1. Catalogs
  2. Alicat Scientific
  3. Technical Data for CODA-Series Mass Flow Controllers

Technical Data for CODA-Series Mass Flow Controllers

Technical Data for CODA-Series Mass Flow Controllers

Technical Data for CODA-Series Mass Flow Controllers

Product catalog summary
Technical Data Overview for CODA-Series Mass Flow Controllers
Specifications: The CODA-Series Mass Flow Controllers are designed to handle flow rates from 40 to 100,000 grams per hour. They operate within an ambient temperature range of 0–60°C and a fluid temperature range of −35–70°C. The devices offer ingress protection of IP40 or IP67 and are constructed with 316L stainless steel, FKM, and FFKM materials, with optional EPDM or PCTFE.
Sensor and Control Performance: The mass flow accuracy for liquids is ±0.2% of reading or ±0.05% of full scale, while for gases it is ±0.5% of reading or ±0.05% of full scale. The repeatability is ±0.05% of reading or ±0.025% of full scale. The steady state control range is 2%–100% of full scale, with a density accuracy of ±5 kg/m³ and a measurable density range of 100–2,000 kg/m³. The viscosity range is 0–200 cP.
Temperature Sensitivity: The mass flow zero shift is ±0.01% of full scale per °C from tare temperature, and the mass flow span shift is ±0.005% of reading per °C from 25°C.
Valve and Response Times: The valve function is normally closed. Typical control response times are <500 ms for 40 g/h–10,000 g/h and <800 ms for 30,000–100,000 g/h. Indication response times are <100 ms for 40 g/h–10,000 g/h and <200 ms for 30,000 g/h–100,000 g/h.
Communications: The controllers support analog I/O options of 0–5 Vdc, 0–10 Vdc, and 4–20 mA, as well as digital I/O options including Serial (USB-C), RS-232, RS-485, Modbus RTU, EtherCAT, and EtherNet/IP. Power requirements vary with flow rates, with 4 W for 40–10,000 g/h and 5 W for 30,000–100,000 g/h, both at 9–30 Vdc.
Range-Specific Technical Data: The document provides detailed specifications for process connections, recommended inlet filters, nominal pressure drops, and proof pressures for different flow ranges. For example, a 40 g/h flow uses a 1/4″ VCR®-compatible male connection with a 2μ filter and ≥6 PSID nominal pressure drop.
Dimensions and Weight: The dimensions and weight vary by flow range. For 40–10,000 g/h, the dimensions are approximately 4.318 in height, 5.138 in width, and 1.122 in depth, with a weight of about 2.0 lb. For 30,000–100,000 g/h, the dimensions are approximately 5.304 in height, 5.945 in width, and 1.575 in depth, with a weight of about 3.0 lb.
See more

Catalog excerpts

Technical Data for CODA-Series Mass Flow Controllers-1

Technical Data for CODA-Series Mass Flow Controllers 40 to 100,000 grams per hour full scale Standard specifications. Consult Alicat for available options. SENSOR AND CONTROL PERFORMANCE Liquid: ±0.2% of reading or ±0.05% of full scale, whichever is greater Gas: ±0.5% of reading or ±0.05% of full scale, whichever is greater Mass Flow Accuracy Repeatability ±0.05% of reading or ±0.025% of full scale, whichever is greater Steady State Control Range Density Accuracy Measurable Density Range Viscosity Range Zero Stability ±0.05% of full scale (included in mass flow accuracy) Mass flow zero shift: ±0.01% of full scale per °C from tare temperature2 Mass flow span shift: ±0.005% of reading per °C from 25°C Temperature Sensitivity Valve Function Typical Control Response Time Typical Indication Response Time 1 Density reading and density accuracy are independent of the mass flow reading and mass flow accuracy. 2 Mass flow zero shift for 40 g/h is ±0.025% of full scale per °C from tare temperature. MECHANICAL Ambient: 0–60°C Fluid: −35–70°C Consult Alicat for additional options Operating Temperature Range Ingress Protection Wetted Materials 316L stainless steel, FKM & FFKM standard; EPDM or PCTFE optional Consult Alicat for additional wetted materials options COMMUNICATIONS Analog I/O Options Serial (USB-C); RS-232 or RS-485 (DB-15 or M12) Modbus RTU, EtherCAT, EtherNet/IP Power Requirements Digital Update Rate Analog Update Rate RANGE-SPECIFIC TECHNICAL DATA Process connections Recommended inlet filter Full scale flow (g/h) Nominal pressure drop (H₂0) Proof Pressure (PSIA)³ Mounting Options

 Open the catalog to page 1
Technical Data for CODA-Series Mass Flow Controllers-2

Technical Data for CODA-Series Mass Flow Controllers 40 to 100,000 grams per hour full scale Standard specifications. Consult Alicat for available options. Tare Button Power Light Outlet Port Tapped Mounting Holes DIMENSIONS Full Scale Flow 40–10,000 g/h 30,000– 100,000 g/h

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.