1. Catalogs
  2. Ametek Calibration
  3. nVision Reference Recorder MPa Datasheet

nVision Reference Recorder MPa Datasheet

nVision Reference Recorder MPa Datasheet
1 / 10 PagesView full catalog

nVision Reference Recorder MPa Datasheet

Product catalog summary
Pressure Module (PM) Specifications:
The accuracy of MPa (Gauge Pressure) modules varies by range: from 0 to 30% of the range, the accuracy is ± (0.0075% of Full Scale), and from 30 to 110% of the range, it is ± (0.025% of Reading). Vacuum specifications are applicable only to the 300 kPa, 1 MPa, and 3 MPa models. The accuracy of MPaA (Pressure with BARO module) varies by module and range.
Differential Pressure Measurement:
The Tare function enhances measurement uncertainties when using two modules with the same full-scale pressure range. Without Tare, uncertainties are calculated using the root sum squares method.
Sensor Details:
Wetted materials include 316 stainless steel and Viton®. The barometric reference accuracy is ± 0.5 mbar.
Current, Voltage, & Switch Test Module (MA20):
This module measures 4-20mA loop, supply voltages, and switch open/closure. The current accuracy is ± (0.015% of reading + 0.002 mA), and the voltage accuracy is ± (0.015% of reading + 0.002 VDC).
Temperature Module (RTD100):
Calibrated for Pt100 RTD sensors conforming to DIN/IEC 60751, with an accuracy of ± (0.015% of reading + 0.02 Ω).
nVision Chassis (NV):
Operating temperature ranges from -20 to 50°C. It features a 255 x 160 pixel graphical display and is intrinsically safe with specific battery types.
Data/Communication:
Includes a digital interface via mini-USB and a datalogging capacity of approximately 1,000,000 data points.
Enclosure:
Weighs 680 g and is IP66 and IP67 rated with impact-resistant housing.
Certifications:
ATEX and IECEx certified for hazardous locations, complying with Electromagnetic Compatibility and Pressure Equipment Directives.
Standard Delivery:
Includes nVision Recorder, CD Manual, Calibration Certificate, and accessories.
Complementary Products:
Offers fittings, hoses, pumps, and software for calibration.
Overview: This document provides detailed specifications and ordering information for nVision pump systems and reference recorders, including descriptions of various modules, pump systems, and case options available for customization.
Specifications:
  • Modules: Lists different modules such as Pressure (PM), Current, Voltage, and Switch (MA20), Temperature (RTD100), and Blank Plate (BNKPLT), each with specific configurations and options.
  • Pressure Range and Resolution: Details the pressure range and display resolution for different pressure modules, ranging from 300 kPa to 100 MPa.
Ordering Information:
  • Sample part numbers illustrate how to order nVision systems with specific module configurations and options.
  • For systems with only one module, a blank plate (BNKPLT) must be specified for the empty slot.
Pump Systems Overview:
  • Outlines various pump systems (A to H) with their respective pressure ranges and case options, including specific part numbers and pressure capabilities, ranging from pneumatic to hydraulic hand pumps.
  • Case options include aluminum and waterproof (Pelican Case) for certain systems.
Additional Information:
  • The RTD100 temperature module does not include a probe; options for probes are listed with part numbers and descriptions.
  • Ordering information specifies that MPa versions are available directly from the factory in the USA.
See more

Catalog excerpts

nVision Reference Recorder MPa Datasheet-1

nVision Reference Recorder MPa Pressure Module (PM) ACCURACY MPa (Gauge Pressure) MPaA (Pressure with BARO module) 0 to 30% of Range: ± (0.0075% of Full Scale) 30 to 110% of Range: ± (0.025% of Reading) Vacuum: For 300 kPa and 1 MPa ± (0.06% of Full Scale*) For 3 MPa ± (0.06% of Full Scale*) ±1 LSD * Full Scale = -99.9 kPa 10, 30, 70, and 100 MPa modules 0 to 30% of Range: ± (0.015% of Full Scale) 30 to 110% of Range: ± (0.05% of Reading) 1.38 to 100.00 kPaA: ± 0.08 kPaA 100.00 to 400.00 kPaA: ± (0.025% of Reading) + 0.03 kPaA 1 MPa module 0.00138 to 0.10000 MPaA: ± 0.00008 MPaA 0.10000 to 0.40000 MPaA: ± 0.00010 MPaA 0.40000 to 1.10000 MPaA: ± (0.025% of Reading) 3 MPa module 0.0014 to 0.1000 MPaA: ± 0.0001 MPaA 0.1000 to 1.000 MPaA: ± 0.0003 MPaA 1.000 to 3.1000 MPaA: ± (0.025% of Reading) Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. All models indicate vacuum, but vacuum specification applies to 300 kPa, 1 MPa, and 3 MPa models only. Not recommended for continuous use at high vacuum. Refer to XP2i-DP data sheet for gauges that are intended for continuous high vacuum use. 10 MPa module 0.1000 to 3.1000 MPaA: ± 0.0015 MPaA 3.1000 to 10.1000 MPaA: ± (0.05% of Reading) 30 MPa module 0.100 to 9.100 MPaA: ± 0.005 MPaA 9.100 to 30.100 MPaA: ± (0.05% of Reading) 70 MPa module 0.100 to 21.100 MPaA: ± 0.011 MPaA 21.100 to 70.100 MPaA: ± (0.05% of Reading) 100 MPa module 0.100 to 30.100 MPaA: ± 0.015 MPaA 30.100 to 100.100 MPaA: ± (0.05% of Reading) 4793.Q 2302 • nVision MPa Page 1 of 10

