catalog to page 6Patented temperature calibrator technology. • Patented DLC system that insures high uniformity in the calibration zone. All models. • Patented fast cooling technology. RTC-700 only. • Patented cooling technology giving the broadest temperature range for ultracooler calibrators. RTC-159 only. Beside patents the RTC series offers unique Jofra features and functions: • Automatic Reference sensor protection • Automatic Intelligent Recalibration Information - IRI • Selectable Silent or fast mode operation AMETEK Denmark A/S was ISO-9001 certified in September 1994 by Bureau Veritas Certification Denmark....
Open the catalog to page 7This instrument is warranted against defects in workmanship, material and design for two (2) years from date of delivery to the extent that AMETEK will, at its sole option, repair or replace the instrument or any part thereof which is defective, provided, however, that this warranty shall not apply to instruments subjected to tampering or, abuse, or exposed to highly corrosive conditions. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED AND AMETEK HEREBY DISCLAIMS ALL OTHER WARRANTIES, INCLUDING, WITHOUT LIMITATION, ANY WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR...
Open the catalog to page 8• RTC-158/168/250 A/B/C only (liquid bath) - OPTIONAL : • 1 liquid bath kit consisting of : - 1 sensor basket with temperature equalizer tube - 2 lids for transportation / calibration (RTC-158/250) - 1 insulating ring for spill tray (RTC-168) - 1 stirring magnet remover (RTC-158/250) - 1 liquid drainage syringe - 1 oil material safety data sheet When reordering, please specify the part numbers according to the list of accessories, section 11.0. Optional parts can also be found in the list of accessories Caution Do not use the RTC-158/168 insulation plug (black POM) with the RTC-250 instrument...
Open the catalog to page 9Dimensioning drawings
Open the catalog to page 10Read this manual carefully before using the instrument! Please follow the instructions and procedures described in this manual. They are designed to allow you to get the most out of your calibrator and avoid any personal injuries and/or damage to the instrument. The screen menus shown in this manual represent the menus displayed when using a B-version. /v-®\ These calibrators contain Electrical and Electronic circuits and must be recycled or disposed of properly (in accordance with the WEEE Directive 2012/19/EU). • Always supply the calibrator using a power circuit which is separate from essential...
Open the catalog to page 11• Never use heat transfer fluids such as silicone, oil, paste, etc. in the dry-block calibrators. These fluids may penetrate the calibrator and cause electrical hazard, damage or create poisonous fumes. • The calibrator must be switched off before any attempt to service the instrument is made. There are no user serviceable parts inside the calibrator. • When cleaning the well or the insertion tube, REMEMBER to wear goggles when using compressed air in the dry-block calibrator and cleaning oil in the liquid bath calibrator. • Use protection shield when calibrating at high temperatures (RTC-700)...
Open the catalog to page 12• Liquid baths should only be operated by trained personal. • Heat transfer fluids must only be used in calibrators prepared as a liquid bath. If these fluids are heated above specified temperature they will create noxious or toxic fumes. Proper ventilation must be used. • To avoid hazards from treating fluids in a wrong manner, always reduce the "Max. SET-temperature allowed” in the CALIBRATOR SETUP MENU according to the specifications of the fluid to be used. If using a calibrator outside of the fluids specifications there is a risk of fire hazards, personal Injury or chemical release. By reducing...
Open the catalog to page 13To prevent this from happening, the insertion tube and the well must be dried. This is done by heating up the calibrator to min. 100°C/212°F until all water left has evaporated. Remove the insulation plug while heating up. It is very important that humidity in the well and insertion tube is removed to prevent corrosion and frost expansion damages. • Do not use the instrument if the internal fan is out of order. • Before cleaning the calibrator, you must switch it off, allow it to cool down and remove all cables. • Always place power cable and cables to attached accessories, such as sensors, in...
Open the catalog to page 14• Scratches and other damage to the insertion tubes should be avoided by storing the insertion tubes carefully when not in use. • The insertion tube must never be forced into the well. The well could be damaged as a result, and the insertion tube may get stuck. • Before using new insertion tubes for the calibration, the insertion tubes must be heated up to maximum temperature - 250°C (482°F) / 700°C (1292°F) - for a period of minimum 30 minutes (RTC-250/700 A/B/C only). • The insertion tube must always be removed from the calibrator after use. The humidity in the air may cause corrosion oxidation...