SENSORS, TEST & CALIBRATION
Open the catalog to page 1AMETEK continues to develop new techniques to improve the performance, accuracy, convenience, and functionality of the popular Jofra calibration products. By doing so, we maintain our position as the leading worldwide manufacturers of temperature dry-block calibrators. We are proud to introduce our top model RTC (Reference Temperature Calibrator) series, which is more sophisticated than any existing calibrators. Features & Models The RTC offers many new fantastic features such as: ■ Patented DLC (Dynamic Load Compensation) system for perfect temperature uniformity in the insert. ■ Unique, intelligent...
Open the catalog to page 2Liquid Bath/Large Diameter Insert The RTC-158/250 calibrators are fitted with a 160 mm (6.3") deep well with a diameter of 63.5 mm (2.5"), which are twice the size of any other dry-block. And, they can be used both as dry-block calibrators and as liquid calibration baths with a magnetic stirrer. With these options, it is now possible to calibrate even more temperatures sensors simultaneously and to calibrate both large and odd-shaped sensors—which is not possible with the remaining product range. The RTC-158/250 calibrators can be used without an external reference sensor. But if an STS-200 reference...
Open the catalog to page 3DLC–Dynamic Load Compensation DLC To bring our well documented active dual-zone technology to an even higher level, we have developed the patented DLC system. This feature makes it possible to perform top calibration specifications without being affected by the actual load, e.g. many sensors or very big sensors. The DLC sensor improves on the RTC calibrator’s already advanced dual-zone technology by controlling the homogeneity in not only the well, but inside the insert where the sensors-under-test are placed during calibration. The DLC sensor measures the temperature homogeneity in the insert...
Open the catalog to page 4Unique Temperature Performance Upper Zone the reference sensor and be up and running immediately. The RTC series of calibrators provide precision With the intelligent sensors, AMETEK has eliminated a source of error, and the system is now giv- temperature calibration of sensors, what- ing a fail-safe plug’n’play calibration system. ever the type or format. This is accomplished Unique Reference Sensors through an innovative active dual-zone heatTemperature Difference Main Zone downloading sequence with risk of errors is no longer necessary. Secondly, the user can change The STS-200 reference sensors...
Open the catalog to page 5Fast Temperature Calibration Intelligent Recalibration Information Time is money! This is why all the RTC calibrators have an increased heating and cooling speed In order to comply with ISO, SOP’s and the FDA, it is imperative that the calibration equipment compared to all other calibrators. Heating and cooling speeds have been increased by up to 25%. never exceeds the expiry of the calibration certificate. RTC Series calibrators are constantly check- The implication is savings in both production downtime and general calibration costs. ing calibration dates on the calibrator, as well as any connected...
Open the catalog to page 6The user can decide whether to read the built-in reference sensor or the more accurate angled reference sensor from the large, easy-to-read LCD display of the calibrator. The external sensor The integrated support rod is part of the reduced weight and the internal sensor readings are independent of one another. philosophy. It is lightweight and very easy to mount on the SET-Follows-TRUE (models B & C only) RTC. Two fixing holes are integrated in the calibrator where the support rods can be mounted. The “SET-Follows-TRUE” makes the instrument tune in until the temperature reading of the external...
Open the catalog to page 7Maximum and Minimum Temperature Up to 20 different temperature steps may be pro- From the setup menu, the user can select the maximum and minimum temperature limit for the grammed including the hold time for each step. calibrator. This function prevents damage to the sensor-under-test caused by excessive tempera- Upon completion of an auto-step routine, the user tures, and it helps reducing sensor drift from exposures of too high temperatures. This feature can can easily read the results for the sensor-under- be locked with an access code. test on the RTC display. Results from twenty autostep...
Open the catalog to page 8JofraCal Calibration Software JofraCal is a highly versatile calibration software that is supplied together with the RTC calibrators. The software ensures easy calibration of all kinds of temperature sensors, such as RTD's, thermocouples, transmitters, and thermoswithes. Furthermore, it can be used for pressure calibration i.e. pressure gauges and pressure switches. JofraCal integrates with Jofra calibration instruments. As for temperature calibrators, it is the whole range of temperature calibrators. Regarding pressure calibrators, it integrates with the Crystal XP2i and nVision. JofraCal also...
Open the catalog to page 9Specifications Functional Specifications Temperature Range Accuracy with External STS Ref. Sensor (models B and C) RTC-158 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.04°C/±0.07°F RTC-250 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.07°C/±0.13°F 12-month period. Relative to reference standard. Specifications by use of the external JOFRA STS-200 reference sensor. Accuracy with Internal Ref. Sensor (models A, B, and C) RTC-158 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.18°C/±0.32°F RTC-250 ....
Open the catalog to page 10Specifications RTD Sensor Under Test Input (model B) F.S. (range). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400 ohm Accuracy (12 months). . . . . . . . . . . . . ±(0.002% Rdg.+0.002% F.S) F.S. (range). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4000 ohm Accuracy (12 months). . . . . . . . . . . . ±(0.005% Rdg. + 0.005% F.S.) 2-wire. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . add 50 mOhm Temperature RTD Type Input and curves for many different resistance sensors such...