Group: AMETEK
Catalog excerpts

XP2i Pressure Gauge psi U.S. PATENT D612277 ALL DIMENSIONS IN MM (IN) PANEL MOUNT FLANGE MAX PANEL THICKNESS 11.7 (0.46) Standard Accuracy Version (-10 to 50° C) 0 to 20% of Range: ± (0.02% of Full Scale*) 20% to 110% of Range: ± (0.1% of Reading) Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Vacuum: ± (0.25% of Full Scale**) -S5 Accuracy Option (-10 to 50° C) PORT POSITION FOR -F4 AND -RP OPTIONS All models indicate vacuum, but vacuum specification applies to 15 psi, 30 psi, 100 psi, and 300 psi models only. Not recommended for continuous use below -14.5 psi. Refer to XP2i-DP data sheet for gauges that are intended for continuous high vacuum use. Vacuum to 110% of Range: ± (0.05% of Full Scale*) -S2 Accuracy Option **Full Scale = -14.5 psig Vacuum to 110% of Range: ± (0.02% of Full Scale*) 15 to 25° C ± (0.05% of Full Scale*) -10 to 15° C, 25 to 50° C, typical SENSOR DIAPHRAGM SURFACE Single Line Display: 16.9 (0.67") high mm Dual Line Display: Main Display: 13.9 mm (0.55") Secondary Display: 5.3 mm (0.21") Display Rate: 4 readings/second (standard) 8 readings/second (PSV mode) *Full Scale is 100% of the range Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification. LCD readable in sunlight with bright backlight. PSV mode is intended for Pressure Safety Valve testing, and requires software to enable. (ConfigXP is a free download from ametekcalibration.com. ) 1/4 INCH MBSP ADAPTER ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS
Open the catalog to page 1
XP2i Pressure Gauge psi | POWER The XP2i is Intrinsically Safe only if powered by one of the following battery types. Approved Battery Type Approved Battery Type Exia Intrinsically Safe and Non-incendive for Hazardous Locations: Class I, Division 1, Groups A, B, C and D, Temperature Code T4 or T3C. For hazardous location product warnings, refer to the operation manual. XP2i complies with the Electromagnetic Compatibility and the Pressure Equipment Directives. Ultra Low Power: > 1 year, typical** Uses 3 alkaline AA (LR6) batteries. Use of backlight and/or RS232 reduces operating time....
Open the catalog to page 2
XP2i Pressure Gauge psi |ENCLOSURE Weight includes batteries. Submersible to I m for30 minutes [IEC 60529]. Nickel plating over low copper, marine grade aluminum. LCD protected from impact damage by 1.5 mm (0.06") thick polycarbonate lens under polyester window. Skydrol® compatible. For hazardous location product warnings, refer to the operation manual. |SENSOR Wetted Materials: (wrench tight) 316 stainless steel (finger tight) 316 stainless steel and Viton® (internal o-ring) Diaphragm Seal Fluid: Silicone Oil Connection: Crystal CPF* Female | STORAGE TEMPERATURE Temperature Range: -40 to...
Open the catalog to page 3
Pump System* AMETEK offers a variety of solutions for pressure generation and measurement. Our line of products for pressure and a waterproof carrying case. ► Ordering a Pump System Only Any pump system, carrying case, and connection fittings for an XP2i pressure gauge may be ordered separately from the gauge. Enter XP2I-NONE followed by the Pump System part number and the Carrying Case option code. System B pump system with a waterproof carrying case. generation includes everything from small pneumatic hand pumps to a precision, hydraulic pressure comparator capable of generating up to 15...
Open the catalog to page 4
XP2i Pressure Gauge psi PUMP SYSTEMS OVERVIEW Pump System Case Options Part Number Pressure Range Hand Pump Included Pump Waterproof (Pelican Case) -27 inHg to 580 psi / - 0.91 to 40 bar System H
Open the catalog to page 5All AMETEK Calibration Instruments catalogs and technical brochures
-
PTC-125 Cooler
12 Pages
-
MTC 650A Spec Sheet
8 Pages
-
MTC 650A - Dry-block
8 Pages
-
PTC Series
13 Pages
-
mACal - mA Loop Calibrator
5 Pages
-
Crystal Pressure Fittings
13 Pages
-
CPF
4 Pages
-
AMETEK Calibration Brochure
12 Pages
-
Sanitary Sensors Brochure
4 Pages
-
30 series
4 Pages
-
JofraCloud
6 Pages
-
nVision Reference Recorder
9 Pages
-
DTI-1000 & STS-100/102
10 Pages
-
ETC Series - Dry-block
8 Pages
-
PTC-125 - Ultra Cooler
12 Pages
-
RTC-156/157 Dry-block calibrator
12 Pages
-
CTC Series - Dry-block
8 Pages
-
RTC-700 Dry-block calibrator
12 Pages
-
RTC-159 - Ultra Cooler
12 Pages
Archived catalogs
-
nVision Lab Brochure
7 Pages
-
DTI050 and STS-050
4 Pages
-
Type T - Deadweight tester
8 Pages
-
m1m Data Sheet
4 Pages
-
AMC910 - Signal calibrator
4 Pages
-
CSC101 - Signal Calibrator
4 Pages
-
CSC201 - Signal calibrator
4 Pages
-
HPC40 Series
7 Pages
-
ITC Series - Dry-block
8 Pages