1. Catalogs
  2. AGC Instruments
  3. Trace N2 Gas Analysers

Trace N2 Gas Analysers

Trace N2 Gas Analysers

Trace N2 Gas Analysers

Product catalog summary
Introduction:
The document introduces the AGC Series 100 SED, a gas chromatography system designed for continuous monitoring of nitrogen (N2) in argon (Ar) or helium (He) at parts per billion (ppb) levels. This system is based on the High Frequency Argon Discharge Detector (ADD) technology.

Principal:
The system measures the emission wavelength intensity of nitrogen molecules to determine trace amounts of nitrogen in argon or helium matrix gases. The response is proportional to nitrogen concentration, ranging from low ppb to high parts per million (ppm).

Technique:
The system uses a spectral emission technique with a quartz ionisation chamber. The sample gas is ionised by a high-frequency oscillator, and the nitrogen emission wavelength is detected through a narrow band pass filter and a photomultiplier tube.

Applications:
The AGC Series 100 SED is applicable in various industries, including argon or helium purification plants, the steel industry, air separation plants, welding gas manufacturing, specialty gas manufacturing, doping gas manufacturing, filling stations, the semiconductor industry, chemical plants, and wafer fabs in the electronic industry.

Specifications:
  • Range: 0.01 – 50 ppm
  • Display: Digital Output (4-½ digit meter)
  • Response Time: 0.01 – 10 ppm < 60 secs (95%)
  • Minimum Detection Limit (MDL): < 0.01 ppm
  • Accuracy: Better than 0.01% of Full Scale
  • Zero Drift: <20 ppb over 24 hours at ±2°C; <50 ppb over 24 hours at ±5°C
  • Span Drift: <10 ppb at approximately 5 ppm N2 in Ar over 24 hours at ±2°C and ±5°C
  • Interference: Relative to O2, CH4, and H2O with specified effects on N2 readings
  • Sample Flow: 50 – 500 mL/min
  • Sample Pressure: 30 to 600 KPa
  • Fittings: 2 x 1/8” VCR inlets, 2 x 1/8” Swagelok outlets
  • Output: Analogue 0-20mA or 4 – 20mA
  • Alarms: 4 Signal Alarm Contacts
  • Electrical Supply: 110/220/240 V, 50/60 Hz; 100 VA
  • Mounting: Bench or 19” rack
  • Dimensions: W = 19” (480 mm); H = 4U (180 mm); D = 450 mm
  • Nett Weight: 25 kgs
  • Certification: CE

Contact Information:
Headquarters: AGC Instruments Ltd, Shannon Free Zone West, Shannon, Co. Clare, Ireland. Contact: +353 61 471632, [email protected]
China Representative Office: AGC Instruments Ltd, Beijing, China. Contact: +86 10 6021 5983, [email protected]

Event Information:
Visit AGC Instruments at ARABLAB, Booth Number 626.
See more

Catalog excerpts

Trace N2 Gas Analysers-1

Series 100 Trace N2 Gas Analyser Analysis of N2 in Argon or in Helium to ppb levels About: The AGC Series 100 SED is engineered for the continuous monitoring of N2 in Ar or N2 in He. Its N2 specific detector has been derived from our extremely successful High Frequency Argon Discharge Detector (ADD) system, the Series 100 HF-ADD Gas Analyser. AGC Instruments has been providing this system configuration for several years to the gas manufacturing companies and especially to argon and helium specific plants throughout the world. Spectral Emission Detector (SED) The emission wavelength intensity of the Nitrogen molecule is selected to determine the trace amount of Nitrogen in Argon or Helium Matrix Gas. The response is proportional to the nitrogen concentration in low ppb to high ppm range. Argon or Helium Purification Plants Steel Industry Air Separation Plants Welding Gas Manufacturing Using a spectral emission technique, a unique quartz ionisation chamber is utilised in which the sample gas is ionised by a high frequency tuned oscillator with the resultant emission wavelength intensity of the nitrogen component is directed through a narrow band pass filter and is detected via a photomultiplier tube. Speciality Gas Manufacturing Plants Doping Gas Manufacturing Filling Stations for Cylinder & Trucks Semiconductor Industry Chemical Plants Wafer Fabs in Electronic Industry Visit us at ARABLAB Gas Chromatography since 1965

 Open the catalog to page 1
Trace N2 Gas Analysers-2

Specifications: Range: Digital Output (4-½ digit meter) Better than 0.01% of Full Scale (1) Zero Drift* : *(1) Over 24 hours at dT of operating temperature ±2°C Zero Drift*(2): *(2) Over 24 hours at dT of operating temperature ±5°C) Span Drift*(3): *(3) At approximately 5 ppm N2 in Ar over 24 hours at dT of sample gas of ±2°C) Span Drift*(4): *(4) At approximately 5 ppm N2 in Ar over 24 hours at dT of sample gas of ±5°C Interference*(5) (Relative to O2): 2ppm O2 will reduce the N2 by 20ppb and 0.1ppm will have no effect. *(5) On a 5 ppm reading N2 in Ar by an increase of O2 concentration from...

 Open the catalog to page 2