Save Time, Save Money Best In Class Industrial Calibrator Fast Calibration The new generation CTC temperature calibrators All Jofra temperature calibrators feature a feature improved accuracy by as much as 25%. They purpose-dedicated temperature regulator, are now the most accurate industrial class tempera- which provides a very fast heating and cooling ture calibrator on the market. time, and a short stabilization time. Performing a three point temperature calibration procedure is fast and easy. Wide Temperature Range The CTC series covers a wide temperature range The CTC series is designed...
Open the catalog to page 2Intelligent Accuracy External Reference Sensor Self Calibration Each CTC calibrator has a C model that includes a Adjust the internal reference sensor to give the signal input for an external reference sensor. The same readings as the external reference sensor by sensor improves the accuracy even more. Our exter- performing a Self Calibration. After the Self Calibra- nal reference sensors have been developed to match tion the calibrator can be used without the external reference sensor. Regular Self Calibration is needed to ensure the accuracy. The Self Calibration is simple and fast to perform...
Open the catalog to page 3■ Multi-Information Display Status Bar - Shows information about recalibration due status, hot and cold safety warnings, and date and time. Calibration Settings - Shows the calibration settings and stability criteria for the reference sensor. Can be accessed and changed directly in the display by use of the arrow keys. Calibration Status Shows current status of the calibrator, like heating or cooling, expected time to stability, or stability achieved. Shows the current possibilities of the function keys. The function keys serve as shortcuts to the main functions like Set, Preset, Auto step, and...
Open the catalog to page 4Useful Features The CTC is a very versatile calibrator series with many integrated functions - you can run the calibration in four different ways. Preset Mode MVI – Mains Power Variance Immunity improves temperature stability Unstable mains power supplies are a major contributor to calibration inaccuracies. Traditional temperature calibrators often become unstable in industrial environments where large electrical motors, heating elements, and other devices are periodically cycled on and off. The cycling of supply power can cause lower quality temperature regulators to perform inconsistently,...
Open the catalog to page 5I Special Features Multi Sensor Calibration Insert heat loss due to mass loading from multiple or large sensors can be a challenge for most dry-block calibrators. The CTCs advanced feedback algorithms combined with the external reference sensor effectively addresses these challenges and makes the accuracy even better. When switching on the calibrator or connecting the reference sensor, the calibrator immediately warns you if any of the calibration certificates are overdue. A buzzer and warning appears. The recalibration interval can be set from 1 to 99 months. All STS reference sensors are plug...
Open the catalog to page 6User Specified Settings Calibration Software Included The CTC is supplied with our highly versatile calibration software JofraCal. All calibrations can be documented with a Silent Mode Operation The CTC calibrator can be programmed to run in silent operation. This function is an advantage if calibrating in a laboratory or an office. If used in silent operation the calibrator is not using its full cooling potential. Online Firmware Upgrade certificate, given that the CTC is controlled from a You can download firmware and software upgrades at type in the UUT temperature. JofraCal documents www.ametekcalibration.com....
Open the catalog to page 7Specifications CTC-155 JOFRAW calibration §t\ Please note: All specifications are given with an ambient temperature 23°C/73.4°F ± 3°C/5.9°F and specified at 115 V/230 V. ★) In lower 30 mm incl. Stability, Uniformity, 12-month drift (typical), Hysteresis, Resolution, Load (Max 6 mm) and calibration laboratory uncertainty. ★ ★) Measured after the stability indicator has been on for 10 minutes. Measuring time is 30 minutes. ★ ★★) Stability when exposed to maximum temperature for 100 hours. Stability will depend on actual use of the probe. Sensing Element Response Time Sensing Element Response Time...
Open the catalog to page 8Specifications CTC-350 Temperature Temperature Range Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 to 350°C / 82 to 662°F Lowest calibration temperature . . . . . . . . . .ambient +5°C / 41°F Accuracy CTC-350 With Internal Reference Sensor @ 350°C / 662°F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.45°C /±0.81°F @ 200°C / 392°F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.40°C /±0.72°F CTC-350 With External Reference Sensor STS-120-A-935 @ Full Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.25°C /±0.45°F Stability...
Open the catalog to page 9Mains Power Voltage . . . . . . . . . . . . . 115 V (103-127) / 206 V (180-254) Max Power Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1150 VA Frequency, US deliveries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Hz ±3 Frequency, non US deliveries . . . . . . . . . . . . . . .50 Hz ±3, 60 Hz ±3 Physical Specifications Dimension L x W x H . . 248x148x390 mm / 9.76x5.83x15.35 in Weight CTC-652 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.0 kg / 19.8 lb Insert Dimensions (diameter x length) CTC-652 . . . . . . . . ....
Open the catalog to page 10Mains Power Voltage . . . . . . . . . . . . . . . . . . . . . 115 V (90-127) / 230 V (180-254) Max Power Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1150 VA Frequency, US deliveries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Hz ±3 Frequency, non US deliveries . . . . . . . . . . . . . . .50 Hz ±3, 60 Hz ±3 Physical Specifications Dimension L x W x H . . 248x148x305 mm / 9.76x5.83x12.01 in Weight CTC-660 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1 kg / 13.4 lb Immersion Depth CTC-660 . . . . . . . . . . . . . . ....