ACF5000

ACF5000
1 / 16 PagesView full catalog

ACF5000

Product catalog summary
Overview: The ACF5000 is an advanced FTIR emission monitoring system designed for continuous and precise measurement of multiple gases such as HCl, HF, H2O, CO, CO2, SO2, NO, NO2, CH4, NH3, N2O, H2CO, O2, and VOCs. It can measure up to 15 components and is known for its stability, accuracy, and reliability, using a hot wet extractive measurement technique.
System Description: The ACF5000 addresses the need for environmental monitoring by accurately measuring low-concentration pollutants from combustion processes. It features a robust design that minimizes drift and reduces the need for frequent adjustments. The system is software-expandable for additional components and includes a cost-effective air-driven injector pump.
Applications: This system is versatile, suitable for use in municipal waste incinerators, biomedical and sludge incinerators, hazardous waste incinerators, gasification and pyrolysis processes, cement kilns, solvent recovery, DeNOx and DeSOx processes, crematoria, and metal smelters.
System Devices and Subassemblies: The ACF5000 includes a sampling system with a heated probe tube, filtering device, sample gas line, and protective cover. It also comprises a sample gas conditioning block, FTIR spectrometer, optional oxygen and VOC analyzers, and a control and display unit. An air purification unit provides zero gas for the spectrometer and reference gas for the oxygen analyzer.
Operation: The system operates automatically with software that displays results and status messages. It supports both local and remote operation, self-diagnosis, and automatic correction for dry/wet basis measurements. FTIR results are updated approximately every 30 seconds, and the system includes remote diagnosis and preventive maintenance features.
Technical Data: The system measures various components with specific ranges and performance metrics. It features a free-standing cabinet design with optional air conditioning, IP54 protection class, and various input/output signal options. It requires specific power supply conditions and is designed for indoor use under certain environmental conditions.
Certification: The ACF5000 is certified according to EN 15267 and EN 14181, making it suitable for European Directives and compliant with US EPA standards.
See more

Catalog excerpts

ACF5000-1

ACF5000 Multi-component FTIR emission monitoring system

 Open the catalog to page 1
ACF5000-2

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J • Continuous, quantitative and selective measurement of HCl, HF, H2O, CO, CO2, SO2, NO, NO2, CH4, NH3, N2O, H2CO, O2 and VOC (other gases on request) • Maximum 15 measuring components (standard), simple upgrade on request • Proven hot wet extractive measurement technique • High stability, accuracy and reliability through proven FTIR technology • Fully integrated VOC and O2 analyzers (optional) • Unique air-driven injector pump, no moving parts, low condensate to handle • QAL3 automatic span drift check without test...

 Open the catalog to page 2
ACF5000-3

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J System description Application and design As a result of the growing requirements in the field of environmental monitoring, increasing number of pollutants and with lower concentrations has to be measured from combustion processes. World leader in stack gas monitoring systems for decades and pioneer in FTIR technology, ABB Analytics is offering an inexpensive and forward-looking system with the ACF5000 multi-component FTIR emission monitoring system. Recognized by the process industries for their ruggedness, the ABB...

 Open the catalog to page 3
ACF5000-4

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J — … System description Operation The software installed in the system controller operates the analysis system completely automatically. It allows the following functions: • Display of all measured results and clear status messages • Manual operation of the system for commissioning and service • Local operation for service purposes via Ethernet interface • Remote diagnosis via UMTS • Self-diagnosis of the FTIR spectrometer and archival of the status signal and measured data • Optional automatic correction for dry/wet...

 Open the catalog to page 4
ACF5000-5

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J Measured components and measuring ranges Lowest ranges 0 to 40 vol.% 0 to 30 vol.% 0 to 75 mg/m3 0 to 60 ppm 1) FTIR spectrometer performance is based on an optical path length of 3.2 m (10.5 ft) 2) VOC measured with FTIR (for process measurement only) is composed of the organic compounds methane (CH4), ethane (C2H6), propane (C3H8), benzene (C6H6), toluene (C7H8) and dichloromethane (CH2O2). Measuring ranges within ignition limits cannot be provided. Other measured components and measuring ranges on request. Performance...

 Open the catalog to page 5
ACF5000-6

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J ... Technical data Power supply Input voltage 230/400 V 3 Ph, N, PE or 120/208 V 3 Ph, N, PE or 100/200 V 3 Ph, N, PE (optional, with transformer), ± 10 %, 48 to 62 Hz. Non-floating PEN conductor not allowed. Power consumption System incl. probe filter at power-up during operation Heated probe tube type 42 Heated sampling line TBL01 Air conditioning unit (optional) Prepared for UPS (optional) approx. 2200 VA approx. 1500 VA approx. 800 VA approx. 90 VA/m approx. 1000 VA approx. 500 VA Overvoltage category/pollution...

