Intended Use: The document outlines the specifications and measurement conditions for the Ultrasonic Transducer S1802 0.05A0D2PS, designed for generating shear horizontal wave modes.
Main Technical Specifications:
  • Wave Mode: Shear horizontal
  • Nominal Frequency: 50 kHz
  • Operating Frequency: 50 ± 10 kHz
  • Double Conversion Ratio: 70 dB or better
  • Relative Frequency Bandwidth: > 75%
  • Electric Capacity: 650 ± 100 pF
  • Maximum Excitation Pulse Voltage: < 200 V
  • Connector Type: ERN.00.250
  • Dimensions: < 43ר15 mm
  • Weight: < 20 g
  • Operating Temperature Range: -20 °C to +50 °C
Measurement Conditions and Equipment Used: The transducer's characteristics are measured in combination with a reference transducer. The tested transducer acts as a transmitter, while the reference transducer acts as a receiver. The double conversion ratio is calculated based on the amplitude of the received signal and the excitation pulse.
Excitation Signal: Square pulse with 200 V amplitude and 10 μs duration.
Receiver Parameters: Integrating amplifier with 0.01 Hz – 400 kHz bandwidth, 4 kΩ input resistance, and 10 μV equivalent input noise voltage.
Environmental Conditions: Temperature at 24 °C and relative humidity at 25%.
Measured Characteristics:
  • Echo Pulse Duration: 41.3 μs
  • Operating Frequency: 49.2 kHz
  • AFC Frequency Maximum: 53.2 kHz
  • Double Conversion Ratio AFC Maximum: -93.3 dB
  • Lower AFC Frequency: 28.0 kHz
  • Transducer Delay: 4.0 μs
  • Upper AFC Frequency: 70.4 kHz
  • Relative Frequency Bandwidth (at -6 dB): 86.0%
See more

Catalog excerpts

S1802 – shear wave DPC transducer 50 KHz-1

Ultrasonic transducer S1802 0.05A0D2PS PASSPORT Intended use Main technical specifications Type of generated wave mode: Nominal frequency: Operating frequency: Double conversion ratio: Relative frequency bandwidth: Electric capacity of the piezoelectric element: Maximum excitation pulse voltage: Connector type: Overall dimensions: Weight: Operating temperature range: Shear horizontal 50 kHz (50 ± 10) kHz 70 dB or better > 75 % (650 ± 100) pF < 200 V ERN.00.250 < 43ר15 mm < 20 g -20 °C to +50 °C Measurement conditions and equipment used The measurement of the tested DPC transducer characteristics occurs in combination with the reference DPC transducer, whereby both transducers are connected by their tips with the nip force of 4 N. The tested transducer operates as a trans- mitter and the reference transducer operates as a receiver of ultrasonic waves. The double conversion ratio Srel is determined as a ratio value between the received signal amplitude on the reference transducer and excitation pulse amplitude on the tested transducer. Excitation signal: square pulse with the amplitude 200 V, duration 10 μs, equal to half period of the nominal. Receiver parameters: integrating amplifier with the bandwidth 0.01 Hz – 400 kHz, input resistance 4 kΩ, equivalent input noise voltage 10 μV. Environmental temperature 24 °C, rel. 25 %. conditions: Amplitude-frequency response Level, rel.un. Measured characteristics Shape of the measured pulse Time, μs Echo pulse duration τ: AFC frequency maximum ƒp: Lower AFC frequency ƒl: Upper AFC frequency ƒu: ©Acoustic Control Systems - ACS Group Frequency, kHz Operating frequency ƒc: Double conversion ratio AFCmaximum Srel: Transducer delay td: Relative frequency bandwidth (at -6 dB) Bw:

 Open the catalog to page 1

All ACS-Solutions GmbH catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.