1. Catalogs
  2. AMETEK Calibration Instruments
  3. Jofra Insertion Tubes for Dry-Block Calibration

Jofra Insertion Tubes for Dry-Block Calibration

Jofra Insertion Tubes for Dry-Block Calibration

Jofra Insertion Tubes for Dry-Block Calibration

Product catalog summary
Temperature Specification Overview:
The document outlines the temperature ranges for various insertion tubes used in JOFRA dry-block calibrators. Key models include:
  • ITC-155: -23 to 155°C / -9 to 311°F
  • ITC-320: 33 to 320°C / 91 to 608°F
  • ITC-650: 33 to 650°C / 91 to 1202°F
  • CTC-140: -17 to 140°C / -1 to 284°F
  • CTC-320: 33 to 320°C / 91 to 608°F
  • CTC-650: 33 to 650°C / 91 to 1202°F
  • CTC-1200: 300 to 1205°C / 572 to 2200°F
  • MTC-140: -17 to 140°C / -1 to 284°F
  • MTC-320: 33 to 320°C / 91 to 608°F
  • MTC-650: 33 to 650°C / 91 to 1202°F
Usage Guidelines:
Recommendations for handling and maintaining the calibrators include:
  • Use the insert removal tool for safe removal.
  • Remove inserts post-calibration to prevent difficulty later.
  • Wear protective gear when handling extreme temperatures.
  • Ensure cleanliness of sensors and inserts for optimal thermal conductivity.
  • Regular cleaning of inserts and wells is advised to remove contaminants.
Precautions:
Important precautions include:
  • Avoid forcing inserts or sensors into the well.
  • Prevent foreign materials from entering the well.
  • Do not use oils or liquids to enhance conductivity due to leakage risks.
  • Use only specified metal inserts for compatibility.
Common Issues:
Incorrect insertion tube usage can lead to calibration inaccuracies, emphasizing the importance of using the correct tube.
Insert Compatibility and Materials:
Materials used for inserts vary by model:
  • CTC-140 A and CTC-320 A/B: Aluminum
  • CTC-650 A/B: Brass
  • CTC-1200 A: High-temperature steel alloy
Specifications refer to the sensor's outer diameter.
Insert Types and Codes:
Detailed tables provide spare part numbers for predrilled and undrilled inserts in both metric and imperial measurements for various models.
Overview: The document details technical specifications and ordering information for multi-hole inserts used in temperature calibrators, specifically models ATC-125/140/156/157 and ITC-155, and outlines AMETEK's product offerings.
Specifications: Lists multi-hole inserts with dimensions in metric and inches, designed for specific calibrator models, each with a matching insulation plug for optimal performance.
Ordering Information: Standard inserts require an insert code when ordering with a new calibrator. Custom inserts need specification of instrument and sensor size. Contact details for global ordering are provided.
Product Offerings: AMETEK provides a range of calibration instruments including:
  • Temperature Calibrators: Portable dry-block calibrators, precision thermometers, and liquid baths.
  • Pressure Calibrators: Electronic systems for field use, ranging from -25 mbar to 1000 bar.
  • Signal Instruments: Tools for process signal measurement and simulation.
  • M&G Pressure Testers & Pumps: High-accuracy pressure testers and generators.
  • Lloyd Instruments: Materials testing machines and software.
  • Davenport Polymer Test Equipment: Tools for measuring moisture-sensitive PET polymers.
  • Chatillon Force Measurement: Handheld force gauges and motorized testers.
  • Newage Testing Instruments: Hardness testers and related software.
Application Kit: Includes a kit for calibrating sanitary sensors using custom-made inserts with reference sensors.
Contact Information: Provides contact details for AMETEK offices in the UK, France, Germany, Denmark, USA, India, Singapore, and China for support and inquiries.
See more

Catalog excerpts

Jofra Insertion Tubes for Dry-Block Calibration-1

for temperature dry-blocks Insertion TubesInserts for JOFRA dry-blocks ITC-155 -23 to 155°C / -9 to 311°F From time to time we get information from end-users of JOFRA dry-block calibrators that something is wrong with the calibrator, as it is not able to meet the temperature specifications. This is often due to the fact that the customer is using a wrong insertion tube, although he is in good faith. Therefore, you can save a lot of time and irritations by using the right insertion tube. Does and dont's in the use of inserts Important things to REMEMBER when using inserts: • Remove inserts from the well with the insert removal tool. • Remove the inserts after calibration, as it might cause problems to remove them later on if they stay in the well for longer periods. • Take care when removing very hot or cold inserts and use adequate protective equipment if necessary. • Make certain that the sensors and inserts are free from any particles of dirt before inserting them into the well. Inserts are designed to fit perfectly into the well thus maximizing the thermal conductivity. If not the insert may be jammed. • Remove regularly build-up of dirt and dust from inserts and wells with a cloth or detergent. Important things to AVOID when using inserts: • Do not force inserts or sensors into the well. • Do not drop any foreign materials into the well. • Do not use oil or other liquids to optimise the thermal conductivity due to risk of leakage, which will destroy the other components. • Do not use inserts of metals dissimilar to that of the block, unless specified by the manufacturer as this may cause a different degree of expansion and contraction. ISO 9001 Manufacturer Specification Sheet AS-Insertion Tubes AMETEK TEST & CALIBRATION

