Catalog excerpts
Agilent N8900 Series Autoranging System DC Power Supplies N8920A-N8924A (5 kW, 208 VAC) N8925A-N8930A (10 kW, 208 VAC) N8931A-N8937A (15 kW, 208 VAC) N8940A-N8944A (5 kW, 400 VAC) N8945A-N8950A (10 kW, 400 VAC) N8951A-N8957A (15 kW, 400 VAC) Data Sheet • Autoranging output – does the job of multiple power supplies! • 5, 10, and 15 kW maximum output power • Up to 1500 V and up to 510 A • 14 different voltage, current, and power combinations available in 208 or 400 VAC (28 total models) • Just the right amount of performance at just the right price • Easily parallel units to create “one” power supply with > 100 kW of power • Built-in voltage and current measurement Family of Affordable, Autoranging System DC Power Supplies The Agilent Technologies N8900 Series provides 5, 10, and 15 kW autoranging, single-output programmable DC power for ATE applications that require just the right amount of performance at just the right price. The N8900 Series power supplies’ autoranging output characteristic enables unprecedented flexibility by offering a wide range of voltage and current combinations at full power. Power supplies with “rectangular,” or traditional, output characteristics provide full power at only one voltage and current combination. Just one N8900 does the job of multiple power supplies. It’s like having many power supplies in one! The N8900 Series provides stable output power, built-in voltage and current measurements, and autoranging output voltage and current from 80 to 1500 V and 10 to 510 A. These supplies offer many system-ready features like multiple standard I/O interfaces to simplify and accelerate test-system development and compact 3U design to save rack space. If more power is required, easily parallel multiple N8900 units to create “one” power supply with > 100 kW of total output power. The builtin master/slave control enables programming as if it’s just one big power supply; no need to program each supply individually. • High power density, 15 kW in only 3U (5.25 inch/13.34 cm) • Protection from over-voltage, over-current, and overtemperature • LAN (LXI Core), USB, GPIB, and analog interfaces all standard Autoranging output – like having many power supplies in one!
Open the catalog to page 1Autoranging output – does the job of multiple power supplies The N8900 power supplies’ autoranging output characteristic makes it much more flexible than rectangular, or traditional, output characteristic power supplies because they expand the power curve, giving the user more voltage and current combinations in one power supply. It’s like having many rectangular power supplies in one. For example, the 1000 V, 30 A, 10 kW model is capable of 1000 V and 10 A at 10 kW as well as 333.3 V and 30 A at 10 kW. If it were a rectangular output, the specifications would be 1000 V, 10 A, 10 kW. At...
Open the catalog to page 2Simplify system connections The N8900 Series power supplies comes standard with GPIB, Ethernet/LAN, USB 2.0, and analog interfaces giving you the flexibility to use your I/O interface of choice today and safeguard your test setup for the future. There is no need to worry whether or not you are choosing the right interface when they all come standard. The N8900 is fully compliant with the LXI Core specification. Remote access and control The built-in Web server provides remote access and control of the instrument via a standard browser. This control goes above and beyond the LXI...
Open the catalog to page 3Performance specifications All specifications pertain to >2% of rated voltage and >1% of rated current N8920A / N8940A Output voltage ripple and noise CV p-p1 Load effect (change from 0% to 100% of full load) Voltage Programming accuracy (23 °C ± 5 °C) Voltage Measurement accuracy (23 °C ± 5 °C) Voltage Load transient recovery time (time for output voltage to recover within 1% of its rated output for a load change from 10% to 90% of its rated output current) Time Figure 4. Parallel operation for more power (cables not included)
Open the catalog to page 4Supplemental characteristics (typical) N8920A / N8940A Output response time: Time from 10% to 90%, or 90% to 10%, of total voltage excursion Up, full load3 Down, full load Down, no load Command response time < 25 ms Remote sense compensation Volts/load lead Over-voltage protection Range Source effect (±10% of AC input rating) Voltage Output current ripple and noise CC rms Programming resolution Voltage Measurement resolution Voltage Output terminal isolation (negative terminal to ground) 300 V Acoustic noise declaration Idle fan speed 208 VAC input (N8920A - N8937A) Nominal input voltage...
