HPC50 Series Pressure Calibrator Datasheet MPa
1 /7Pages

HPC50 Series Pressure Calibrator Datasheet MPa

HPC50 Series Pressure Calibrator Datasheet MPa
1 /7Pages

Catalog excerpts

HPC50 Series Pressure Calibrator Datasheet MPa-1

ACCURACY • PRESSURE MEASUREMENT MPa (Gauge Pressure) X 18 to 28° C 30 to 110% of Range: ±(0.035% of Reading) Vacuum* : ±(0.05% of Full Scale**) X -20 to 50° C 0 to 30% of Range: ±(0.015% of Full Scale) 30 to 110% of Range: ±(0.050% of Reading) Vacuum* : ±(0.05% of Full Scale**) * Applies to 3 MPa and lower ranges only. Vacuum Range = -1 MPa. ** Full Scale is the numerical value of the positive pressure range. Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. All models indicate vacuum, but vacuum specification applies to 100 kPa, 300 kPa, 1 MPa, and 3 MPa models only. APMi and ATMi MODULE CONNECTIONS Not recommended for continuous use at high vacuum. Refer to XP2i-DP data sheet for gauges that are intended for continuous high vacuum use. The BARO option allows you to toggle between gauge and absolute pressure. Exposure to environmental extremes of temperature, shock, and/ or vibration may warrant a more frequent recertification period. SENSOR DIAPHRAGM SURFACE APMi modules must be exercised and re-zeroed whenever exposed to significant changes in environmental conditions to achieve these specifications. To exercise a module, cycle the module between zero (ambient barometric pressure) and the pressure of interest. A properly exercised module will return to a zero reading (or return to the same ambient barometric reading). All absolute accuracies are equivalent to the gauge pressure accuracies, except as noted below. 300 kPa Range: Gauge Accuracy + 0.03 kPaA 1 MPa Range: Gauge Accuracy + 0.00001 MPaA MPaA (Absolute Pressure with BARO Option) 100 kPa Range: Gauge Accuracy + 0.03 kPaA 0 to 30% of Range: ±(0.01% of Full Scale) mA CONNECTIONS Ø 4 mm SHIELDED BANANA JACKS 1/4 INCH MNPT ADAPTER 1/4 INCH MBSP ADAPTER ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS SENSORS, TEST

 Open the catalog to page 1
HPC50 Series Pressure Calibrator Datasheet MPa-2

| DIFFERENTIAL PRESSURE The Tare function can improve differential pressure measurement uncertainties. Requires the use of an equalizing valve. Full Scale Range of Both Sensors STANDARD DELIVERY_ • ISO 17025 Accredited Calibration Certificate, NIST Traceable • Your choice of adapters (1/4" NPT, 1/4" BSP, or 1/4" M20) • Protective Boot—required for Intrinsic Safety • Test Leads, red and black with clips • Velco strap • User manual • Mini-USB Cable All welded construction on sensors above 300 kPa. (The 100 kPa sensor may have Viton o-ring seal.) Metal to metal cone seal; O-ring can be removed if...

 Open the catalog to page 2
HPC50 Series Pressure Calibrator Datasheet MPa-3

mA CONNECTIONS Ø 4 mm SHIELDED BANANA JACKS CURRENT & VOLTAGE MEASUREMENT Connection: 4 mm jacks Current (mA) Input Accuracy: mA Range: Percent Range: Max Allowable Current: Resolution: Units: Input Resistance: Voltage Burden @ 20mA: Voltage Burden @ 50mA: HART Resistor: ±(0.015% of rdg + 0.002 mA) 0 to 55 mA 0-20, 4-20, 10-50 93.3 mA 0.001 mA or 0.01% mA, scaling, % error, and % flow < 4.99 Ω < 0.10 V < 0.250 V 250 Ω Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Inputs protected by a resettable fuse. mA can be displayed as a percentage,...

 Open the catalog to page 3
HPC50 Series Pressure Calibrator Datasheet MPa-4

EXTERNAL MODULES The HPC50 Series has two identical ports to connect external pressure or temperature modules. For details on the modules, see the links below. Pressure Measurement APMi AND ATMi CONNECTIONS See the APMi datasheet. Temperature Measurement Digital Interface: mini-USB The mini USB will power the HPC50 Series with or without the batteries installed. Do not use mini USB connection in a hazardous area. For hazardous location product warnings, refer to the operation manual. Screen: 320 x 240 pixel graphical display Display Rate: 3 readings/second (standard) LCD readable in sunlight....

 Open the catalog to page 4
HPC50 Series Pressure Calibrator Datasheet MPa-5

Cell Voltage: 1.5 V (Alkaline Batteries) CERTIFICATIONS Uses 3 alkaline AA (LR6) batteries. ATEX BATTERY LIFE No External Modules: 40 Hours One External Module: 25 Hours All hours are based on operation without the use of the backlight. Use of the backlight will decrease battery life. II 1G IEx ia IIC T4/T3 Ga FTZU 18 ATEX 0043X Ex ia IIC T4/T3 Ga IECEx FTZU 18.0012X Two External Modules: 12 Hours ENCLOSURE Weight: 567 g (20.0 oz) Rating: IP66/67 Housing: PC/PBT plastic Weight is for dual sensor model with protective boot installed. LCD protected from impact damage by 0.5 mm (0.02") thick polycarbonate...

 Open the catalog to page 5
HPC50 Series Pressure Calibrator Datasheet MPa-6

Display Resolution (Add one digit of resolution for differential mode. ) Pump System* No Pump .. System A.. System A.. System B . . System B . . System C . . System C . . System D.. System D.. System E . . System F . . System F . . System G .. System G .. System H . pressure fitting. HPC52-30MPA-70MPA-BARO-BSP ... Dual Sensor (30 MPa/70 MPa) HPC50 with the BARO option and a 1/4" male BSP pressure fitting. NPT pressure fitting; a System G pump system; and a waterproof carrying case. AMETEK offers a variety of solutions for pressure generation and measurement. Our line of products for pressure...

 Open the catalog to page 6
HPC50 Series Pressure Calibrator Datasheet MPa-7

PUMP SYSTEMS OVERVIEW Pump System Case Options Part Number Pressure Range Hand Pump Included Pump Waterproof (Pelican Case) ametekcalibration.com © 2022 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Ob

 Open the catalog to page 7