Agilent U1450A/U1460A Insulation Resistance Testers Data Sheet Accomplish more in a day’s work with the Agilent U1450A/U1460A series insulation resistance testers. With a single tool, get remote testing capability and the fastest way to generate accurate test reports letting you carry out tests with full confidence. • Remote testing coupled with report generation capability (Windows PC, iOS/Android) • 50 V/100 V/250 V/500 V/1000 V test voltages • Adjustable test voltages from 10 V up to 1.1 kV range ** • Up to 260 GΩ insulation resistance range • Full-featured 4.5 digit OLED DMM (66,000 counts) * • Timed/PI/DAR test • IP 67 certified and drop proof up to 3 meters (10 feet) • (–40 to +55 °C) operating temperature range ** • CAT III 1000 V/CAT IV 600 V safety ratings * ** Do MORE. Accomplish MORE. Expect MORE.

Open the catalog to page 1
Agilent U1450A/U1460A - 2

Test remotely and eliminate MORE reporting errors Increase test efficiency and eliminate more reporting errors with the U1450A/U1460A Series' complimentary report generating software. 1. Agilent Handheld Meter Logger Software (Windows PC)1 2. Agilent Insulation Tester Application (iOS/Android)2 Eliminate the conventional data entering process and get error-free automated test reports generated in a tabulated or graphical form for easy interpretation and analysis - essential for troubleshooting, commissioning and preventive maintenance tasks. Better yet: All five insulation testers are...

Open the catalog to page 2
Agilent U1450A/U1460A - 3

Capable of handling MORE abuse in industrial settings The insulation resistance testers are housed in robust overmold enclosures, certified to IP 67 for protection against water and dust. The robust tool is tested to withstand drops up to 3 meters (10 feet) high and operates over harsh temperature range of –40 to +55 °C1. Built to handle tough conditions and certified to stringent industrial standards, the U1450A/U1460A series is what you need to face the demands of everyday tasks. Test the MOST applications with the same device The adjustable insulation test voltage from 10 V to 1.1 kV in...

Open the catalog to page 3
Agilent U1450A/U1460A - 4

Take a Closer Look OLED front panel Non-Contact AC Voltage Detection (Vsense) OLED display with 2000:1 contrast ratio and 160 degree viewing angle Timed/PI/DAR test Test lock button Adjustable test voltage ranges from 10 V to 1.1 kV (selectable in “Setup” menu) Earth bond resistance test Low pass ilter Auto AC or DC signal identiication IP 67 certiied – waterproof and dust proof CAT III 1000 V/CAT IV 600 V safety ratings

Open the catalog to page 4
Agilent U1450A/U1460A - 5

U1450A/U1460A Series Comparison Chart at a glance Industry Basic features Insulation test voltage range Insulation resistance range Earth bond resistance range Timed only Adjustable insulation test voltages Live circuit test inhibit (30 V, 50 V, 75 V) Calculate cable length by capacitance Automatic discharge for capacitive circuits under test Calibrated for sine wave Calibrated for sine wave Calibrated for sine wave AC/DC voltage, resistance, continuity, capacitance AC/DC voltage, resistance, continuity, capacitance AC/DC voltage, resistance, continuity, capacitance, diode test AC/DC...

Open the catalog to page 5
Agilent U1450A/U1460A - 6

Speciications U1451A/U1452A/U1452AT insulation resistance speciications Insulation resistance specifications with accuracy of ± (% of reading + number of least significant digit) Test voltage Test current The voltage indication on the display refers to the voltage at the DUT (device under test), and the accuracy is according to the DC voltage measurement. The default test voltage is shown in the table below. ΩMega position (Mark) Test voltage Test accuracy The test voltage across a resistor of a value of UN × (1000 Ω/V) shall not differ by more than 10% relative to noload (open) voltage, as...

Open the catalog to page 6
Agilent U1450A/U1460A - 7

U1453A/U1461A insulation resistance speciications Insulation resistance specifications with accuracy of ± (% of reading + number of least significant digit) Test voltage Test current The voltage indication on the display refers to the voltage at the DUT (device under test), and the accuracy is according to the DC voltage measurement. The marked test voltage and the actual test voltage may be different if the test voltage is adjusted in the Setup. Refer to the table below for more details. ΩMega position (Mark) Default test voltage (factory) Adjustable test voltage (user) 2. Incremental 1V...

Open the catalog to page 7
Agilent U1450A/U1460A - 8

U1451A/U1452A/U1452AT earth bond resistance specifications Earth-bond resistance specifications with accuracy of ± (% of reading + number of least significant digit)1' Range Resolution U1451A U1452A/U1452AT Open circuit voltage /. The following statements are true for earth-bond resistance tests: • Overload protection: < 2 Vand 0.44 A/1000 V; 10 x 35 mm 30 kA fast-acting fuse ' Short circuit: > 200.0 mA as resistance <2U 2. The accuracy is specified after the Nuii function is used to subtract the test iead resistance and thermal effect (by shorting the test ieadsj U1453A/U1461A earth bond...

Open the catalog to page 8
Agilent U1450A/U1460A - 9

The following specifications are a requirement for European labeling. Intrinsic uncertainty Earth-bond resistance Insulation resistance Based on the test voltage and range. Refer to insulation resistance 30% /. The maximum resistance to meet the standard of EN61557-1, 5.2'A, which indicates the maximum amount aiiowed as iess than 30% Intrinsic uncertainty (A) Temperature (E3) Operating uncertainty /. Specification confidence ievei to 99.73% as coverage factor up to 3 2. Temperature range is from 0 to 35 °C 3. Test voitage/'maximum range for different modeis: The following specifications are...

Open the catalog to page 9
Agilent U1450A/U1460A - 10

U1453A/U1461A adjustable DC test voltage speciications Adjusted DC test voltage specifications with accuracy of ± (% of reading + number of least significant digit)1, 2 Range Rate current The minimum test voltage may be set from 10 V The indication at the rated output voltage across a resistor with a value of UN × (1000 Ω/V) shall not differ by more than 10% relative to the indicated value, as a result of possibly present AC voltage components in the output voltage, when a capacitor of 1 µF is connected in parallel with the insulation resistance to be measured U1451A/U1452A/U1452AT DC...