Performance Specifications
Sensitivity for NO detection ranges from 250 to 425 nA/ppm at 2ppm NO. The response time to reach 90% of the final value from zero to 2ppm NO is less than 25 seconds. Zero current in zero air at 20°C is between 10 to 80 nA, with noise levels at ±2 standard deviations equivalent to 80 ppb. The sensor's performance warranty covers up to 20 ppm NO, with linearity error at full scale being less than ±1 ppb. The overgas limit for stable response is 50 ppm.

Lifetime and Drift
The zero drift is 0 to 50 ppb equivalent change per year in lab air, while sensitivity drift is 0 to -20% change per year. The operating life is warranted for more than 24 months until the signal reduces to 50% of the original.

Environmental Conditions
Sensitivity at -20°C is 35 to 60% of the output at 20°C, and at 50°C, it is 120 to 135%. Zero current changes from 20°C are -10 to -50 nA at -20°C and 200 to 350 nA at 50°C.

Cross Sensitivity
The sensor shows less than 20% sensitivity to 5ppm H2S, less than 7% to 5ppm NO2, less than 4% to 5ppm Cl2 and SO2, and less than 0.1% to 100ppm H2, 5ppm NH3, and 5% Vol CO2. Halothane sensitivity at 100ppm is less than 0.1%.

Key Specifications
The bias voltage is +200 mV, with an operating temperature range of -30 to 50°C, pressure range of 80 to 120 kPa, and humidity range of 15 to 85% rh. The storage period is 6 months at 3 to 20°C in a sealed pot. The recommended load resistor is 33 to 100 Ω, and the sensor weighs less than 6 grams.

Additional Information
All sensors are tested under ambient conditions with a 10 ohm load resistor unless otherwise stated. Customers are advised to test sensors under their own conditions to ensure suitability. For disposal instructions, contact Alphasense or its distributor. For further information, visit the Alphasense website.
See more

Catalog excerpts

NO-A4-1

NO-A4 Nitric Oxide Sensor 4-Electrode 4-Electrode Technical Specification Figure 1 NO-A4 Schematic Diagram Top View PERFORMANCE Sensitivity Response time Zero current Noise* Range Linearity Overgas limit Bottom View Side View nA/ppm in 2ppm NO t90 (s) from zero to 2ppm NO nA in zero air at 20°C ±2 standard deviations (ppb equivalent) ppm NO limit of performance warranty ppb error at full scale, linear at zero and 5ppm NO maximum ppm for stable response to gas pulse * Tested with Alphasense AFE low noise circuit Zero drift Sensitivity drift Operating life ENVIRONMENTAL Sensitivity @ -20°C Sensitivity @ 50°C Zero @ -20°C Zero @ 50°C ppb equivalent change/year in lab air % change/year in lab air, monthly test months until 50% original signal (24 month warranted) (% output @ -20°C/output @ 20°C) (% output @ 50°C/output @ 20°C) nA change from 20°C nA change from 20°C CROSS SENSITIVITY % measured gas H2S sensitivity NO2 sensitivity % measured gas Cl2 sensitivity % measured gas SO2 sensitivity % measured gas H2 sensitivity % measured gas CO sensitivity % measured gas NH3 sensitivity % measured gas CO2 sensitivity % measured gas Halothane sensitivity KEY SPECIFICATIONS Bias voltage Temperature range Pressure range Humidity range Storage period Load resistor Weight mV (working electrode potential is above reference electrode) °C kPa % rh continuous months @ 3 to 20°C (stored in sealed pot) Ω (AFE circuit is recommended) g At the end of the product’s life, do not dispose of any electronic sensor, component or instrument in the domestic waste, but contact the instrument manufacturer, Alphasense or its distributor for disposal instructions. NOTE: all sensors are tested at ambient environmental conditions, with 10 ohm load resistor, unless otherwise stated. As applications of use are outside our control, the information provided is given without legal responsibility. Customers should test under their own conditions, to ensure that the sensors are suitable for their own requirements. Alphasense Ltd, Sensor Technology House, 300 Avenue West, Skyline 120, Great Notley. CM77 7AA. UK Telephone: +44 (0) 1376 556 700 Fax: +44 (0) 1376 335 899 E-mail: [email protected] Website: www.alphasense.com

 Open the catalog to page 1
NO-A4-2

Perf erformance Data NO-A4 Performance Data Figure 2 Sensitivity Temperature Dependence 140 Figure 2 shows the temperature dependence of sensitivity at 2ppm NO This data is taken from a typical batch of sensors. Figure 3 Zero Temperature Dependence 400 Figure 3 shows the variation in zero output of the working electrode caused by changes in temperature, expressed as nA. This data is taken from a typical batch of sensors. Contact Alphasense for futher information on zero current correction. Figure 4 shows response to 200ppb NO. This was tested with the ISB circuit. Technical Specification Time...

 Open the catalog to page 2

All Alphasense catalogs and technical brochures

  1. SO2-B4

    2  Pages

  2. NO2-B43F

    2  Pages

  3. NO-B4

    2  Pages

  4. H2S-B4

    2  Pages

  5. CO-B4

    2  Pages

  6. PID-AR5

    2  Pages

  7. PID-A15

    2  Pages

  8. PID-AY5

    2  Pages

  9. PID-AH5

    2  Pages

  10. SO2-D4

    2  Pages

  11. PID-AH2

    2  Pages

  12. CH-D3

    2  Pages

  13. O2-G2

    2  Pages

  14. NO2-D4

    2  Pages

  15. NO-D4

    2  Pages

  16. IRM-AT

    2  Pages

  17. H2S-D4

    2  Pages

  18. PH3-BE

    2  Pages

  19. OPC-R2

    2  Pages

  20. CL2-D4

    2  Pages

  21. CO-D4X

    2  Pages

  22. SOH-A2

    4  Pages

Archived catalogs

  1. CO2-D1

    2  Pages

  2. SO2-B4

    2  Pages

  3. SO2-A4

    2  Pages

  4. PID-AH

    2  Pages

  5. PID-A1

    2  Pages

  6. O3-B4

    2  Pages

  7. O3-A4

    2  Pages

  8. NO2-B4

    2  Pages

  9. NO2-A4

    2  Pages

  10. NO-B4

    2  Pages

  11. H2S-B4

    2  Pages

  12. H2S-A4

    2  Pages

  13. CO-B4

    2  Pages

  14. CO-A4

    2  Pages

  15. COH-A2

    4  Pages

  16. Dual Sensor

    4  Pages

  17. CO-AX

    2  Pages

  18. SOH-A2

    4  Pages

  19. VOC Sensor

    2  Pages

  20. CO-AE

    2  Pages

  21. CO-AF

    2  Pages

  22. CO-BF

    2  Pages

  23. CO-BX

    2  Pages

  24. CO-CE

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.