1. Catalogs
  2. Agilent Technologies - Life Sciences and Chemical
  3. Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory

Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory

Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory

Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory

Product catalog summary
Introduction
Flame Atomic Absorption Spectrometry (FAAS) requires precise sample preparation, which can be time-consuming and error-prone. The SIPS 10 and SIPS 20 Sample Introduction Pump Systems are accessories for Agilent’s SpectrAA FAAS instruments that automate sample preparation, enhancing productivity and reducing costs.

Features of SIPS
The SIPS systems automate several tasks:
  • Preparation of up to 10 calibration standards from a single bulk standard.
  • Intelligent dilution of over-range samples.
  • Preparation of standard additions calibration from a single standard.
  • Introduction of analytical spikes using a single standard.
  • Addition of chemical modifiers.
  • Extension of the dynamic range of FAAS.
The SIPS 10 offers in-line calibration and sample dilution, while the SIPS 20 adds capabilities for in-line spiking and sample modification.

Operation of SIPS
The SIPS uses a peristaltic pump system that ensures solution flow is proportional to pump speed, allowing accurate control of solution volume. This design extends the pump tubing's lifetime by releasing pressure at the end of a run.

Automatic Calibration and Dilution
The SIPS systems prepare calibration standards from a single multi-element bulk standard, reducing preparation time and error risk. The SpectrAA software automatically determines standard concentrations. The SIPS also automatically dilutes samples outside the calibrated range, ensuring result integrity and eliminating costly re-analysis.

Performance and Accuracy
Table 1 and Figure 4 demonstrate the SIPS system's dilution accuracy, with a maximum dilution error of 1.2%.

Modifier Addition and In-line Spiking
The SIPS 20 automates the addition of chemical modifiers and in-line spiking of samples, reducing preparation time and contamination risk.

In-line Standard Additions Calibration
The SIPS 20 automates the addition of standards during analysis, ensuring calibration within the linear range for maximum accuracy.

Conclusion
The SIPS systems significantly enhance the efficiency and accuracy of FAAS by automating sample preparation and calibration processes, making them valuable tools for laboratories using Agilent’s SpectrAA instruments.
See more

Catalog excerpts

Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory-1

• Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory Technical Overview Flame Atomic Absorption Spectrometry (FAAS) typically involves element-specific sample preparation which can be time consuming and prone to operator-introduced errors. The SIPS 10 and SIPS 20 Sample Introduction Pump System (SIPS) are optional accessories for Agilent's SpectrAA FAAS instruments. They offer a range of powerful features that automate many tedious and error prone sample preparation tasks. These greatly simplify and speed up the sample preparation process, improving productivity and reducing costs. The combination of the SIPS and Agilent's Fast Sequential capabilities results in an easy to use, highly productive and cost effective Atomic Absorption solution.

 Open the catalog to page 1
Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory-2

Whether you are using an autosampler, or doing manual sample introduction, a SIPS accessory, combined with the SpectrAA software, will automatically: • Prepare up to 10 calibration standards from a single bulk standard, automating the instrument calibration process • Intelligently dilute over range samples • Prepare standard additions calibration from a single standard • Introduce analytical spikes, using a single standard • Add chemical modifiers, e.g. ionization suppressants • Extend the dynamic range of FAAS to that approaching ICP-OES The SIPS accessory is available in two versions: • The...

 Open the catalog to page 2
Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory-3

The SIPS 20 comprises two pumps (Figure 2). One delivers the sample and the second delivers a standard or modifier to the nebulizer. Controlling the relative speeds of both pumps allows the amount of standard or modifier added to the sample to be precisely regulated. This allows in-line 'spiking' of samples. No More Tedious Standard preparation The SIPS 10 and 20 are capable of preparing calibration standards from a single multi-element bulk standard. This reduces sample preparation time and the risk of operator error during standard preparation, which can lead to costly re-runs. Using the concentration...

 Open the catalog to page 3
Increasing Productivity and Simplifying Sample Preparation with the SIPS accessory-4

Modifier addition Flame AA methods routinely require the addition of chemical modifiers for accurate results. These modifiers are normally added during the sample preparation process. Using the second pump of the SIPS 20, modifiers can be automatically added during the analysis, taking a repetitive step out of the sample preparation process. The amount of modifier added is controlled by the pump speed, and this pump speed remains constant during sample dilutions to ensure consistent blank correction and matrix modification; something that must be accounted for when adding modifiers manually. In-line...

 Open the catalog to page 4

All Agilent Technologies - Life Sciences and Chemical catalogs and technical brochures

  1. MS500

    12  Pages

  2. 4200 MP-AES

    12  Pages

  3. Agilent B1507A

    22  Pages

  4. Agilent M9703A

    16  Pages

  5. I/O Hardware

    16  Pages

  6. VEE 9.32

    24  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.