Digital, modular pyrometer with exchangable sensor heads for non-contact temperature measurement of various surfaces in different temperature ranges. The Impac® Series 600 is a modular, digital pyrometer series that provides a customizable design with easy installation and maintenance. The long wavelength IN 600 is best suited for non-contact temperature measurement on non-metallic or coated metallic objects. The IS 600, IGA 600, and IGA 600/23 are short wavelengths digital pyrometers for measurements of metals, ceramics or graphite. The IN 600/5 is specifically designed to measure the temperature of glass surfaces. The modular concept allows for various combinations of the system components. The standard configuration includes an electronic converter box, a sensor cable, and a removable sensor head to allow for easy exchange. Alternatively, each converter box can connect up to two sensor heads or up to two optional multi-sensor boxes with connections for up to four sensor heads each. To ensure maximum flexibility, the system allows for any combination of the available sensor heads, even with different wavelengths. Spectral Range Sensor Heads IN 600: 8 to 14 µm IS 600: 0.7 to 1.1 µm IGA 600: 1.45 to 1.8 µm PRODUCT HIGHLIGHTS Removeable sensor heads for easy exchange without the need to remove the complete sensor cable and without recalibration Connection of up to 8 sensor heads to one converter box (8 measuring points in one pyrometer system). If wanted, sensor heads with different wavelengths can be connected to the same converter box. Converter box available with or without display Repeatability IN 600 & IN 600/5: 0.5% of measured °C value or 0.8 °C IS 600, IGA 600 & IGA 600/23: 0.25% of measured °C + 1 °C Optional communications: analog and/or RS-232/RS-485 and/or Profinet ©2022 Advance
Open the catalog to page 1Measurement Specifications Temperature Ranges Spectral Ranges Measurement Uncertainty IN 600: (ε =1, t90 = 1 s, Tamb = 15 to 30°C) T = 0 to 700 °C: 0.7% oR or 1°C T = 0 to −20 °C: 2 °C T = −20 to −40 °C: 3 °C IS 600, IGA 600 & IGA 600/23: 0.5 % oR + 2 °C IN 600/5: 0.5% oR or 0.8 °C TK: 0.03% per °C or 0.05°C per °C (25°C ambient temperature) Repeatability2 IN 600 & IN 600/5: 0.5% of measured °C value or 0.8°C, whichever value is greater IS 600, IGA 600 & IGA 600/23: 0.25% of measured °C + 1 °C Electrical Specifications Power Supply 5 VDC (when connected via USB) or 10 to 30 VDC, ripple <...
Open the catalog to page 2TECHNICAL DATA CONTINUED Environmental Specifications Protection Class Sensor heads (standard) IN 600, IS 600, IGA 600, IGA 600/23, IN 600/5 -20 to 70 °C (-4 to 158 °F), with non-condensing conditions. Sensor head, sensor cables, multi-sensor box, and analog box -20 to 85 °C (-4 to 185 °F), with non-condensing conditions. Storage Temperature Relative Humidity Standard sensor heads Converter box with display Converter box without display Converter box: Aluminium Sensor head: Stainless steel Interface and Communication Digital Interface Optional: RS-232/RS-485 (switchable) or Profinet Analog...
Open the catalog to page 3Converter Box and Multi Analog Box Sensor Head and Connector
Open the catalog to page 4DIMENSIONS CONTINUED
Open the catalog to page 5The IN 600 and IN 600/5 sensor heads are available with 3 different optics: 1N (2:1 distance ratio), 2N (10:1 distance ratio) and 3N (20:1 distance ratio). The spot sizes at different measuring distances are shown in the tables below; intermediate values have to be interpolated. Optics 1N, 2:1 Measuring Distance a [mm] Optics 2N, 10:1 Measuring Distance a [mm] Optics 3N, 20:1 Measuring Distance a [mm]
Open the catalog to page 6OPTICS CONTINUED
Open the catalog to page 7REFERENCE NUMBERS ICB 600 converter box with display, no digital interface ICB 600-RS converter box with display, RS-232/RS-485 ICB 600-PN converter box with display, Profinet ICB 600-N converter box without display, no digital interface ICB 600-N-RS converter box without display, RS-232/RS-485 ICB 600-N-PN converter box without display, Profinet ICB-A converter box, single analog ICB-8A converter box, multi analog Sensor and Extension Cables PN Sensor cable, with connector, 3 m Sensor cable, with connector, 15 m Extension cable, with connector, 15 m Sensor cable, w/o connector, 1 m Sensor...