 Open the catalog to page 1
Jofra Insertion Tubes for Dry-Block Calibration-2

USING INSERTS Using inserts from other JOFRA calibrators Insertion tube Well Calibrator Often customers believe that they can use inserts from other models of JOFRA temperature calibrators. Consequently, we get complaints about errrors in the temperature specifications. Even if the diameter is right, the length might not be. In this case the insertion tube is too tall. This gives a heat flow out of the top of the well and the temperature uniformity will change dramatically. Using inserts that are too small Insertion tube Well Calibrator Using insertion tubes with diameters that are too small...

 Open the catalog to page 2
Jofra Insertion Tubes for Dry-Block Calibration-3

PREDRILLED INSERTS FOR CTC AND MTC SERIES Inserts for CTC-140 A and CTC-320 A/B are made of aluminum. Inserts for CTC-650 A/B are made of brass. Inserts for CTC-1200 A are made of high-temperature steel alloy. All specifications on hole sizes are referring to the outer diameter (OD) of the sensor-under-test. The correct clearance size is applied in all predrilled inserts. Inserts, undrilled Spare part no. for predrilled inserts - metric (mm) Spare part no. for predrilled inserts - imperial (inch) Note 1: Use the insert code, when ordered as the standard insert together with a new calibrator....

 Open the catalog to page 3
Jofra Insertion Tubes for Dry-Block Calibration-4

JOFRAAI# calibration #1\ PREDRILLED INSERTS FOR ATC AND ITC SERIES - 4 MM REFERENCE HOLE JOFRA dry-block insert compatibility and materials: ATC-320 = ATC-650 = ITC-320 = ITC-650 (made of brass) ATC-125 = ATC-155 = ATC-156 (made of aluminum) ATC-157 = ITC-155 (made of aluminum) All specifications on hole sizes are referring to the outer diameter (OD) of the sensor-under-test. The correct clearance size is applied in all predrilled inserts. (ATC-157 A/B and ITC-155 A) Spare part no. for predrilled inserts with 4 mm reference hole Note: All inserts (metric and inches) are supplied with a hole for...

 Open the catalog to page 4
Jofra Insertion Tubes for Dry-Block Calibration-5

Spare part no. for predrilled inserts with 1/4 in (6.35 mm) reference hole Note: All inserts (metric and inches) are supplied with a hole for the 4 mm OD reference sensor. Note 1: Use the insert code, when ordered as the standard insert together with a new calibrator. Note 2: ATC-157 / ITC-155: 12 mm, 13 mm, 7/16 in and 1/2 in inserts are delivered without the 4 mm reference hole, but supplied with a matching insulation plug. Insertion Tubes

 Open the catalog to page 5
Jofra Insertion Tubes for Dry-Block Calibration-6

Spare part no. for multi-hole inserts - metric (mm) Note: All multi-hole inserts (metric and inches) for ATC-125/140/156/157 and ITC-155 are supplied with a matching insulation plug. Note: Remember to use matching insulation plugs. Note 1: Use the insert code, when ordered as the standard insert together with a new calibrator. Insertion Tubes

 Open the catalog to page 6
Jofra Insertion Tubes for Dry-Block Calibration-7

Multi-hole M05 (ATC-125 A/B) Multi-hole M06 (ATC-125 A/B) Multi-hole M02 (ATC-140/250 A/B) Multi-hole M04 (ATC-157 A/B / ITC-155 A) Multi-hole M09 (ATC-157 A/B / ITC-155 A) Spare part no. for multi-hole inserts - imperial (inch) Note: All multi-hole inserts (metric and inches) for ATC-125/140/156/157 and ITC-155 are supplied with a matching insulation plug. Note: Remember to use matching insulation plugs. Note 1: Use the insert code, when ordered as the standard insert together with a new calibrator. Insertion Tubes

 Open the catalog to page 7
Jofra Insertion Tubes for Dry-Block Calibration-8

ORDERING CUSTOM-MADE SPECIAL INSERTS Custom-made special inserts are made on request. When ordering custom made special inserts a few details have to be noticed: • Specify the instrument for which the inserts are ordered • Specify the sensor size for each hole Please contct us for custom-made inserts. APPLICATION KIT FOR CALIBRATION OF SANITARY SENSORS At picture 1 you see a custom made insert and our STS-102 A cable reference sensor placed in a JOFRA ATC-156 B dry-block calibrator. At picture 2, the sanitary sensor has been fitted into the insert and is ready for calibration. Note that the design...

 Open the catalog to page 8

All AMETEK Calibration Instruments catalogs and technical brochures

  1. PTC-125 Cooler

    12  Pages

  2. PTC Series

    13  Pages

  3. CPF

    4  Pages

  4. 30 series

    4  Pages

  5. JofraCloud

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.