D1471 – normal-beam dual-crystal transducer 4 MHz

D1471 – normal-beam dual-crystal transducer 4 MHz

D1471 – normal-beam dual-crystal transducer 4 MHz

Product catalog summary
Main Technical Specifications
  • Measurement Conditions and Equipment: The system uses a rectangular pulse with an amplitude of 20 V and a duration of 50 ns. An amplifier with a bandwidth of 0.01-15.00 MHz and an input impedance of 3.6 kΩ is used. Signals are filtered using a digital bandpass filter with a bandwidth from 1 to 14 MHz.
  • Excitation: A 200 Ω damping resistor is connected in parallel to the transducer.
  • Cable: Double LEMO-LEMO cable with a wave resistance of 50 Ω and a length of 3 m.
  • Calibration Blocks: Standard steel samples with ultrasonic longitudinal wave velocity of 5910 m/s and varying thicknesses from 0.7 mm to 100.0 mm.
Measurement Results
  • Backwall Echo Pulse: For a 20 mm thickness, the transducer type is contact straight beam double with a nominal frequency of 4 MHz and a nominal echo pulse duration of 0.8 µs.
  • Transducer Specifications: Nominal relative bandwidth is 40%, nominal sensitivity is minus 60 dB, piezoelement diameter is 12 mm, and nominal piezoelement capacity is 1500 ± 150 pF. The connector type is 2 x LEMO 00.250, with an operation temperature range from -30 to +50 °C. Dimensions are 36 x 24 mm, and weight is 58 g.
Reverberation Noise Characteristics (RNC): The level of 0 dB corresponds to the excitation pulse amplitude. The axes are marked minus the ultrasound delay time in the prisms.
Calculated Parameters and Acceptance Results
  • Work Frequency: 3.9 MHz, within the tolerance of 3.2 to 4.8 MHz.
  • Echo Pulse Duration: 0.51 µs, less than the maximum of 0.8 µs.
  • Relative Spectrum Bandwidth: 80%, exceeding the minimum of 40%.
  • Sensitivity: 41 dB, less than the maximum of 60 dB.
  • Amplitude Difference on 20 mm Sample: 55, more than the minimum of 26.
  • Other Parameters: Echo pulse amplitude is 185 mV, spectrum maximum frequency is 3.8 MHz, spectrum bandwidth is 3.1 MHz, lower spectrum frequency is 2.4 MHz, upper spectrum frequency is 5.5 MHz, and delay is 2.0 µs.
See more

Catalog excerpts

D1471 – normal-beam dual-crystal transducer 4 MHz-1

Ultrasonic piezoelectric transducer D1471 DATA SHEET Main technical specifications Transducer type Nominal frequency Nominal echo pulse duration Nominal relative band width Nominal sensitivity Piezoelement diameter Coordinating inductance Nominal piezoelement capacity Connector type Operation temperature range Dimensions Weight contact straight beam double 4 MHz 0,8 µs 40 % minus 60 dB 12 mm absent 1500  150 pF 2 x LEMO 00.250 о from minus 30 to plus 50 С 36  24 mm 58 g Measurement conditions and used equipment Excitation Rectangular pulse with amplitude 20 V and duration 50 ns, equal to half-period of nominal frequency oscillations. Amplifier with 0,01-15,00 MHz bandwidth and 3,6 kΩ input impedance. Effective noise level, normalized to the amplifier input level, is less than 20 µV. Signals are pre-cleaned from low-frequency components using a digital bandpass filter with a bandwidth from 1 to 14 MHz. Damping resistor 200 Ω (connected in parallel to the transducer). Double LEMO-LEMO with wave resistance 50 Ω and 3 m length. Calibration blocks Standard parallel-sided steel samples, ultrasonic longitudinal wave velocity 5910 m/s, with thickness 0,7; 1,0; 1,5; 2,5; 10,0; 20,0; 30,0; 50,0; 100,0 mm. Measurement results Backwall echo pulse for 20 mm thickness and its

 Open the catalog to page 1
D1471 – normal-beam dual-crystal transducer 4 MHz-2

Reverberation noise characteristics (RNC) for the transducer without acoustic load and the curve of backwall echo signal level for steel samples of different The level of 0 dB corresponds to the excitation pulse amplitude. The time and thickness axes are marked minus the ultrasound delay time in the prisms. Calculated parameters and acceptance results Value Work frequency (mean of border spectrum frequencies), MHz Echo pulse duration (at minus 20 dB level from maximum), µs Relative spectrum bandwidth (at minus 6 dB level), % Sensitivity (bottom echo pulse and excitation pulse amplitudes’ ratio),...

 Open the catalog to page 2

All ACS-Solutions GmbH catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.