Open the catalog to page 8REFERENCE NUMBERS (CONTINUED) Electrical Connection Cables PN Electrical connection cable, 2 m, 10-wire, with digital connection (1 m) Electrical connection cable, 15 m, 10-wire, with digital connection (1 m) Electrical connection cable, 2 m, for power supply and thermocouple output Electrical connection cable, 2 m, 7-wire, analog out only Electrical connection cable, 15 m, 7-wire, analog out only Electrical conncetion cable, 2 m, 10-wire, analog out Electrical connection cable, 15 m, 10-wire, analog out Electrical connection cable, 2 m, 12-wire, analog out Electrical connection cable, 15...
Open the catalog to page 9ABOUT ADVANCED ENERGY Advanced Energy (AE) has devoted more than three decades to perfecting power for its global customers. AE designs and manufactures highly engineered, precision power conversion, measurement and control solutions for mission-critical applications and processes. Our products enable customer innovation in complex applications for a wide range of industries including semiconductor equipment, industrial, manufacturing, telecommunications, data center computing, and medical. With deep applications know-how and responsive service and support across the globe, we build...
Open the catalog to page 10All ADVANCED ENERGY catalogs and technical brochures
-
CSM Power solutions Catalog
12 Pages
-
Horticultural Lighting sheet
1 Pages
-
LCM brochure
4 Pages
-
eVos me Datasheet
2 Pages
-
eVos datasheet
2 Pages
-
eVerest datasheet
12 Pages
-
Navigator Networks datasheet
8 Pages
-
Paramount Plus datasheet
8 Pages
-
Ascent AMS datasheet
6 Pages
-
Ascent AP datasheet
8 Pages
-
Ascent DMS datasheet
8 Pages
-
Medical Power Supplies Brochure
20 Pages
-
SCR Thyro-PX DC datasheet
8 Pages
-
Impac IGAR 6 Datasheet
8 Pages
-
Luxtron M-1000 Datasheet
4 Pages
-
Luxtron M-1100 Datasheet
4 Pages
-
Glass Solutions Brochures
20 Pages
-
Petrochemical Solutions Brochure
16 Pages
-
Impac Pyrometer Overview
28 Pages
-
Basic Bus Module
2 Pages
-
Thyro S Series Datasheet
10 Pages
-
Thyro PX Series Datasheet
11 Pages
-
Thyro A Series Datasheet
10 Pages
-
Thyro A+ Series Datasheet
8 Pages
-
FBG Analyzer Series
4 Pages
-
ON-SITE INFRARED SURVEY
4 Pages
-
xgen series
56 Pages
-
THYRO-FAMILY ACCESSORIES
12 Pages
-
HIGH VOLTAGE POWER SUPPLIES
8 Pages
-
Apex 3k5PIN Conversion
4 Pages
-
UV 400 Pyrometer Datasheet
8 Pages
-
Luxtron FOT probes
2 Pages
-
Luxtron M924 FOT OEM-Module
4 Pages
-
E2T Pulsar4 Datasheet
8 Pages
-
E2T Quasar 2 Datasheet
4 Pages
-
FlareSpection Datasheet
4 Pages
-
ThermalSpection 724DV Datasheet
12 Pages
-
BoilerSpection Datasheet
4 Pages
-
Innova 3731 Datasheet
8 Pages
-
Gas Sensing Overview
16 Pages
-
XSOLO
8 Pages
-
XF
8 Pages
-
COOLX®600
8 Pages
-
COOLX®1800
12 Pages
-
COOLX®1000
11 Pages
-
POWER FOR GLASS COATING
4 Pages
-
ASCENT AMS DC POWER SUPPLIES
6 Pages
-
IMPAC IS 50-LO/GL
4 Pages
-
IMPAC IPE 140
8 Pages
-
IMPAC IPE 140/39
2 Pages
-
IMPAC IN 140/5 SERIES
8 Pages
-
IMPAC SERIES 8 PRO
8 Pages