DS1200-Exd-AMT-Ex
3Pages

{{requestButtons}}

Catalog excerpts

DS1200-Exd-AMT-Ex - 1

Flame Proof Single Channel Dew Point Hygrometer for Ex Hazardous Areas DS1200 Exd AMT Ex is a single channel, online hygrometer with isolator designed to accurately measure trace moisture in process gases u lising a range of dedicated, ultra high capacitance, field mountable transmi ers. The DS1200 Exd AMT Ex is designed to take dewpoint readings in Ex Hazardous areas Zones 1, 21, 2 and 22. It is Flameproof to ATEX and IECEx Cer fica on for use in poten ally explosive atmospheres. ATEX and IECEx Certified Enclosure. IP66 Rating. The system incorporates an automa c calibra on (AutoCal) facility and encompasses measurement needs from 120°C to +20°C ( 184°F to +68°F) dewpoint. The sensor/transmi er is highly sensi ve and has been designed and condi oned for long term stability. Addi onally, the transmi er is ATEX cer fied for intrinsic safety so can be used in hazardous environments with the isolator located inside the DS1200 Exd Flameproof Unit. There are three Hot Key front panel key func ons to gain rapid access to set alarms trip points and choice of displayed engineering units. These func ons can be locked to prevent unauthorised tampering. Side View Users may choose to view the trace moisture levels in °C dewpoint, °F dewpoint, ppm(v), ppb(v), g/m³ or lb/MMSCF on the 4 digit, 14 segment LED display. The factory calibrated, but fully configurable, 4 20mA output is automatically linearly correlated with all the selected engineering units. DS1200 Exd AMT Ex contains two independent, fully user programmable alarm contacts to provide all the versa lity required for any applica on. Choice of transmi ers covering an overall range from 120°C to +20°C dewpoint ( 184°F to +68°F) Units selectable for °C Td, °F Td, ppm(v), ppb(v), g/m³ or lb/MMSCF Calibra ons traceable to Interna onal Humidity Standards Automa c calibra on (AutoCal) adjustment to maintain accuracy between laboratory calibra ons GUBH Flameproof Enclosure (type “d”) Ex II 2 G D DS1200 Display has a flexible power input of 22V 253V AC or DC Opera on 50/60Hz Two independent fully programmable alarms relays Linear 4 20mA retransmission signal Hot Key func ons for rapid access to set alarms and units, can be locked to prevent tamper Ex cer fied sensor/transmi er for Hazardous Area use (via isolator housed in Exd encl

Open the catalog to page 1
DS1200-Exd-AMT-Ex - 2

Alpha Moisture Systems Model DS1200-Exd MODEL AMT-Ex Dewpoint Transmitter only TRANSMITTER TYPE: 3 Wire, 4-20mA Loop Powered Transmitter OUTPUT SIGNAL: 4 to 20mA Linear OPERATING VOLTAGE: 12-28V dc. Reverse polarity protected. SENSING ELEMENT: Ultra High Capacitance - Aluminium Oxide Type RANGES: See How to Order Step 1 Below INTRINSIC SAFETY CERTIFICATION STANDARDS: EN 60079-0 Ex II 1GD Ex ia IIC T4 Ex iaD 20 T135°C (-20°C ≤ Ta ≤ +60°C) AUTOCAL: Field calibration / Span check facility. Activated and operated via two button switches on the Transmitter body. Locked in normal operating mode...

Open the catalog to page 2
DS1200-Exd-AMT-Ex - 3

Alpha Moisture Systems Model DS1200-Exd Select a Range Below Holder Logging SR: -110°C to -20°C (-166°F to -4°F) dewpoint PL: -100°C to 0°C (-148°F to +32°F) dewpoint RD: -80°C to -20°C (-112°F to -4°F) dewpoint GY: -80°C to 0°C (-112°F to +32°F) dewpoint SD: -100°C to +20°C (-148°F to +68°F) dewpoint LD: -120°C to +20°C (-184°F to +68°F) dewpoint HD: -65°C to +20°C (-85°F to +68°F) dewpoint Select a Display Unit °C Dewpoint °F Dewpoint ppm(V) lb/MMSCF g/m3 Voltage Output Current Output Select a Retransmission Output Select Alarm 1 Configuration High Alarm, Direct action (normally...

Open the catalog to page 3

All Alpha Moisture Systems Ltd catalogs and technical brochures

  1. DS1500-Exd

    2 Pages

  2. Model 6020

    3 Pages

  3. DS1500

    3 Pages