D1771 – normal-beam dual-crystal transducer 5 MHz

D1771 – normal-beam dual-crystal transducer 5 MHz

D1771 – normal-beam dual-crystal transducer 5 MHz

Product catalog summary
Main Technical Specifications
  • Transducer Type: Contact straight beam double
  • Nominal Frequency: 4 MHz
  • Nominal Echo Pulse Duration: 0.6 µs
  • Nominal Relative Bandwidth: 90%
  • Nominal Sensitivity: -60 dB
  • Piezoelement Diameter: 12 mm
  • Nominal Echo Pulse Delay in Protector: 0.15 µs
  • Nominal Piezoelement Capacity: 1500 ± 150 pF
  • Connector Type: LEMO 00.250
  • Operation Temperature Range: -30 to +50 °C
  • Dimensions: 23×44×15 mm
  • Weight: 22 g
Measurement Conditions and Equipment
  • Excitation: Rectangular pulse with 20 V amplitude and 125 ns duration
  • Receiver: Amplifier with 0.01-15 MHz bandwidth and 400 Ω input impedance
  • Damping Resistor: 100 Ω (parallel to transducer)
  • Cable: Single LEMO-LEMO, 50 Ω wave resistance, 1.2 m length
  • Calibration Blocks: Standard steel samples with ultrasonic longitudinal wave velocity of 5910 m/s
Measurement Results
  • Backwall echo pulse for 20 mm thickness and its spectrum
  • Reverberation noise characteristics (RNC) without acoustic load
  • Backwall echo signal level for steel samples of varying thickness
Calculated Parameters and Acceptance Results
  • Work Frequency: 4.1 MHz (within 3.2 – 4.8 MHz tolerance)
  • Echo Pulse Duration: 0.41 µs (<= 0.6 µs tolerance)
  • Relative Spectrum Bandwidth: 85% (within 70 – 110% tolerance)
  • Sensitivity: -39 dB (>= -60 dB tolerance)
  • Echo Pulse Amplitude: 211 mV
  • Delay: 2.3 µs
  • Spectrum Maximum Frequency: 4.1 MHz
  • Lower Spectrum Frequency: 2.4 MHz
  • Upper Spectrum Frequency: 5.8 MHz
  • Spectrum Bandwidth: 3.5 MHz
See more

Catalog excerpts

D1771 – normal-beam dual-crystal transducer 5 MHz-1

Ultrasonic piezoelectric transducer D1771 DATA SHEET Main technical specifications Transducer type: Nominal frequency: Nominal echo pulse duration: Nominal relative band width: Nominal sensitivity: Piezoelement diameter: Nominal echo pulse delay in protector: Nominal piezoelement capacity: Connector type: Operation temperature range: Dimensions: Weight: Contact straight beam double 4 MHz 0.6 µs 90 % -60 dB 12 mm 0.15 µs 1500 ± 150 pF LEMO 00.250 from -30 to +50 ◦ C 23×44×15 mm 22 g Measurement conditions and used equipment Excitation: Rectangular pulse with amplitude 20 V and duration 125 ns, equal to half-period of nominal frequency oscillations. Amplifier with 0.01-15 MHz bandwidth and 400 Ω input impedance. Effective noise level, normalized to the amplifier input level, is less than 20 µV. Damping resistor: 100 Ω (connected in parallel to the transducer). Single LEMO-LEMO with wave resistance 50 Ω and 1.2 m length. Calbration blocks: Standard parallel-sided steel samples, ultrasonic longitudinal wave velocity 5910 m/s, with thickness 100 mm, 50 mm, 30 mm, 20 mm, 10 mm, 2.5 mm, 1.5 mm, 1 mm, 0.7 mm. Measurement results Backwall echo pulse for 20 mm thickness and its

 Open the catalog to page 1
D1771 – normal-beam dual-crystal transducer 5 MHz-2

Reverberation noise characteristics (RNC) for the transducer without acoustic load and the curve of backwall echo signal level for steel samples of different thickness The level of 0 dB corresponds to the excitation pulse amplitude. Calculated parameters and acceptance results Parameter Work frequency (Mean of border spectrum frequencies) , MHz Echo pulse duration (at -20 dB level from maximum) , µs Relative spectrum bandwidth (at -6 dB level) , % Sensitivity (bottom echo pulse and excitation pulse amplitudes’ ratio), dB Echo pulse amplitude, mV Spectrum maximum frequency, MHz Lower spectrum...

 Open the catalog to page 2

All ACS-Solutions GmbH catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.