Intended Use: The S1806 ultrasonic transducer is designed for generating shear horizontal wave modes.
Main Technical Specifications:
  • Nominal Frequency: 100 kHz
  • Operating Frequency: 90-110 kHz
  • Double Conversion Ratio: 70 dB or better
  • Relative Frequency Bandwidth: Greater than 70%
  • Electric Capacity: 450 ± 200 pF
  • Maximum Excitation Pulse Voltage: Less than 200 V
  • Connector Type: ERN.00.250
  • Dimensions: Less than 45ר15 mm
  • Weight: Less than 20 g
  • Operating Temperature Range: -20 °C to +50 °C
Measurement Conditions and Equipment Used: The transducer's characteristics are measured in conjunction with a reference transducer. The tested transducer acts as a transmitter, while the reference transducer serves as a receiver. The double conversion ratio is calculated by comparing the received signal amplitude to the excitation pulse amplitude. The excitation signal is a square pulse with a 200 V amplitude and 10 μs duration. The receiver uses an integrating amplifier with a bandwidth of 0.01 Hz to 400 kHz, input resistance of 4 kΩ, and equivalent input noise voltage of 10 μV. Measurements are conducted at 24 °C with 25% relative humidity.
Measured Characteristics:
  • Echo Pulse Duration: 22.8 μs
  • Operating Frequency: 84.7 kHz
  • AFC Frequency Maximum: 98.0 kHz
  • Double Conversion Ratio AFC Maximum: -102.2 dB
  • Lower AFC Frequency: 46.5 kHz
  • Transducer Delay: 2.8 μs
  • Upper AFC Frequency: 122.8 kHz
  • Relative Frequency Bandwidth (at -6 dB): 90.1%
Environmental Conditions: Measurements are taken at a temperature of 24 °C and relative humidity of 25%.
See more

Catalog excerpts

S1806 - shear wave DPC transducer 100 KHz-1

Ultrasonic transducer S1806 DATA SHEET Intended use Main technical specifications Type of generated wave mode: Nominal frequency: Operating frequency: Double conversion ratio: Relative frequency bandwidth: Electric capacity of the piezoelectric element: Maximum excitation pulse voltage: Connector type: Overall dimensions: Weight: Operating temperature range: Shear horizontal 100 kHz (100 ± 10) kHz 70 dB or better > 70 % (450 ± 200) pF < 200 V ERN.00.250 < 45ר15 mm < 20 g -20 °C to +50 °C Measurement conditions and equipment used The measurement of the tested DPC transducer characteristics occurs in combination with the reference DPC transducer, whereby both transducers are connected by their tips with the nip force of 4 N. The tested transducer operates as a trans- mitter and the reference transducer operates as a receiver of ultrasonic waves. The double conversion ratio Srel is determined as a ratio value between the received signal amplitude on the reference transducer and excitation pulse amplitude on the tested transducer. Excitation signal: square pulse with the amplitude 200 V, duration 10 μs, equal to half period of the nominal. Receiver parameters: integrating amplifier with the bandwidth 0.01 Hz – 400 kHz, input resistance 4 kΩ, equivalent input noise voltage 10 μV. Environmental temperature 24 °C, rel. 25 %. conditions: Amplitude-frequency response Level, rel.un. Measured characteristics Shape of the measured pulse Time, μs Echo pulse duration τ: AFC frequency maximum ƒp: Lower AFC frequency ƒl: Upper AFC frequency ƒu: ©Acoustic Control Systems - ACS Group Frequency, kHz Operating frequency ƒc: Double conversion ratio AFCmaximum Srel: Transducer delay td: Relative frequency bandwidth (at -6 dB) Bw:

 Open the catalog to page 1

All ACS-Solutions GmbH catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.