1. Catalogs
  2. AMETEK Calibration Instruments
  3. HPC550Ex/552Ex - Handheld Pressure Calibrator

HPC550Ex/552Ex - Handheld Pressure Calibrator

HPC550Ex/552Ex - Handheld Pressure Calibrator

HPC550Ex/552Ex - Handheld Pressure Calibrator

Product catalog summary
Ex Safety Certifications: The HPC550Ex and HPC552Ex calibrators are certified for use in explosive atmospheres, adhering to ATEX and IECEx standards. They are engineered to prevent ignition from electrical, sparking, or heating effects.
Pressure Range and Accuracy: These calibrators cover a pressure range from 25 mbar to 700 bar (0.35 to 10,000 psi) with a high accuracy of ±0.025% of reading plus 0.01% full scale, and are temperature compensated for field use.
Features: They include intelligent features such as damping, leak test, % error calculation, min/max, and switch test, along with a high accuracy thermometer using a Pt100 sensor.
Design and Usability: Designed for hazardous areas in the oil, gas, and chemical industries, they feature a clear graphical display, intuitive menu, and capabilities for mA, switch, and temperature measurements.
Pressure Measurement: The calibrators support 15 engineering units and maintain high accuracy across various pressure ranges, including compound, gauge, absolute, and differential ranges.
Environmental and Physical Specifications: Operating within -10 to 45°C, they have an IP54 ingress protection rating and are powered by four 1.5 VDC alkaline batteries, offering over 35 hours of battery life.
Applications: Suitable for technicians in petrochemical plants, oil platforms, refineries, and other explosive environments.
Ordering Information: Detailed ordering information is provided for different models and pressure ranges to help users select the appropriate calibrator.
Pressure Calibrator Specifications: Various pressure calibrators are outlined, with ranges from -0.96 to 700 bar, including absolute, gauge, and compound pressures.
Calibration Certificate: A NIST traceable calibration certificate is available to ensure accuracy and reliability.
Accessories: Optional accessories include a soft case for transport and a Pt100 temperature sensor. The soft case is not suitable for hazardous areas and includes compartments for components.
AMETEK Test & Calibration Instruments: An overview of AMETEK's offerings, including temperature and pressure calibrators, signal instruments, and more, highlighting their applications and features.
Contact Information: Contact details for AMETEK offices in the UK, France, Germany, Denmark, USA, India, Singapore, and China are provided for customer support and inquiries.
See more

Catalog excerpts

HPC550Ex/552Ex - Handheld Pressure Calibrator-1

Handheld Pressure Calibrator HPC550Ex Family » Ex Safety Certifications ATEX: II 2 G Ex ia IIB T3 Gb (Ta=-10°C to + 45°C) KEMA 10 ATEX 1068 X Ex ia IIB T3 Gb (Ta=-10°C to +45°C) IECEx CSA 10.0013 Intrins ically safe » Wide pressure range » High accuracy » True field calibrator Fully temperature compensated » Single or dual sensor versions Chose between HPC550Ex single senror calibrator, or HPC552Ex dual sensor calibrator. A HPC552Ex might be an alternative to two separate calibrators, saving investment and cost of ownership » Full set of intelligent features Wide selection of useful functions and features. Damping, leak test, % error calculation, min/max, switch test, etc. » Thermometer High accuracy thermometer, Pt100 sensor. » Intrinsically Safety - Apparatus - IECEx and EU Ex standards JOFRA HPC550/552Ex are intended for use in explosively hazardous areas and designed to prevent ignition by electrical, sparking or heating effects. JOFRA HPC550/552Ex calibrators feature deadweight tester accuracy in a modern digital package. The HPC550/552Ex can be used in a very broad range of applications from simple tool type jobs to complex calibration jobs in custody transfer systems. Including potential hazardous areas in Oil, Gas and Chemical industries. The HPC550/552Ex series has been designed to meet high accuracy pressure calibration applications and faciliate your tasks. HPC550/552Ex offers features such as user configurable information display, 15 different pressure units, transmitter supply, mA input, % error calculation, voltage measurement, serial communication, and external pressure module capability. The accuracy of the HPC550/552Ex calibrators is specified in % of reading to ensure an even better accuracy and wider applicable pressure range. The HPC is temperature compensated from 0 to 45ºC/32 to 113ºF for on-site operation. It is a truly superior pressure calibrator for laboratory and field use, bringing laboratory accuracy into the field. The pressure calibrator of choice in potential explosive atmospheres. The intrinsically safe HPC550/552Ex calibrators complies with ATEX / IECEx certifications and are an extension of JOFRA´s well known pressure calibrator range. HPC550/552Ex meet the needs of technicians who install, maintain, calibrate and troubleshoot equipment in petro-chemical plants, on oil platforms, in refineries, governmental bodies, commercial organizations and any location subject to the risk of potential explosion hazardous atmospheres. ISO 9001 Manufacturer Specification Sheet SS-HPC550Ex

