Sodium Monitor Model 8036
8Pages

{{requestButtons}}

Catalog excerpts

Sodium Monitor Model 8036 - 1

Sodium Monitor Model 8036 ƒÞ Wide measuring range (0.01ƒÝgkg–1 to 10mgkg–1) – covers all power station applications ƒÞ Full sample temperature compensation – maximises measurement accuracy ƒÞ 'Pumpless' liquid handling section – low maintenance ƒÞ Automatic range changing – switches the current output in steps covering two decades ƒÞ Simple calibration sequence – automatically controlled by the microprocessor Data Sheet SS/8036_5

Open the catalog to page 1
Sodium Monitor Model 8036 - 2

Sodium Monitor Model 8036 SS/8036_5 2 The prevention of damage to the boilers on modern power stations becomes more and more critical as the cost of repairs and out of service plant continually rises. To reduce the effect of boiler tube corrosion and the formation of scale on the inside of the tubes, the levels of impurities in the various parts of the steam/water cycle must be carefully monitored. Sodium ions are normally the most abundant species in solution found in boiler plant, therefore the measurement of sodium provides a valuable indication of the overall purity of the solution....

Open the catalog to page 2
Sodium Monitor Model 8036 - 3

Sodium Monitor Model 8036 SS/8036_5 3 General Information The ABB 8036 Sodium Monitor is a microprocessor-based instrument which uses a sodium ion selective electrode together with a reference electrode to carry out a potentiometric determination of the sodium ion concentration in the sample. The monitor is designed to meet the exacting standards for sodium measurement of nuclear and fossil-fuelled power stations, providing fast response and high accuracy at both high and low sodium concentration levels. The electronic transmitter unit is housed in a wall-mounted steel case. The liquid...

Open the catalog to page 3
Sodium Monitor Model 8036 - 4

Sodium Monitor Model 8036 SS/8036_5 4 Pre-amplifier Constant Head Unit Standard Solution Container Sample/Calibrate Solenoid Valve Sample Pressure Inlet Relief Valve Drain Flowcell Reference Electrode Sodium Electrode Thermistor Valve Supply Output to Transmitter 24V Supply Temperature Compensation Entrainment T Piece Reagent Heater Reagent Solution Container Schematic of Liquid Handling Section Calibration Method A calibration solution bottle containing a sodium solution of known concentration is placed in a tray on top of the liquid handling case. The value of the solution is fed into the...

Open the catalog to page 4
Sodium Monitor Model 8036 - 5

Sodium Monitor Model 8036 SS/8036_5 5 8036-100 Inputs/Outputs 8036-100 Transmitter Unit Sodium EIL8036 Mains Supply 100/240V Temperature Sensor Input Input from Pre-amplifier Alarm 1 Output Alarm 2 Output Calibration Mode Indication Output Calibration Fail Indication Output Calibration Change-over Valve 24V Solenoid Supply Two Isolated current Outputs 0 to 10mA, 0 to 20mA or 4 to 20mA Range Indication Outputs 1234 Transmitter Unit The current from the pre-amplifier representing the output from the electrode pair and the temperature information is processed by the main electronic section...

Open the catalog to page 5
Sodium Monitor Model 8036 - 6

Sodium Monitor Model 8036 SS/8036_5 6 Specification Ranges 0.01ƒÝgkg–1 to 1mgkg–1 or 0.1ƒÝgkg–1 to 10mgkg–1 internally selectable Accuracy „b10% of concentration or „b0.02ƒÝgkg–1 whichever is the greater Reproducibility „b5% of concentration at constant temperature Response time 1 to 100ƒÝgkg–1 less than 4 mins. for 90% step change 100 to 1ƒÝgkg–1 less than 6 mins. for 90% step change Outputs Two isolated current outputs in the range 0 to 10, 0 to 20 or 4 to 20mA. Max. impedance 1kƒÇ logarithmic or linear. Remote range indication Two voltage-free contacts rated 250V 2A non-inductive...

Open the catalog to page 6
Sodium Monitor Model 8036 - 7

Sodium Monitor Model 8036 SS/8036_5 7 Model 8036 Electrical Connections TB2 TB5 TB6 TB4 TB1 TB3 TB7 RNG 1 RNG 2 RNG 3 RNG 4 0V Remote Indication of Measurement Range 0V L N E Remote Cal IOUT 1 IOUT 2 0V 0V +– +– NO COM NC NO COM NC NO COM NC NO COM NC Calibration Indication Calibration Fail Alarm 1 Alarm 2 Th1 S1 S2 Th2 0V SOL 24S 24V SOL2 Remote Switch Inhibit External Alarm Contacts Connections to Sensor Unit Mains Power Supply 1234 Ordering Information Supplied with monitor Instruction manual Low level sodium and reference electrode 2.5m of connection cable Additional option Sensor unit...

Open the catalog to page 7
Sodium Monitor Model 8036 - 8

Sodium Monitor Model 8036 SS/8036_5 Kent-Taylor Sodium EIL 8036 Kent-Taylor Sodium EIL 8036 490mm 8036-100/200 With Door Open Interconnecting Cable Gland Terminal Box 200mm 10mm 20mm 300mm 300mm 230mm 8.5mm Earth Stud M6 20mm 230mm Earth Stud M6 56mm Terminal Box 200mm 10mm 400mm Clean Drain Contaminated Drain 25m 75mm 25m 40mm 300m 230m 8.5m 35mm To Fit 10mm Flexible Tube 160mm 20mm Earth Stud M6 330mm Inter- Connecting Conduit 1/4" O.D. Stainless Steel Tube Connect to a good known earth 6.3mm O.D. Compression Fitting 332m Drains Sample Inlet Interconnection Conduit ABB Limited Oldends...

Open the catalog to page 8

All ABB Measurement & Analytics catalogs and technical brochures

  1. ProKiln GAC400

    24 Pages

  2. AV1 and AV2

    24 Pages

  3. SCK

    20 Pages

  4. INDI-BOY DSP

    8 Pages

  5. 9QGPS4500

    8 Pages

  6. 9QGPS3100P

    12 Pages

  7. 9QGPS2960

    12 Pages

  8. 9QGPS2625

    8 Pages

  9. MB3600-HP10

    5 Pages

  10. MB3600-CH10

    8 Pages

  11. ACF5000

    16 Pages

  12. Magnos28

    10 Pages

  13. EL3060

    8 Pages

  14. HP30

    4 Pages

  15. Aztec ATS430

    12 Pages

  16. LMS200

    12 Pages

  17. PGC5000

    8 Pages

  18. MR Series

    2 Pages