Overview: The document is a specification sheet for the RTC-158 and RTC-250 Reference Temperature Calibrators by JOFRA, part of AMETEK. These calibrators are designed for high accuracy and stability in temperature calibration, offering both liquid bath and dry-block configurations.
Specifications:
  • Accuracy: High accuracy down to ±0.04°C using external reference sensors.
  • Stability: Excellent stability of 0.01°C.
  • Temperature Range: RTC-158 operates from -22 to 155°C, and RTC-250 from 28 to 250°C.
  • Homogeneity: Active dual-zone block ensures good temperature homogeneity.
  • Dynamic Load Compensation (DLC): Ensures temperature uniformity, available in B and C models.
  • Intelligent Sensors: Plug’n’play calibration system with intelligent reference sensors.
  • Communication: USB port for easy communication.
Features:
  • Multi-Model Options: Available in models A, B, and C, with varying features such as input for reference sensors and DLC sensors.
  • Display: Easy-to-read color VGA display with intuitive navigation.
  • Portability: Lightweight design with a functional carrying case.
  • Multi-Hole Inserts: Designed to accommodate various sensor sizes.
  • Integrated Support Rod: Lightweight and easy to mount, useful in bath calibration.
Advantages:
  • Liquid Bath vs. Dry-Block: Liquid baths allow calibration of sensors that do not fit into insertion tubes, while dry-blocks offer easier handling and no hazardous liquids.
  • Calibration Efficiency: Capable of calibrating up to 24 sensors simultaneously with the ASM multi-scanner.
  • Software Integration: JOFRACAL software supports on-line calibration and documentation.
Additional Features:
  • MVI: Mains power Variance Immunity for stable operation in fluctuating power conditions.
  • Auto-Stepping: Programmable temperature steps with results display.
  • Work Orders: Allows stand-alone operation and prevents unauthorized changes.
Hardware and Software Requirements:
  • PC Requirements: Minimum hardware includes Intel Pentium II 1.4 GHz processor, 128MB RAM, and USB port. Software requires Microsoft Windows ME, XP, or 7.
JOFRACAL Software: This versatile software supports calibration of various temperature and pressure sensors, integrating seamlessly with JOFRA instruments. It allows for manual data entry and storage for later analysis.
DLC System: The Dynamic Load Compensation (DLC) system enhances calibration accuracy by maintaining temperature homogeneity, even with varying sensor loads. It allows for simultaneous calibration of multiple sensors and compensates for sensor production tolerances.
Input Specifications: The calibrators support a range of input types, including RTD and thermocouple inputs, with specific accuracy levels detailed for each type.
Physical and Operational Details: The document outlines the physical dimensions, weight, power consumption, and environmental operating conditions for the RTC-158/250 models.
Accessories: Various inserts and a liquid bath kit are available for different calibration needs, with specific part numbers provided for ordering.
Model Variants: Three models are available: Model A (basic), Model B (full with DLC and reference sensor inputs), and Model C (middle with DLC and reference sensor inputs). All models feature dual-zone heating/cooling, USB communication, and programmable temperature settings.
Ordering Information: The RTC-158 series operates from -22 to 155°C, while the RTC-250 series operates from 28 to 250°C. Power supply options vary by region, and insert types include undrilled and multi-hole options. Calibration certificates are available in various forms, including traceable and accredited options.
AMETEK Overview: AMETEK offers a range of test and calibration instruments under various brands, including JOFRA for temperature calibrators, M&G for pressure testers, and Chatillon for force measurement. The company provides global support with contact information for offices in the UK, France, Germany, Denmark, USA, India, Singapore, and China.
See more

Catalog excerpts

RTC-158/250 Dry-block / Liquid Bath Calibrator-1

Reference Temperature Calibrator Down to ±0.04°C using the external reference sensor. 4-wire True-Ohm-Measurement technology is used » Excellent stability 0.01˚C » Wide temperature range » Excellent temperature homogeneity id bath for liqu use Both -block nd dry a Unique active dual-zone block ensures good temperature homogeneity in the calibration zone Patent Dynamic Load Compensation g! Perfect temperature uniformity in the pendin Liquid Bath insert, even when calibrating large sensors or many sensors at a time. B and C models only » Temperature uniformity indicator Shows the degree of temperature uniformity in the insert when using the new DLC technology. B and C models only In combination with the stirrer the newly developed sensor basket ensures virtually zero axial and radial gradients in the calibration zone » Intelligent reference sensors JOFRA reference sensors are supplied with intelligent plugs, holding the calibration data (coefficients) of the reference sensor. This is a truly plug’n’play calibration system AMETEK continues to develop new techniques to improve performance, accuracy, convenience and functionality of the well-known JOFRA calibration products. By doing so, we maintain our position as the leading worldwide manufacturer of temperature dry-block calibrators. Advantages of the combined liquid bath/dry-block calibrator Calibration of many sensors at a time due to more space for example in connection with validation of many thermocouples, which saves time • Automatic calibration of as many as 24 sensors at a time For customers, who only want to use liquid baths For calibration of odd sizes and shapes of sensors including sanitary sensors WET = no need for inserts, that fit the sensors DRY = more space for calibration of special sensors Industries who need to calibrate many sensors at a time or short sensors can benefit from the big well JOFRACAL software and RTC B-models can handle on-line calibration and documentation of multiple sensors calibrated simultaneously All RTC calibrators communicate via an easy-to-use USB port Best performing dry-block with regard to the EURAMET/cg-13/v.01 guideline for testing of dry-blocks ISO 9001 Manufacturer Specification Sheet, SS-RTC-158/250