 Open the catalog to page 1
HPC550Ex/552Ex - Handheld Pressure Calibrator-2

Clear graphical display Large backlit graphical display shows current status and mode, 1, 2 or 3 measurement windows, to suit your applications and demands. Perform semi-automatic pressure switch tests, leak test, % error calculation, switch in HART resistor or other calibration tasks through the outstanding functions and features of the HPC550Ex. Intuitive menu system Combination of “soft keys” and cursor keypad ensures easy operation. Temperature measurement High accuracy temperature measurement port by use of he optional Pt100 sensor. A superb instrument for easy and convenient temperature...

 Open the catalog to page 2
HPC550Ex/552Ex - Handheld Pressure Calibrator-3

Media compatibility Pressure: differential ranges (HPC550Ex only) Engineering units (built-in) Nickel plated brass and 316 stainless steel Pressure connection Pressure overload Pressure accuracy ambient temp. Pressure accuracy ambient temp. F.S. (full scale) is the numerical value of the positive pressure range. Temperature effect -10 to 0°C / 14 to 32°F ±0.005% F.S./°C. Accuracy includes hysteresis, nonlinearity, repeatability and reference standard uncertainty, 1 Year typical long-term stability, operated inside the rated temperature span and pressure range. For optimal performance “zero” the...

 Open the catalog to page 3
HPC550Ex/552Ex - Handheld Pressure Calibrator-4

CONFIG LOWER MORE JOFRA HPC Ex features a custom set up display, 1, 2 or 3 lines / measurement chancels, easy to configure for the indi- vidual task

 Open the catalog to page 4
HPC550Ex/552Ex - Handheld Pressure Calibrator-5

This table shows the resolutions that can be obtained by the HPC calibrators throughout all engineering units. HPC550Ex/HPC552Ex Ranges and Overpressure Ratings BAR Range (bar) Proof pressure - maximum allowable pressure without a shift in calibration. Burst pressure - sensor damaged or destroyed, some risk of personal injury. Static pressure - differential units only. Maximum allowed common mode pressure between both points. Compound ranges - the data for the 1 bar range also applies to the -1 to +1 bar compound range, the data for the 2 bar range also applies to the -1 to +2 bar compound range....

 Open the catalog to page 5
HPC550Ex/552Ex - Handheld Pressure Calibrator-6

ORDERING INFORMATION Order number Model Pressure range 25MD 0 to 700 bar / 10,000 psi gauge Calibration certificate G NIST traceable calibration certificate Accessories (Optional) C Soft case (Not to be used in hazardous areas) Temperature sensor Pt100 Pressure system IND Calibrator only Sample order number HPC550, -0.82 to 35 bar, NIST traceable certificate, soft case, temperature sensor, calibrator only Handheld Pressure Calibrator • HPC550Ex/552Ex

 Open the catalog to page 6
HPC550Ex/552Ex - Handheld Pressure Calibrator-7

ORDERING INFORMATION Order number Pressure range (Port 1 - lowest pressure, Port 2 - highest pressure) 0 to 700 bar / 10,000 psi gauge Calibration certificate G NIST traceable calibration certificate Accessories (Optional) C Soft case (Not to be used in hazardous areas) Temperature sensor Pt100 Pressure system IND Calibrator only Sample order number HPC550Ex/552Ex • Handheld Pressure Calibrator HPC552, port 1: -0.82 to 35 bar, port 2: 0 to 350 bar, NIST traceable certificate, soft case, temperature sensor, calibrator only

 Open the catalog to page 7
HPC550Ex/552Ex - Handheld Pressure Calibrator-8

ACCESSORIES Useful soft case - Option C Not to be used in hazardous areas! The HPC550Ex can be delivered with a soft case as an option. The handy soft case is designed for protection during transport. The soft case has separate compartments for HPC550Ex (w/Velcro strap), test leads, test hoses, temperature probe, and JOFRA APM pressure module. A shoulder strap ensures convenient transportation when climbing ladders, etc. and the handy strap on the back makes it possible to hang the instrument on a valve, ladder etc. The manual and calibration documents fit into a pocket on the front of the soft...