NO-B4

NO-B4

NO-B4

Product catalog summary
Performance Specifications
Sensitivity ranges from 500 to 850 nA/ppm at 2ppm NO, with a response time of less than 45 seconds from zero to 2ppm NO. The zero current in zero air at 20°C is between 30 to 200 nA, and noise is ±2 standard deviations, equivalent to 15 ppb. The sensor's range is up to 20 ppm NO, with a linearity error of less than ±1 ppb at full scale. The overgas limit is 50 ppm for a stable response to a gas pulse.
Lifetime
The zero drift is 0 to 50 ppb equivalent change per year in lab air, and sensitivity drift is 0 to -20% change per year. The operating life is over 24 months until the signal reaches 50% of the original, with a 24-month warranty.
Environmental Specifications
Sensitivity at -20°C is 60 to 90% of the output at 20°C, and at 40°C, it is 97 to 110%. Zero current at -20°C is 0 to 30 nA, and at 40°C, it is 100 to 200 nA.
Cross Sensitivity
The sensor shows less than 10% sensitivity to H2S at 5ppm, less than 4% to NO2, less than 3% to Cl2, less than 5% to SO2, and less than 0.1% to H2, NH3, and CO2. Sensitivity to CO is less than 0.3%, and to O3 is less than 4%. Halothane sensitivity is less than 0.1% at 100ppm.
Key Specifications
The bias voltage is +200 mV, with an operating temperature range of -30 to 40°C, pressure range of 80 to 120 kPa, and humidity range of 15 to 85% rh. The storage period is 6 months at 3 to 20°C in a sealed pot. The recommended load resistor is 33 to 100 Ω, and the sensor weighs less than 13 g.
Figures and Diagrams
Figures illustrate the temperature dependence of sensitivity and zero output, as well as the response to 200ppb NO. The use of the Alphasense ISB circuit reduces noise to 15ppb, with potential for further digital smoothing.
Additional Notes
All sensors are tested under ambient conditions with a 10 ohm load resistor unless otherwise stated. Customers are advised to test sensors under their own conditions to ensure suitability. Disposal instructions are provided by Alphasense or its distributors.
See more

Catalog excerpts

NO-B4-1

NO-B4 Nitric Oxide Sensor Technical specifications Version 1.0 NO-B4 Nitric Oxide Sensor – 4-Electrode Figure 1 NO-B4 Schematic Diagram Bottom View Bottom View Side View Side View Dimensions are in millimetres PERFORMANCE (± 0.1 mm). Sensitivity nA/ppm at 2ppm NO nA/ppm in 2ppm 500 to 850 Performance Response Sensitivity 2ppm NO time t90 (s) from zero to NO Response time t90 (s) from zero to 2ppm NO < 45 Zero current current nA in zeroinair at 20°C Zero nA zero air at 20°C 30 to 200 Noise* ±2 standard deviations (ppb equivalent) Noise ± 2 standard deviations (ppb equivalent) 15 Range ppm NO limitperformance warranty of performance warranty 20 Range ppm NO limit of Linearity ppb error at full scale, linear at zero and 5ppm NO < 1 Linearity Overgas limit ppb error at full scale, linear at to gas pulse 5ppm± NO zero and maximum ppm for stable response 50 Overgas limit with Alphasense IBS low noise circuit maximum ppm for stable response to gas pulse Tested * * Tested with Alphasense ISB low noise circuit Zero drift Sensitivity drift ppb Operating life ppb equivalent change/year in lab air % change/year in lab air, monthly test equivalent change/year (24-month warranted) in lab air months until 50% original signal Zero drift Sensitivity drift % change/year in lab air, monthly test Operating life months % (output @ -20°C/output @signal 2ppmmonth warranted) until 50% original 20°C) @ (24 NO Sensitivity @ -20°C 60 to 90 Environmental ENVIRONMENTAL Zero @ -20°C nA 0 to 30 nA Sensitivity Zero-20°C (% output @ -20°C/output @ 20°C) @ 2ppm 100 to 200 @ @ 40°C NO Sensitivity @ 40°C (% output @ 50°C/output @ 20°C) @ 2ppm NO H S sensitivity % measured gas @ 5ppm H S (after 3 mins) < -10 Cross Sensitivity Zero @ -20°C nA NO sensitivity % measured gas @ 5ppm NO (after 3 mins) <4 Zero @ 40°C nA Cl sensitivity % measured gas @ 5ppm Cl <3 CROSS SENSITIVITY H sensitivity % measured gas @ 100ppm H < 0.1 H2S sensitivity % measured gas @5ppm CO 2S (after 3 minutes) 5ppm H CO sensitivity % measured gas @ < 0.3 NH sensitivity % measured gas @ 5ppm @ 5ppm NH NO2 sensitivity % measured gas NO2 (after 3 < 0.1 minutes) CO sensitivity % measured gas @ 5% volume CO < 0.1 Cl2 sensitivity % measured gas @100ppb 5ppm O Cl2 O sensitivity % measured gas @ <4 SO2 sensitivity % measured gas @100ppm 5ppm Halothane SO2 Halothane sensitivity @ < 0.1 H2 sensitivity % measured gas @ 100ppm H2 Bias voltage % measured gas @ 5ppm +200 mV (working electrode potential is above reference electrode) Key Specifications sensitivity CO CO Temperature range °C -30 to 50 NH3 sensitivity % measured gas @ 5ppm NH3 Pressure range kPa 80 to 120 CO2 sensitivity range% measured gas @ 5% Vol CO2 Humidity % rh continuous 15 to 85 O3 sensitivity period% measured@ 3 to 20°C (stored in sealed pot) 3 Storage months gas @ 100ppb O 6 Load resistor � (AFE circuit is recommended) Halothane sensitivity @ 100ppm Halothane 33 to 100 2 KEY SPECIFICATIONS Bias voltage mV (working electrode potential is above reference electrode) +20 Temperature range °C -30 to 4 For further information on the performance of this sensor, on other sensors in the range or any other T: +44 (0)1376 556700 E: [email protected] W: www.alphasense.com subject, please contact Alphasense Ltd. or visit our website at “www.alphasense.com