Open the catalog to page 5Supplemental characteristics (typical) - all models Supplemental characteristics Analog programming: Input range Specified accuracy + 0.2% of rating Input impedance Temperature coefficients: (after 30 minute warmup) Voltage 50 PPM/°C of rated output voltage 50 PPM/°C of rated output current Series operation not recommended Parallel operation Master-slave Savable states Nonvolatile memory Interface capabilities GPIB, USB 2.0, 10/100 LAN SCPI - 1993, IEEE 488.2 compliant interface; Requires Agilent I/O Libraries 16.3 or later Environmental conditions Environment Indoor use, installation...
Open the catalog to page 6Outline diagrams DC Output Bus-Bar Detail 80 V and 200 V models
Open the catalog to page 7Available models Figure 5. N8924A autoranging system DC power supply 4. The N8900 Series are autoranging power supplies. The "Current § Max Voltage" and "Voltage § Max Current" are listed to show the full range of voltage and current combintations possible due to the autoranging capability.
Open the catalog to page 8Options None AC input voltages If the AC input voltage where the power supply will be used is: • 187 to 229, please choose a 208 VAC model (N8920A-N8937A) • 360 to 440 VAC, please choose a 400 VAC model (N8940A-N8957A) Rack mount kit for Agilent racks (coming soon) Use standard rack rails for non-Agilent racks. Line cords and terminations (plugs) Due to the number of different line cords and terminations around the world, the N8900 power supplies do not come with line cords or terminations. Users will need to supply their own dependent on the local laws and codes of the country/region where...
Open the catalog to page 9All Agilent Technologies - Life Sciences and Chemical catalogs and technical brochures
-
Agilent 280 Series AA Systems
12 Pages
-
FLAME ATOMIC ABSORPTION SPECTROSCOPY
104 Pages
-
Inorganic Standards
40 Pages
-
InfinityLab LC Supplies
148 Pages
-
General Chromatography
164 Pages
-
Sample Preparation (2018/2019)
144 Pages
-
MS500
12 Pages
-
FLAME ATOMIC ABSORPTION SPECTROSCOPY
104 Pages
-
Agilent training
27 Pages
-
AGILENT VACUUM PRODUCTS CATALOG
536 Pages
-
Vacuum Measurement Catalog
44 Pages
-
AGILENT ION Pumps
53 Pages
-
AGILENT DIFFUSION PUMPS
35 Pages
-
AGILENT Rotary Vane Pumps
29 Pages
-
Agilent InfinityLab LC Series
16 Pages
-
3100 fractionator
8 Pages
-
4200 MP-AES
12 Pages
-
Agilent GPC/SEC Solutions
12 Pages
-
Agilent?s biomass solutions
12 Pages
-
X-ray Crystallography
12 Pages
-
Agilent ProPulse NMR System
5 Pages
-
An autosampler like no other
4 Pages
-
Agilent U1450A/U1460A
19 Pages
-
Agilent B1507A
22 Pages
-
Agilent M9703A
16 Pages
-
W1905 Radar Model Library
16 Pages
-
M8048A ISI Channels
7 Pages
-
Agilent J-BERT M8020A
27 Pages
-
I/O Hardware
16 Pages
-
Advanced Design System
16 Pages
-
RF & Microwave Attenuators
4 Pages
-
E4980AL Precision LCR Meter
8 Pages
-
FieldFox Handheld Analyzers
8 Pages
-
AFM/SPM Accessories
20 Pages
-
VEE 9.32
24 Pages
-
89600 VSA Software
9 Pages
-
Command Expert
2 Pages
-
E5061B Network Analyzer
8 Pages
-
E5071C ENA Network Analyzers
18 Pages
-
N9038A MXE EMI Receiver
2 Pages
-
N9322C Basic Spectrum Analyzer
10 Pages
-
Digital Multimeters
25 Pages
-
M9361A PXI Downconverter
2 Pages
-
Power Meters and Power Sensors
34 Pages
-
E7515A UXM Wireless Test Set
2 Pages
-
Agilent Power Products
31 Pages
-
E5063A Network Analyzer
8 Pages
-
Leak Detection Catalog
41 Pages
-
Agilent 7820A GC brochure
8 Pages
-
Agilent GC-MSD and QTOF
96 Pages
-
Agilent GC-MSD 5977
39 Pages
-
Agilent Intro BERLIN
24 Pages
-
1290 Infinity Quaternary LC
24 Pages
-
1260 Infinity Bio-inert
26 Pages
Archived catalogs
-
Diffusion Pumps Catalog
37 Pages