CO-D4X

CO-D4X

CO-D4X

Product catalog summary
Technical Specifications Overview:
The CO-D4X Carbon Monoxide Sensor is a miniature sensor designed for detecting carbon monoxide levels. It features a sensitivity range of 18 to 45 nA/ppm in 400ppm CO, with a response time of less than 25 seconds from zero to 400ppm CO at 22°C. The sensor has a zero current equivalent of less than ±4 ppm in zero air and a resolution of less than 1.5 ppm equivalent RMS noise. It supports a performance range up to 1,000 ppm with a linearity error of ±40 ppm CO at full scale.

Lifetime and Environmental Specifications:
The sensor exhibits a zero drift of less than 0.5 ppm equivalent change per year in lab air and a sensitivity drift of less than 6% change per year. It has an operating life of over 18 months until 80% of the original signal is maintained. The sensor operates effectively within a temperature range of -20°C to 50°C, a pressure range of 80 to 120 kPa, and a humidity range of 15 to 90% rh. It is recommended to store the sensor for up to 6 months at 3 to 20°C in a sealed pot.

Cross Sensitivity and Key Specifications:
The sensor has a filter capacity of 20,000 ppm·hrs for H2S and exhibits various sensitivities to other gases, including less than 0.1% for H2S at 20ppm, less than 6% for NO2 at 10ppm, and less than 60% for NO at 50ppm. The recommended load resistor is between 10 to 47 ohms, and the sensor weighs less than 2 grams.

Performance Data and Graphical Analysis:
Figures included in the document illustrate the sensor's sensitivity and zero output variations with temperature changes, as well as its response to increasing CO concentrations. The data is derived from typical sensor batches, showing mean values and ±95% confidence intervals.

Additional Notes:
All sensors are tested under ambient environmental conditions with a 10-ohm load resistor unless otherwise stated. The document emphasizes the importance of testing sensors under specific conditions to ensure suitability for particular applications. Disposal instructions are provided for the end of the product's life, advising against domestic waste disposal.
See more

Catalog excerpts

CO-D4X-1

Monoxide CO-D4X Carbon Monoxide Sensor Technical specifications Version 1.0 Miniature Size CO-D4X Carbon Monoxide Sensor – Miniature Size Figure 1 CO-D4X Schematic Diagram PATENTED and PATENT PENDING Bottom View Side View Bottom View Side View PERFORMANCE Sensitivity Response time Sensitivity Performance Zero current Response Resolution time Zero current RangeResolution Linearity Range Overgas limit Linearity nA/ppm in 400ppm CO 18 to 45 < 25 t90 (s) from zero to 400ppm CO % 22°C 18 to 45 ppm nA/ppm in 400ppm CO air equivalent in zero <±4 < 25 RMSt90 (s) from zero to equivalent) noise (ppm 400ppm CO % 22°C < 1.5 ppm equivalent in zero air <±4 ppm RMS noise (ppm equivalent) warranty limit of performance 1,000 < 1.5 ppm ppm limit of performance warranty CO error at full scale, linear at zero and1,000 400ppm CO ± 40 maximum ppmatfor scale, linear at zero and 400ppm CO ± 40 2,000 ppm CO error full stable response to gas pulse ppm equivalent change/year in lab air % change/year in lab air, monthly test ppm equivalent change/year in lab air < 0.5 months until 80%lab air, monthly test (18 month warranted) % change/year in original signal <6 Overgas limit Zero drift Sensitivitydrift Zero drift Operating life drift Sensitivity Operating life maximum ppm for stable response to gas pulse months until 80% original signal (18-month warranted) ENVIRONMENTAL Sensitivity @ -20°C % (output @ -20°C/output @ 20°C) @ 400ppm CO Environmental CROSS Cross Sensitivity SENSITIVITY Filter capacity Filter capacity H2S H S sensitivity sensitivity NO2 NO sensitivity sensitivity Cl2 Cl sensitivity sensitivity NO sensitivity NO sensitivity SO2 SO sensitivity sensitivity H sensitivity H2 sensitivity C H sensitivity C2H4 NH sensitivity sensitivity NH3 sensitivity 2 Sensitivity @ @ -20°C % (output @ 50°C/output20°C)20°C) @CO 50°C @ @ 400ppm 400ppm CO Sensitivity % (output @ -20°C/output @ N/A Zero @ -20°C@ 50°C ppm % (output @ 50°C/output @ from20°C CO equivalent change 20°C) @ 400ppm Sensitivity N/A Zero @ 50°C ppm ppm equivalent change from 20°C equivalent change from 20°C Zero @ -20°C <±2 ppm equivalent change from 20°C ppm·hrs H S ppm·hrs HS 2 % measured gas @ 20ppm H S 2S H % measured gas @ 20ppm % measured gas @ 10ppm NO 2 NO % measured gas @ 10ppm % measured gas @ 10ppm % measured gas @ 10ppm Cl Cl2 % measured gas @ 50ppm % measured gas @ 50ppm NO NO % measured gas @ 20ppm SO % measured gas @ 20ppm SO2 % measured gas @ 400ppm H % measured gas @ 400ppm C H 2 H % measured gas @ 400ppm % measured gas @ 400ppm NH 2H4 C % measured gas @ 20ppm % measured gas @ 20ppm NH3 2 Temperature range °C Pressure range OC kPa KEY Temperature range Humidity range % rh (see note below) SPECIFICATIONS Pressure range kPa months @ 3 to 20°C (stored in sealed pot) Storage period Humidity range %rh (see note below) Load resistor � (recommended) Weight g Storage period months @ 3 to 20OC (stored in sealed Load resistor Weight Note: Above 85% rh and 40OC a maximum continuous exposure period of 10 days is warranted. Where such exposure occurs the For at lower %rh and temperature sensor, on other sensors in days. T:sensor will [email protected] W: www.alphasense.com to rest further information on the performance of this levels for several the range or any other +44 (0)1376 556700 E: normal electrolyte volumes when allowed subject, please contact Alphasense Ltd. or visit our website at “www.alphasense.com”. At the end of the product’s life, do not dispose of any electronic sensor, component or inst