 Open the catalog to page 1
NO-B4-2

Technical specifications Version 1.0 Perf erformance Data NO-B4 Performance Data Figure 1 Sensitivity Temperature Dependence Figure 2 Sensitivity Temperature Dependence Perf erformance Data NO-B4 Performance Data 120 Figure 2 shows the temperature dependence of sensitivity at 2ppm NO. % Output (referenced to 20(referenced to(referenced to 20oC) % Output oC) % Output 20oC) Figure 2 Sensitivity Temperature Dependence NO-B4 Performance Data Perf erformance Data 100 120 This data2is taken from a Figure shows the typical batch of sensors. of Figure 1 shows the temperature dependence of temperature...

 Open the catalog to page 2

All Alphasense catalogs and technical brochures

  1. SO2-B4

    2  Pages

  2. NO2-B43F

    2  Pages

  3. H2S-B4

    2  Pages

  4. CO-B4

    2  Pages

  5. PID-AR5

    2  Pages

  6. PID-A15

    2  Pages

  7. PID-AY5

    2  Pages

  8. PID-AH5

    2  Pages

  9. SO2-D4

    2  Pages

  10. PID-AH2

    2  Pages

  11. CH-D3

    2  Pages

  12. O2-G2

    2  Pages

  13. NO2-D4

    2  Pages

  14. NO-D4

    2  Pages

  15. IRM-AT

    2  Pages

  16. H2S-D4

    2  Pages

  17. PH3-BE

    2  Pages

  18. OPC-R2

    2  Pages

  19. CL2-D4

    2  Pages

  20. CO-D4X

    2  Pages

  21. SOH-A2

    4  Pages

Archived catalogs

  1. CO2-D1

    2  Pages

  2. SO2-B4

    2  Pages

  3. SO2-A4

    2  Pages

  4. PID-AH

    2  Pages

  5. PID-A1

    2  Pages

  6. O3-B4

    2  Pages

  7. O3-A4

    2  Pages

  8. NO2-B4

    2  Pages

  9. NO2-A4

    2  Pages

  10. NO-B4

    2  Pages

  11. NO-A4

    2  Pages

  12. H2S-B4

    2  Pages

  13. H2S-A4

    2  Pages

  14. CO-B4

    2  Pages

  15. CO-A4

    2  Pages

  16. COH-A2

    4  Pages

  17. Dual Sensor

    4  Pages

  18. CO-AX

    2  Pages

  19. SOH-A2

    4  Pages

  20. VOC Sensor

    2  Pages

  21. CO-AE

    2  Pages

  22. CO-AF

    2  Pages

  23. CO-BF

    2  Pages

  24. CO-BX

    2  Pages

  25. CO-CE

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.