O3-A4

O3-A4

O3-A4

Product catalog summary
Performance Specifications
  • Sensitivity: -200 to -400 nA/ppm at 100ppb O3
  • Response Time: Less than 15 seconds from zero to 100ppb
  • Zero Current: 10 to 30 nA in zero air at 20°C
  • Noise: ±2 standard deviations equivalent to 5ppb
  • Range: Performance warranty limit at 5ppm O3
  • Linearity: 200 to 500 ppb error at full scale
  • Overgas Limit: Stable response up to 10ppm
Lifetime Specifications
  • Zero Drift: 0 to 50 ppb equivalent change per year
  • Sensitivity Drift: -20% to -35% change per year
  • Operating Life: Greater than 18 months until 50% of original signal
Environmental Specifications
  • Temperature Sensitivity: Data provided for -20°C and 50°C at 500ppb O3
  • Zero Temperature Change: -10 to -30 nA at -20°C; 170 to 360 nA at 50°C
Cross Sensitivity
  • H2S: Less than -70% at 5ppm
  • NO2: 70% to 120% at 5ppm
  • Cl2: Less than 30% at 5ppm
  • NO: Less than -5% at 5ppm
  • SO2: Less than -4% initial transient at 5ppm
  • CO: Less than 0.1% at 5ppm
  • H2: Less than 2% at 100ppm
  • C2H4: Less than 0.1% at 100ppm
  • NH3: Less than 1% at 20ppm
  • CO2: Less than 0.1% at 5%
Key Specifications
  • Temperature Range: -20°C to +50°C
  • Pressure Range: 80 to 120 kPa
  • Humidity Range: 15% to 85% RH non-condensing
  • Flow Rate: Minimum 500 sccm during calibration
  • Bias Voltage: 0 V
  • Storage Period: 6 months at 3 to 20°C
  • Load Resistor: 33 to 100 Ω
  • Weight: Less than 6 g
Figures and Diagrams
  • Figure 1: O3-A4 Schematic Diagram
  • Figure 2: Sensitivity Temperature Dependence
  • Figure 3: Zero Temperature Dependence
  • Figure 4: Response to 200ppb Ozone
Additional Notes

All sensors are tested at ambient conditions with a 47 ohm load resistor unless otherwise stated. Customers should test under their own conditions to ensure suitability. For disposal instructions, contact Alphasense or its distributor.

See more

Catalog excerpts

O3-A4-1

O3-A4 Ozone Sensor Ozone 4-Electrode 4-Electrode Technical Specification Figure 1 O3-A4 Schematic Diagram Top View PERFORMANCE Sensitivity Response time Zero current Noise* Range Linearity Overgas limit Bottom View Side View nA/ppm at 100ppb O3 t90 (s) from zero to 100ppb nA in zero air at 20oC ±2 standard deviations (ppb equivalent) ppm O3 limit of performance warranty ppb error at full scale, linear at zero and 1ppm O3 maximum ppm for stable response to gas pulse * Tested with Alphasense AFE low noise circuit LIFETIME Zero drift Sensitivity drift Operating life ppb equivalent change/year in lab air % change/year in lab air, monthly test months until 50% original signal (12 month warranted) ENVIRONMENTAL Sensitivity @ -20°C Sensitivity @ 50°C Zero @ -20°C Zero @ 50°C CROSS SENSITIVITY H2S sensitivity NO2 sensitivity Cl2 sensitivity NO sensitivity SO2 sensitivity CO sensitivity H2 sensitivity C2H4 sensitivity NH3 sensitivity CO2 sensitivity % measured gas @ % measured gas @ % measured gas @ % measured gas @ % measured gas @ % measured gas @ % measured gas @ % measured gas @ % measured gas @ % measured gas @ KEY SPECIFICATIONS Temperature range Pressure range Humidity range Flow rate Bias voltage Storage period Load resistor Weight °C kPa % rh non-condensing minimum sccm during calibration V months @ 3 to 20°C (stored in sealed pot) Ω (AFE circuit is recommended) g Alphasense Ltd, Sensor Technology House, 300 Avenue West, Skyline 120, Great Notley. CM77 7AA. UK Telephone: +44 (0) 1376 556 700 Fax: +44 (0) 1376 335 899 E-mail: [email protected] Website: www.alphasense.com

 Open the catalog to page 1
O3-A4-2

Perf erfomance Data O3-A4 Perfomance Data Figure 2 shows the temperature dependence of sensitivity at 100ppb O3. This data is taken from a typical batch of sensors. Figure 3 Zero Temperature Dependence 400 350 300 Figure 3 shows the variation in zero output of the working electrode caused by changes in temperature, expressed as nA. This data is taken from a typical batch of sensors. Contact Alphasense for futher information on zero current correction. Figure 4 Response to 200ppb Ozone 420 410 Figure 4 shows response to 200ppb O3. Use of Alphasense AFE circuit reduces noise to 5ppb, with the opportunity...

 Open the catalog to page 2

All Alphasense catalogs and technical brochures

  1. SO2-B4

    2  Pages

  2. NO2-B43F

    2  Pages

  3. NO-B4

    2  Pages

  4. H2S-B4

    2  Pages

  5. CO-B4

    2  Pages

  6. PID-AR5

    2  Pages

  7. PID-A15

    2  Pages

  8. PID-AY5

    2  Pages

  9. PID-AH5

    2  Pages

  10. SO2-D4

    2  Pages

  11. PID-AH2

    2  Pages

  12. CH-D3

    2  Pages

  13. O2-G2

    2  Pages

  14. NO2-D4

    2  Pages

  15. NO-D4

    2  Pages

  16. IRM-AT

    2  Pages

  17. H2S-D4

    2  Pages

  18. PH3-BE

    2  Pages

  19. OPC-R2

    2  Pages

  20. CL2-D4

    2  Pages

  21. CO-D4X

    2  Pages

  22. SOH-A2

    4  Pages

Archived catalogs

  1. CO2-D1

    2  Pages

  2. SO2-B4

    2  Pages

  3. SO2-A4

    2  Pages

  4. PID-AH

    2  Pages

  5. PID-A1

    2  Pages

  6. O3-B4

    2  Pages

  7. NO2-B4

    2  Pages

  8. NO2-A4

    2  Pages

  9. NO-B4

    2  Pages

  10. NO-A4

    2  Pages

  11. H2S-B4

    2  Pages

  12. H2S-A4

    2  Pages

  13. CO-B4

    2  Pages

  14. CO-A4

    2  Pages

  15. COH-A2

    4  Pages

  16. Dual Sensor

    4  Pages

  17. CO-AX

    2  Pages

  18. SOH-A2

    4  Pages

  19. VOC Sensor

    2  Pages

  20. CO-AE

    2  Pages

  21. CO-AF

    2  Pages

  22. CO-BF

    2  Pages

  23. CO-BX

    2  Pages

  24. CO-CE

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.