1. Catalogs
  2. ACS-Solutions GmbH
  3. D2763 – normal-beam dual-crystal ultrasonic transducer 10 MHz

D2763 – normal-beam dual-crystal ultrasonic transducer 10 MHz

D2763 – normal-beam dual-crystal ultrasonic transducer 10 MHz

D2763 – normal-beam dual-crystal ultrasonic transducer 10 MHz

Product catalog summary
Main Technical Specifications
  • Transducer Type: Contact straight beam double
  • Nominal Frequency: 10 MHz
  • Nominal Echo Pulse Duration: 1.2 µs
  • Nominal Relative Band Width: 35%
  • Nominal Sensitivity: -70 dB
  • Piezoelement Diameter: 6 mm
  • Nominal Echo Pulse Delay in Protector: 0.15 µs
  • Nominal Piezoelement Capacity: 800 ± 50 pF
  • Connector Type: LEMO 00.250
  • Operation Temperature Range: -30 to +50 °C
  • Dimensions: 50×18×49 mm
  • Weight: 15 g
Measurement Conditions and Equipment
  • Excitation: Rectangular pulse, 20 V amplitude, 50 ns duration
  • Receiver: Amplifier with 0.01-15 MHz bandwidth, 400 Ω input impedance
  • Damping Resistor: 100 Ω
  • Cable: Single LEMO-LEMO, 50 Ω wave resistance, 1.2 m length
  • Calibration Blocks: Standard steel samples with various thicknesses
Measurement Results
  • Backwall echo pulse and spectrum for 20 mm thickness
  • Reverberation noise characteristics without acoustic load
Calculated Parameters and Acceptance Results
  • Work Frequency: 9.5 MHz (within 8 – 12 MHz tolerance)
  • Echo Pulse Duration: 0.8 µs (<= 1.2 µs tolerance)
  • Relative Spectrum Bandwidth: 36% (within 10 – 60% tolerance)
  • Sensitivity: -59 dB (>= -70 dB tolerance)
  • Echo Pulse Amplitude: 21 mV
  • Delay: 3.3 µs
  • Spectrum Maximum Frequency: 10.2 MHz
  • Lower Spectrum Frequency: 7.3 MHz
  • Upper Spectrum Frequency: 11 MHz
  • Spectrum Bandwidth: 3.6 MHz
See more

Catalog excerpts

D2763 – normal-beam dual-crystal ultrasonic transducer 10 MHz-1

Ultrasonic piezoelectric transducer D2763 DATA SHEET Main technical specifications Transducer type: Nominal frequency: Nominal echo pulse duration: Nominal relative band width: Nominal sensitivity: Piezoelement diameter: Nominal echo pulse delay in protector: Nominal piezoelement capacity: Connector type: Operation temperature range: Dimensions: Weight: Contact straight beam double 10 MHz 1.2 µs 35 % -70 dB 6 mm 0.15 µs 800 ± 50 pF LEMO 00.250 from -30 to +50 ◦ C 50×18×49 mm 15 g Measurement conditions and used equipment Excitation: Rectangular pulse with amplitude 20 V and duration 50 ns, equal to half-period of nominal frequency oscillations. Amplifier with 0.01-15 MHz bandwidth and 400 Ω input impedance. Effective noise level, normalized to the amplifier input level, is less than 20 µV. Damping resistor: 100 Ω (connected in parallel to the transducer). Single LEMO-LEMO with wave resistance 50 Ω and 1.2 m length. Calbration blocks: Standard parallel-sided steel samples, ultrasonic longitudinal wave velocity 5910 m/s, with thickness 100 mm, 50 mm, 30 mm, 20 mm, 10 mm, 2.5 mm, 1.5 mm, 1 mm, 0.7 mm. Measurement results Spectral amplitude, rel. units Backwall echo pulse for 20 mm thickness and its spectrum

 Open the catalog to page 1
D2763 – normal-beam dual-crystal ultrasonic transducer 10 MHz-2

Reverberation noise characteristics (RNC) for the transducer without acoustic load and the curve of backwall echo signal level for steel samples of different thickness The level of 0 dB corresponds to the excitation pulse amplitude. Calculated parameters and acceptance results Parameter Work frequency (Mean of border spectrum frequencies) , MHz Echo pulse duration (at -20 dB level from maximum) , µs Relative spectrum bandwidth (at -6 dB level) , % Sensitivity (bottom echo pulse and excitation pulse amplitudes’ ratio), dB Echo pulse amplitude, mV Spectrum maximum frequency, MHz Lower spectrum...

 Open the catalog to page 2

All ACS-Solutions GmbH catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.