Product catalog summary
Overview: The document describes the PDT Dewpoint Transmitter, a compact, economical solution for measuring dewpoints in gases and dry compressed air. It is particularly suitable for applications requiring readings down to -60 °C dewpoint, such as monitoring refrigerant dryers and cool stores.
Specifications: The transmitter is a 2-wire, 4-20 mA loop-powered device utilizing a bulk polymer sensing element. It offers various dewpoint ranges: -60 °C to +20 °C, -40 °C to +30 °C, and -20 °C to +50 °C. The device is designed for easy installation and minimal maintenance, with a robust zinc alloy body and IP66/NEMA 4 weatherproof protection.
Features and Benefits: The PDT Transmitter provides excellent accuracy (± 2 °C dewpoint), stability, and response speed. It is supplied with a calibration certificate traceable to national and international humidity standards. The device is compatible with DS4000, DS1200, and DIS030 display indicators for enhanced functionality.
Installation and Maintenance: The transmitter can be fitted directly into pipelines or used with a dedicated Dewpoint Sensor Holder/Sample Cell for optimal sampling conditions. This setup allows for easy removal and annual calibration without process interruption.
Technical Data: The transmitter operates within a temperature range of -20 °C to 70 °C and a pressure range of -1 to 50 bar. It is EMC compliant with BS EN ICE 61326-1 standards and features a 24-month warranty against defects.
Ordering Information: The document provides detailed instructions on ordering the transmitter and related components, including cable lengths and fitting options.
Precautions: The sensor should not be exposed to corrosive gases or strong oxidizing agents, as these can damage the sensor and impair calibration accuracy.
See more

Catalog excerpts

PDT Dew point Transmitter Data sheet -40+30°Cdp-1

Dewpoint Transmitter for General Applications Features and Benefits Long-lasting bulk Polymer sensing element offers excellent accuracy, stability and response speed IP66 / NEMA 4 Weatherproof Protection Compact 2-wire 4-20 mA Loop Powered Dewpoint Transmitter Supplied with a calibration certificate traceable to National and International Humidity standards Various dewpoint ranges available, -60 °C to +50 °C Economical solution for higher ‘wetter’ dewpoints, e.g. monitoring refrigerant driers DS4000 Hygrometer (Advanced) Optional Sensor Holder/Sample Cell for optimum sampling conditions The PDT Dewpoint Transmitter utilises a bulk Polymer sensing element. It is an economical choice for higher or ‘wetter’ dewpoints where your application only requires readings down to -60 °C dewpoint. Dewpoint ranges are offered as: -60 °C to +20 °C, -40 °C to +30 °C and -20 °C to +50 °C dewpoint. This compact, 2-wire, 4-20 mA Loop Powered Dewpoint Transmitter is used for trace moisture measurement in gases and dry compressed air. Applications include refrigerant dryers, cool stores, switchgear monitoring and more. Very easy to install, requiring little or no maintenance. Long-lasting bulk Polymer sensing element offers excellent sensitivity, repeatability, and response speed. Available separately, or combined with the DS4000 or DS1200 Display Indicators to make up a Hygrometer System. An alternative option is a combined system integrating the DIS030, a compact 4-20 mA Loop Powered Local Display that mounts directly on to the PDT Transmitter, as pictured above. The PDT can be fitted directly into the pipeline, or our dedicated Dewpoint Sensor Holder/Sample Cell. The Dewpoint Sensor Holder/Sample Cell is highly recommended for optimum sampling conditions and allows the Transmitter to be removed for annual laboratory dewpoint calibration without interrupting the process. PDT - Dewpoint Transmitter Technical Data Transmitter Type Sensing Element Output Signal Power supply Dewpoint Ranges Electromagnetic Compatibility (EMC) Accuracy Operating Temperature (Ambient and Process) Operating Humidity (Ambient) Pressure Range Factory Calibration Sample Flow Rate 2-wire, 4-20 mA Loop Powered Transmitter Polymer Linear 4 to 20 mA with selectable engineering units 24 Vdc (10 Vdc – 30 Vdc) See How to Order on page 3 Complies with BS EN ICE 61326-1 ± 2 °C dewpoint (UK National Physical Laboratory traceable for range -90 °C to +20 °C) -20 °C to 70 °C Maximum 95 %RH Non-Condensing -1 to 50 bar Supplied with a calibration certificate traceable to National and International Humidity standards. We recommend annual laboratory calibration. Flow in

 Open the catalog to page 1
PDT Dew point Transmitter Data sheet -40+30°Cdp-2

Electrical Connection Mechanical Connection Transmitter Body Sensor Protection Ingress Protection Manufacturer’s Warranty M12, 5-pole G1/2” Zinc alloy, PC, ABS 316 Sintered stainless steel filter – 50 micron IP 65 / NEMA 4 when Connector coupled to Transmitter 24 months in case of defective parts or faulty workmanship Measurement Units in millimeters / (inches) ASH-PDT Transmitter Holder DS4000 Hygrometer Display Technical Specifications See Datasheet - Document No. 3037 DS1200 Hygrometer Display Technical Specifications See Datasheet - Document No. 3039 DIS030 Local Display Technical Specifications...

 Open the catalog to page 2
PDT Dew point Transmitter Data sheet -40+30°Cdp-3

Transmitter Range: Example: PDT-60/20 = -60 °C to +20 °C dewpoint, or equivalent in °F For a longer cable length with connecting plug, simply enter the required length after the part number above up to 10 meters maximum: INLET Pipe Fitting The PDT Transmitter comes complete with 2 metres of connecting cable and a connector plug as standard. Transmitter Holder with Fittings • Comes compete with mounting plate. Example: CAB115 - 7 = 7 meters with connector plug. No Fittings 1/4” - Pipe OD 1/8” - Pipe OD 6 mm - Pipe OD OUTLET Pipe Fitting No Fittings 1/4” - Pipe OD 1/8” - Pipe OD 6 mm - Pipe OD...