Product catalog summary
Transducer Specifications
  • Type: Normal beam EMAT transducer with permanent magnet for generating ultrasonic shear waves with linear polarization.
  • Nominal Frequency: 3 MHz
  • Effective Aperture Diameter: 10 mm
  • Inspection Range: 1 to 200 mm (using A1270 EMAT)
  • Lift-off/Through-coating Thickness: Up to 2 mm (for inspection range up to 50 mm)
  • Maximal Excitation Pulse Voltage: 500 V
  • Direct Current Resistance: 1.5 ± 0.3 Ohm
  • Operating Temperature Range: -20 to +60°C
  • Extended Operating Temperature Range: -20 to +800°C (with high-temp ACS probe holder)
  • Overall Dimensions: 35x58 mm
  • Socket Type: LEMO ERN.00.250
  • Weight: 255 g
Measurement Conditions and Equipment
  • Reference Excitation Signal: Unipolar square pulse, 400 V ± 40 V, 200 ± 13 ns duration at 50% max voltage amplitude.
  • Calibration Specimen: CO-2 made of steel 20, longitudinal wave velocity 5930 m/s, shear wave velocity 3247 m/s.
  • Measured Pulse: Echo pulse from the backwall of CO-2, depth 30 mm.
  • Induced Noise: White thermal noise with 2 mV effective amplitude.
Measured Characteristics
  • Echo Pulse Duration: 0.94 µs
  • Echo Pulse Amplitude: 58.8 µV
  • Bandwidth: 2.2 MHz
  • Relative Bandwidth: 76%
  • Amplitude Frequency Response:
    • Peak Frequency: 2.8 MHz
    • Lower Cut-off Frequency: 1.8 MHz
    • Upper Cut-off Frequency: 4.1 MHz
    • Centre Frequency: 3 MHz
Reverberation Noise Curve (RNC)
  • Signal-to-Noise Ratio: 33 dB (backwall signal vs. transducer self-noise), 15 dB (with electromagnetic noise)
  • RNC Level at 5 µs: -153 dB
Autocorrelation Function (ACF)
  • Amplitude of First Maximum: 0.36
  • Time Position of First Maximum: 0.34 µs
Notes: The RNC is normalized by test excitation signal amplitude and presented in a logarithmic scale. The solid line indicates the transducer RNC, while the dashed line shows the amplitude of induced noise combined with the RNC. The dot indicates the echo pulse amplitude received on the CO-2 specimen.
See more

Catalog excerpts

Electromagnetic acoustic transducer S7394-1

Electromagnetic acoustic transducer S7394 DATA SHEET Transducer type: Nominal frequency: normal beam, EMAT transducer with permanent magnet for generating ultrasonic shear waves with linear polarization. 3 MHz Effective aperture diameter Inspection range: 1 up to 200 mm (when using A1270 EMAT) Lift-off / through-coating thickness: Maximal excitation pulse voltage: Direct current resistance if signal inductor: Operating temperature range: Extended operating temperature range: from -20 to +800 ◦ C (when using high-temp ACS probe holder) 35x58 mm LEMO ERN.00.250 255 g Overall dimensions: Type of socket: Weight: MEASUREMENT CONDITIONS AND EQUIPMENT Reference excitation signal: unipolar square pulse with amplitude 400 V ± 40 V, pulse duration 200 ± 13 ns by 50% of the maximum voltage amplitude. Calibration specimen: CO-2 made of steel 20, serial number 006, longitudinal wave velocity 5930 m/s, shear waves velocity 3247 m/s. Measured pulse: echo pulse from the backwall of CO-2, depth 30 mm. Induced noise: white thermal noise with 2 mV effective amplitude, generated in inductor coil placed adjacent to the protector of the transducer. MEASURED CHARACTERISTICS Amplitude frequency response Shape of the measured echo pulse Echo pulse duration: Echo pulse amplitude: Ae : Bandwidth Π: Relative bandwidth Bw : Peak frequency fp : 2.8 MHz Lower cut-off frequency fl : 1.8 MHz Upper cut-off frequency fu : 4.1 MHz Centre frequency f

 Open the catalog to page 1
Electromagnetic acoustic transducer S7394-2

Reverberation noise curve (RNC) Time, µs Signal-to-noise ratio between the backwall signal in the reference block and transducer self-noise: Signal-to-noise ratio between the backwall signal in the reference block transducer self-noise in presence of electromagnetic noise: RNC level at 5 µs: Time, µs Amplitude of the first maximum of ACF: 0.36 Time position of the first maximum of ACF: 0.34 µs Note: The RNC is normalized by test excitation signal amplitude and is given in logarithmic scale. Transducer RNC is indicated by the solid line. The dash line shows the amplitude of induced noise in sum...

 Open the catalog to page 2

All ACS-Solutions GmbH catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.