 Open the catalog to page 1
CO-D4X-2

Technical specifications Version 1.0 Perf erformance Data CO-D4X Performance Data Figure 1 Sensitivity Temperature Dependence Figure 2 Sensitivity Temperature Dependence CO-D4X Performance Data Perf erformance Data o o % Output (referenced to (referenced to 20oC) Zero current Zero current (equivalent ppm, referenced to 20 C) (equivalent ppm, referenced to 20oC) ppm, referenced% Output (referenced to 20oC) Output (nA) Zero current (equivalent to 20oC) % Output 20 C) Figure 2 Sensitivity Temperature Dependence 100 Perf erformance Data CO-D4X Performance Data 140 Mean "+95% conf." "-95% conf." Mean...

 Open the catalog to page 2

All Alphasense catalogs and technical brochures

  1. SO2-B4

    2  Pages

  2. NO2-B43F

    2  Pages

  3. NO-B4

    2  Pages

  4. H2S-B4

    2  Pages

  5. CO-B4

    2  Pages

  6. PID-AR5

    2  Pages

  7. PID-A15

    2  Pages

  8. PID-AY5

    2  Pages

  9. PID-AH5

    2  Pages

  10. SO2-D4

    2  Pages

  11. PID-AH2

    2  Pages

  12. CH-D3

    2  Pages

  13. O2-G2

    2  Pages

  14. NO2-D4

    2  Pages

  15. NO-D4

    2  Pages

  16. IRM-AT

    2  Pages

  17. H2S-D4

    2  Pages

  18. PH3-BE

    2  Pages

  19. OPC-R2

    2  Pages

  20. CL2-D4

    2  Pages

  21. SOH-A2

    4  Pages

Archived catalogs

  1. CO2-D1

    2  Pages

  2. SO2-B4

    2  Pages

  3. SO2-A4

    2  Pages

  4. PID-AH

    2  Pages

  5. PID-A1

    2  Pages

  6. O3-B4

    2  Pages

  7. O3-A4

    2  Pages

  8. NO2-B4

    2  Pages

  9. NO2-A4

    2  Pages

  10. NO-B4

    2  Pages

  11. NO-A4

    2  Pages

  12. H2S-B4

    2  Pages

  13. H2S-A4

    2  Pages

  14. CO-B4

    2  Pages

  15. CO-A4

    2  Pages

  16. COH-A2

    4  Pages

  17. Dual Sensor

    4  Pages

  18. CO-AX

    2  Pages

  19. SOH-A2

    4  Pages

  20. VOC Sensor

    2  Pages

  21. CO-AE

    2  Pages

  22. CO-AF

    2  Pages

  23. CO-BF

    2  Pages

  24. CO-BX

    2  Pages

  25. CO-CE

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.