 Open the catalog to page 1
nVision Reference Recorder MPa Datasheet-2

nVision Reference Recorder MPa| DIFFERENTIAL PRESSURE MEASUREMENT UNCERTAINTIES WITH TARE The Tare function can improve measurement uncertainties on two modules with the same full scale pressure range installed into one nVision Reference Recorder. Requires the use of an equalizing valve. The following specifications apply to the measurement system with a logging interval of 1 second/reading: Full Scale Range of Both Sensors SENSORS, TEST & CALIBRATION

 Open the catalog to page 2
nVision Reference Recorder MPa Datasheet-3

nVision Reference Recorder MPa SENSOR Wetted Materials: ( wrench tight) 316 stainless steel (finger tight) 316 stainless steel and Viton® (internal o-ring) Diaphragm Seal Fluid: Silicone Oil Connection: Crystal CPF Female REFERENCE (BARO) BAROMETRIC Accuracy: ± 0.5 mbar, ± 0.00725 psi Range: 700.0 to 1100.0 mbarA, 10.153 to 15.954 psiA Units and Resolution: mbar. . . . . . . . . . . . . . . psi. . . . . . . . . . . . . . . . . . inHg. . . . . . . . . . . . . . . . mmHg. . . . . . . . . . . . . Pressure Connection: All welded, with a permanently filled diaphragm seal. Metal to metal cone seal;...

 Open the catalog to page 3
nVision Reference Recorder MPa Datasheet-4

nVision Reference Recorder MPa Current, Voltage, & Switch Test Module (MA20) Intended for use with a 4-20mA loop measurement. This module is also capable of measuring supply voltages and has an auxiliary fixed output for use in switch open/closure testing. Each MA20 module includes a super flexible silicone test lead kit (P/N 3952). CURRENT & VOLTAGE MEASUREMENT Current (mA) Input Accuracy: ± (0.015% of rdg + 0.002 mA) Range: 0 to 55 mA (MA20) Max Allowable Current: 93.3 mA Resolution: 0.001 mA or 0.01% Units: mA, % 4-20, % 10-50 Input Resistance: < 17.2 Ω Voltage Burden @ 20mA: < 0.35 V Includes...

 Open the catalog to page 4
nVision Reference Recorder MPa Datasheet-5

nVision Reference Recorder MPa Temperature Module (RTD100) Calibrated for Pt100 RTD / PRT (100 Ohms at 0°C Platinum Resistance Temperature Detector) sensors conforming to DIN/ IEC 60751 (or IEC751) with US, Euro, or Lab calibration curves. An RTD is not included, but each RTD100 includes P/N 3953 RTD Connection Kit. | TEMPERATURE MEASUREMENT Resistance Input Range: 0 - 400 Ohms range for use with 100 Ohm PRTs Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Resolution: 0.01 on all scales Units: °C, K, °F, R, n TCRs: 0.003850 n/n/°C (IEC 60751),...

 Open the catalog to page 5
nVision Reference Recorder MPa Datasheet-6

Crystal \|(pressure #1% nVision Reference Recorder MPa nVision Chassis (NV)| OPERATING TEMPERATURE Temperature Range: -20 to 50° C (-4 to 122° F) < 95% RH, non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification. | DISPLAY Screen: 255 x 160 pixel graphical displayDisplay Rate: 4 readings/second (standard)up to 10 readings/second (recording) LCD readable in sunlight with bright backlight. | POWER The nVision is Intrinsically Safe only if powered by one of the following battery types. Approved Battery Type 4 x AA: 200 hours, typical...

 Open the catalog to page 6
nVision Reference Recorder MPa Datasheet-7

nVision Reference Recorder MPa DATA/COMMUNICATION Digital Interface: mini-USB DATALOGGING Capacity: Approx. 1,000,000 data points* Storage Type: Non-volatile flash memory Fastest Interval: 10 per second Slowest Interval: 1 per hour ENCLOSUREWeight: 680  g (24.0 oz) Rating: IP66 and IP67 Housing: Impact resistant injection molded Keypad and Labels: UV Resistant Polyester Mounting: M4 x 0.7 [8 mm (0.31")] deep threaded insert mounting locations The mini USB will power the nVision with or without the battery pack installed. For hazardous location product warnings, refer to the operation manual....

 Open the catalog to page 7
nVision Reference Recorder MPa Datasheet-8

nVision Reference Recorder MPa | STORAGE TEMPERATURE Temperature Range: -40 to 75° C (-40 to 167° F) Batteries should be removed if stored for more than one month. The following requires the use of our free CrystalControl software | CERTIFICATIONS II 1G Ex ia IIB T4 Ga or T3 SIRA 09 ATEX 2008X Averaging Screen: Data Point Counter: Display Screens: Estimated Recording Time: Live PC Graph: Password Protect: Pressure Switch Test: Remove: Run Tags: Screen Numbers: Secure Documents: Start-up Screen: User Defined Unit: Averages all points in a recording run. Screen for counting the data points logged....

 Open the catalog to page 8
nVision Reference Recorder MPa Datasheet-9

| RANGE & RESOLUTION TABLE nVision Reference Recorder MPa Display Resolution No Pump .. System A .. System A .. System B . . System B . . System C . . System C . . System D System D System E . . System F . . System F . . System G .. System G .. System H . • The Waterproof Case is an option for Systems A, B, and C only . The Waterproof Case is the only option for Systems G and H . AMETEK offers a variety of solutions for pressure generation and measurement. Our line of products for pressure generation includes everything from small pneumatic hand pumps to a precision, hydraulic pressure comparator...

 Open the catalog to page 9

All Ametek Calibration catalogs and technical brochures

  1. FastCalXP

    9  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.