1. Catalogs
  2. Alpha Moisture Systems Ltd
  3. Model DS1200 AMT-Ex Online Dewpoint Hygrometer

Model DS1200 AMT-Ex Online Dewpoint Hygrometer

Model DS1200 AMT-Ex Online Dewpoint Hygrometer

Model DS1200 AMT-Ex Online Dewpoint Hygrometer

Product catalog summary
Specifications:
  • Power Supply: 22Vac - 253Vac 50/60Hz & 20Vdc - 300Vdc
  • Output Signal: Linear 0-20/4-20 mA isolated retransmission signal
  • Alarm Relays: 2 programmable, rated 2A @ 250Vac / 1A @ 24Vdc
  • Power Consumption: 3.5 W Maximum
  • EMC Compliance: BS EN IEC 61326-1
  • Operating Temperature: -20 °C to +60 °C
  • Operating Humidity: Maximum 95 %RH Non-Condensing
  • Dimensions: Height 48 mm, Width 96 mm, Depth 120 mm
  • Ingress Protection: Front panel IP65 / NEMA 4X
  • Warranty: 24 months
Features and Benefits:
  • 4-digit display with user-configurable options for dewpoint in °C or °F, ppm(v), g/m³, or lb/MMSCF
  • Flameproof with ATEX and IECEx Certification for explosive atmospheres
  • Compatible with Alpha Moisture Systems Online Dewpoint Transmitters
  • Selectable 0-20 mA or 4-20 mA isolated retransmission signal
  • Two programmable alarm relays
Applications:
  • Designed for continuous trace moisture measurement in gases and dry compressed air
  • Suitable for safe area installations, with options for hazardous areas using AMT-Ex transmitter
Limitations:
  • Should not be exposed to corrosive gases such as mercury, ammonia, chlorine, or strong oxidizing agents like ozone
Related Products:
  • AMT-Ex for hazardous applications
  • Isolator for use with AMT-Ex
  • AMT and PDT for safe area use
Ordering Information:
  • Specify input range: [A] -4-20mA, [B] -0-20mA
See more

Catalog excerpts

Model DS1200 AMT-Ex Online Dewpoint Hygrometer-1

Dewpoint Hygrometer Features and Benefits User-configurable display of dewpoint in °C or °F dewpoint, ppm(v), g/m³ or lb/MMSCF, factory pre-configured to customer order Linear 0-20/4-20 mA isolated retransmission signal 2 programmable alarm relays Front panel protection to IP65 / NEMA 4X Exd option for installation in Ex Hazardous areas. Flameproof to ATEX and IECEx Certification for use in potentially explosive atmospheres For continuous trace moisture measurement in gases and dry compressed air and dry rooms/chambers. The DS1200 Dewpoint Hygrometer is installed in the safe area (non-hazardous), and is a panel mounted single-channel Indicator specifically designed to work with Alpha Moisture Systems Online Dewpoint Transmitters AMT and PDT transmitter (Safe area use), and the AMT-Ex transmitter (where the transmitter only is installed in the hazardous area). The DS1200 can be factory configured to read in °C dewpoint, °F dewpoint, ppm(v), ppb(v), g/m³ or lbs/MMSCF. The measured value is displayed on a large four-digit LED display. The Indicator offers a selectable 0-20 mA or 4-20 mA isolated retransmission signal and two programmable alarm relays, which can be independently set across the operating range of the Transmitter. The DS1200 is our entry-level Indicator, offering simplicity and functionality for most routine applications. Our fully-featured DS4000 will satisfy more advanced applications requiring a user-configurable display and greater versatility. Power Supply Output Signal Alarm Relays Power Consumption Electromagnetic Compatibility (EMC) Operating Temperature (Ambient) Operating Humidity (Ambient) Cut-out Dimensions Weight Dimensions Ingress Protection Manufacturer’s Warranty 22Vac - 253Vac 50/60Hz & 20Vdc -300Vdc Linear 0-20/4-20 mA isolated retransmission signal 2 programmable alarm relays. Contacts rated 2A @ 250Vac / 1A @ 24Vdc. 3.5 W Maximum Complies with BS EN IEC 61326-1 -20 °C to +60 °C Maximum 95 %RH Non-Condensing 44.5 mm x 91.5 mm 230 g Height 48 mm, Width 96 mm, Depth 120 mm Front panel only IP65 / NEMA 4X 24 months in case of defective parts

 Open the catalog to page 1
Model DS1200 AMT-Ex Online Dewpoint Hygrometer-2

Specify input range: [A] -4-20mA [B] -0-20mA Related Products The Dewpoint Transmitters and Isolator below are fully compatible with the DS1200 Dewpoint Hygrometer AMT-Ex For Hazardous Applications Isolator Used with AMT-Ex. Isolator and DS1200 in the safe area AMT Used with DS1200 in the Safe area PDT Used with DS1200 in the Safe area Corrosive Gases: The Sensor should not be exposed to corrosive gases (or corrosive contaminants in the gas sample) as these can chemically attack the sensor, impairing calibration accuracy and/or damaging it beyond economic repair. Examples of such gases are mercury...

 Open the catalog to page 2

All Alpha Moisture Systems Ltd catalogs and technical brochures

  1. AMT User Manual

    14  Pages

  2. PSS User Manual

    17  Pages

  3. DS1500-Exd

    2  Pages

  4. Model 6020

    3  Pages

  5. DS1500

    3  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.