 Open the catalog to page 1
RTC-158/250 Dry-block / Liquid Bath Calibrator-2

We are proud to introduce our new top model RTC (Reference Temperature Calibrator) series, which is even more sophisticated than any existing calibrators. The RTC offers many new fantastic features, such as: • Patent pending DLC, Dynamic Load Compensation system, for perfect temperature uniformity in the insert • Unique intelligent sensors for plug’n’play connection • USB connector for communication • Easy-to-read color VGA display with perfect overview of the actual status • Intuitive, fast and user-friendly navigation • Lightweight and easy to carry around • New functional carrying case design...

 Open the catalog to page 2
RTC-158/250 Dry-block / Liquid Bath Calibrator-3

These inserts have holes for both the DLC and two reference sensors (4 mm and 1/4”). These holes are marked for correct use. Intelligent reference sensors With this multi-hole insert in the carrying case the user is able to calibrate all the most commonly known sensor sizes. The JOFRA STS-200 intelligent reference sensor as well as the new DLC sensor are both containing all individual calibration data regarding the sensor. Firstly, this means that the time-consuming coefficient downloading sequence with risk of errors is no longer necessary. Secondly, the user can change the reference sensor...

 Open the catalog to page 3
RTC-158/250 Dry-block / Liquid Bath Calibrator-4

The new integrated support rod is part of the reduced weight philosophy. It is lightweight and very easy to mount on the RTC. Two fixing holes are integrated in the calibrator where the support rods can be mounted. The support rod is especially useful in bath calibration. Special designed carrying case AMETEK has designed an all-in-one carrying case. We have now made it possible to store both the STS reference sensors and the DLC sensor in the carrying case with an optimum physical protection. There is room for inserts, insulation plugs and calibration oil as well as compartments for the new...

 Open the catalog to page 4
RTC-158/250 Dry-block / Liquid Bath Calibrator-5

Instrument setups The RTC series allows the user to store up to ten (10) complete instrument setups. You may store all sorts of information including temperature units, stability criteria, use of external reference sensors, resolutions, sensors-under-test (SUT), conversions to temperature, display contrasts, etc. The setup may be recalled at any time. Maximum and minimum temperature From the setup menu, the user can select the maximum and minimum temperature limit for the calibrator. This function prevents damage to the sensor-under-test caused by excessive temperatures and it helps reducing...

 Open the catalog to page 5
RTC-158/250 Dry-block / Liquid Bath Calibrator-6

Unique temperature performance The RTC series of calibrators provides precision temperature calibration of sensors, whatever the type or format. This is accomplished through an innovative active dualzone heating technology. The RTC-series features our well-known active dual-zone heating technology. Each heating zone is independently controlled for precision temperature calibration. The homogeneity in the lower part is close to that of a laboratory liquid bath. The lower zone ensures optimum heat dissipation throughout the entire calibration zone. The upper zone compensates for heat loss from...

 Open the catalog to page 6
RTC-158/250 Dry-block / Liquid Bath Calibrator-7

INSERTS AND LIQUID BATH KIT Undrilled inserts for RTC-158/250 Inserts, undrilled Insert Code1 Undrilled insert Undrilled insert with DLC Undrilled insert with DLC and reference sensor 4 mm and 1/4 in Note 1: Use the insert code when ordered as the standard insert together with a new calibrator. Metric (mm) Multi-hole inserts for RTC-158/250 Spare part no. for multi-hole inserts - metric (mm) Instruments Insert code 1 M01 Note 1: Use the insert code when ordered as the standard insert together with a new calibrator. Imperial (Inch) Multi-hole inserts for RTC-158/250 Spare part no. for multi-hole...

 Open the catalog to page 7
RTC-158/250 Dry-block / Liquid Bath Calibrator-8

FUNCTIONAL SPECIFICATIONS INPUT SPECIFICATIONS Temperature range All input specifications apply to the dry-block of the calibrator running at the respective temperature (stable plus an additional 20 minute period). RTC-158 @ ambient @ ambient @ ambient RTC-250 @ ambient @ ambient @ ambient All input specifications are valid for RTC-158 and RTC-250. Accuracy with internal reference sensor 12-month period. Relative to reference standard. Specs by use of the external STS-200 reference sensor. Excl. sensor drift. Note 1: True ohm measurement is an effective method to eliminate errors from induced...