Overview: The CMH-RPC Precision Chilled Mirror Dewpoint Hygrometer is a high-accuracy instrument designed for laboratory and precision applications. It measures dewpoint temperatures from -95 °C to +20 °C with an accuracy of ±0.2 °C, making it ideal for scientific and research purposes.
Key Features:
  • 3-stage chilled mirror with integral cooling for precise dewpoint measurement.
  • Backlit LCD for displaying readings in °C or °F dewpoint and ppm(V).
  • Optional temperature and pressure sensors available.
  • RS232 serial communication for data output and remote programming.
  • Automatic Balance Control (ABC) for easy measurement and control.
  • Supplied with a calibration certificate traceable to national and international standards.
Technical Specifications:
  • Sensing Element: Three-stage with additional refrigeration.
  • Analogue Output: 0-5 Vdc, 4-20 mA, 0-10 Vdc, or 0-20 mA.
  • Power Supply: 110 Vac or 230 Vac, 50/60 Hz.
  • Sample Flow Rate: 0.25 to 2.4 L/min.
  • Sample Pressure: 0 to 5 bar.
  • Weight: 47 kg; Dimensions: Height 600 mm, Width 520 mm, Depth 435 mm.
Additional Information: The device includes features such as dual alarm capability, manual and programmable automatic balance control, and compliance with BS EN ICE 61326-1 for electromagnetic compatibility. It is supplied with a 24-month warranty for defective parts or workmanship.
Contact Information: For pricing, delivery, or further details, contact Alpha Moisture Systems Ltd. at +44 (0) 1274 733100 or email [email protected]. More information on engineered solutions is available on their website.
See more

Catalog excerpts

Model CMH-RPC Chilled Mirror Dewpoint Hygrometer-1

Precision Chilled-Mirror Dewpoint Hygrometer Features and Benefits A primary measurement instrument, ideal for laboratory and precision applications 3 Stage Chilled Mirror Hygrometer, with integral cooling Precise measurement to -95 °C frost/dew point, 37 ppb(V) High accuracy instrument (±0.2 °C dewpoint) for laboratory or similar applications Backlit LCD displaying readings in °C or °F dewpoint and ppm(V) Optional temperature and pressure sensors are available RS232 Serial comms Pressure measurement can be displayed using the optional pressure transducer 2 programmable alarm relays Automatic Balance Control (ABC), for fast and easy measurement and control Supplied with a Calibration Certificate traceable to National and International Humidity standards CMH-RPC Precision Chilled Mirror Dewpoint Hygrometer accurately measures the dewpoint of a gas by cooling a mirrored surface, over which the gas flows until a layer of dew forms on the surface. Bench or Counter Top Standing - Front panel access to all main components including mirror cleaning - Fast Response Stable - Reliable - Quick Installation. The Chilled Mirror Dew Point Hygrometer utilises a 3-stage cooler combined with an integral refrigeration unit to measure the dewpoint in gases from below -95 °C to +20 °C (-139 °F to +68 °F) dewpoint with an accuracy of ±0.2 °C dewpoint. This primary measurement principle makes the CMH-RPC the ideal instrument for laboratory, scientific, and research applications and a reference instrument for checking and dewpoint calibration of Dewpoint Meters, Dewpoint Transmitters, Dew Meters and Dewpoint Analysers. Please do not hesitate to contact us for price, delivery, or details of your local distributor. Sensing Element Analogue output Serial Output Power Supply Dewpoint Range Electromagnetic Compatibility (EMC) Accuracy Manual Balance Control (MBC) Programmable Automatic Balance Cycle (PABC) Dual Alarm Capability Three Stage With Additional Refrigeration 0-5 Vdc, 4-20 mA, 0-10 Vdc, or 0-20 mA User Scalable to Drive Peripheral Devices Available Simultaneously, One Output per Measurement Parameter Scalable Throughout the Entire Operating Range of Psychometric Variables Programme via Front Panel RS232 Communication With Terminal, Printer, or Computer Remote Programming Data Output Such as Dew Point and Alarm Status 110 Vac or 230 Vac, 50/60 Hz (Specified at Order) -95 °C Dewpoint to +20 °C Dewpoint at 22 °C Ambient Temperature Complies with BS EN ICE 61326-1 +/-0.2 °C (0.36 °F) dewpoint Automatically Corrects for Most Mirror Contaminants Automatically Re-Balances the Optics at Switch On Programmable Interval of Balance Control for Greatest Convenience Increases the Length of Unattended Operation User Programmable Can be set for latching

 Open the catalog to page 1
Model CMH-RPC Chilled Mirror Dewpoint Hygrometer-2

Pressure transducer input for psychometric variable Temperature sensor for RH calculation Factory Calibration Sample Flow Rate Sample Pressure Weight Dimensions Manufacturer’s Warranty Optional Extra Optional Extra Supplied with a calibration certificate traceable to National and International Humidity standards. We recommend annual laboratory calibration 0.25 to 2.4 L/min (0.5/5.0 SCFH) 0 to 5 bar (0 to 75 psia) 47 kg Height 600 mm, Width 520 mm, Depth 435 mm 24 months in case of defective parts or faulty workmanship Component Location CMH-RPC Control Panel On/Off Switch LCD Graphic Data Display...

 Open the catalog to page 2

All Alpha Moisture Systems Ltd catalogs and technical brochures

  1. AMT User Manual

    14  Pages

  2. PSS User Manual

    17  Pages

  3. DS1500-Exd

    2  Pages

  4. Model 6020

    3  Pages

  5. DS1500

    3  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.