 Open the catalog to page 6
ACF5000-7

CABLE GLANDS SIGNALS MAIN SWITCH MAIN SWITCH POWER SUPPLY UPS POWER SUPPLY SAMPLE GAS OUTLET BURNER GAS FID INSTRUMENT AIR IN PURGE GAS TO PROBE ZERO GAS O2 CABLE GLANDS POWER SUPPLY SYSTEM CONTROLLER HEATED SAMPLE GAS LINE ELECTRICAL INLETS / OUTLETS GAS INLETS / GAS OUTLETS RIGHT SIDE LEFT SIDE ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J 7 System design: Exterior view COOLING UNIT

 Open the catalog to page 7
ACF5000-8

LEFT SIDE WALL, INSIDE DOOR OPENED ETHERNET SWITCH, TERMINALS CONTACTORS, RELAYS, CIRCUIT BREAKER NORMAL NET ANALOG SIGNALS, STATUS SIGNALS NEEDLE VALVES, PRESSURE CONTROLLER, COCKS, FLOW METER CIRCUIT BREAKER UPS, FUSES, TERMINALS POWER SUPPLY 24VDC, TERMINALS RIGHT SIDE WALL, INSIDE 8 ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J System design: Interior view

 Open the catalog to page 8
ACF5000-9

PROBE TUBE HEATED SAMPLE GAS LINE TBL01 HEATED FILTER UNIT PFE2 SAMPLE POINT HEATED FTIR CELL FTIR SPECTROMETER ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J Pneumatics diagram

 Open the catalog to page 9
ACF5000-10

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J Electrical connections: Power supply, CE version ANALYSIS CABINET CE VERSION, FUSES UNIPOLAR POWER SUPPLY (BY CUSTOMER) FUSE-PROTECTION EXTERNAL 3x 20 A CABLE 5x 6 mm2 (3x 25A) CABLE GLAND M32 (CABLE DIA. RANGE 7-15mm) UPS POWER SUPPLY (BY CUSTOMER) FUSE-PROTECTION EXTERNAL 20A CABLE 3x 2,5 mm2 CABLE GLAND M25 (CABLE DIA. RANGE 6-13mm) SAMPLE HEATER SAMPLE TUBE TYPE 42(-E84/-E85) CABLE 3x 1,5 mm2 CABLE GLAND M25 (CABLE DIA. RANGE 6-13mm) CABLE 3x 0,75 mm2 CABLE GLAND M20 (CABLE DIA. RANGE 5-10mm) CONNECTION VIA...

 Open the catalog to page 10
ACF5000-11

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J Electrical connections: Power supply, CSA version ANALYSIS CABINET CSA VERSION, FUSES UNIPOLAR POWER SUPPLY (BY CUSTOMER) FUSE-PROTECTION EXTERNAL 3x25A CABLE 5xAWG8 CABLE GLAND M32 (CABLE DIA. RANGE 0.28-0.59") UPS POWER SUPPLY (BY CUSTOMER) FUSE-PROTECTION EXTERNAL 20A CABLE 3xAWG14 CABLE GLAND M25 (CABLE DIA. RANGE 0.24-0.51") SAMPLE HEATER SAMPLE TUBE TYPE 42(-E84/-E85) CABLE 3xAWG16 CABLE GLAND M25 (CABLE DIA. RANGE 0.24-0.51") CABLE 3xAWG20 CABLE GLAND M20 (CABLE DIA. RANGE 0.2-0.39") SAMPLE HEATER FILTER UNIT...

 Open the catalog to page 11
ACF5000-12

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J — Electrical connections: Input, output and status signals -D80.1 Shield Analog outputs 4-20 mA Load max. 600 Ω Digital relay outputs max. 277V AC/DC max. 5A per channel COM 1 System maintenance request COM 2 Digital relay outputs System maintenance mode O2 failure COM 3 -D86 FID/ASP block failure COM 0 Digital relay outputs FTIR failur

 Open the catalog to page 12
ACF5000-13

ACF5000 MULTI-COMPONENT FTIR EMISSION MONITORING SYSTEM | DS/ACF5000-EN REV. J Digital relay outputs max. 277V AC/DC max. 5A per channel -D87 Meas. range feedback meas. comp. 1 COM 0 COM 1 Meas. range feedback meas. comp. 3 COM 2 Digital relay outputs Meas. range feedback meas. comp. 2 Meas. range feedback meas. comp. 4 COM 3 -D88 Meas. range feedback meas. comp. 5 COM 0 COM 1 Meas. range feedback meas. comp. 7 COM 2 Digital relay outputs Meas. range feedback meas. comp. 6 Meas. range feedback meas. comp. 8 COM 3 -D91 Drift check: Port 1 COM 0 COM 1 Drift check: Port 3 COM 2 Digital relay outputs...

 Open the catalog to page 13

All ABB Measurement & Analytics catalogs and technical brochures

  1. ProKiln GAC400

    24  Pages

  2. AV1 and AV2

    24  Pages

  3. SCK

    20  Pages

  4. INDI-BOY DSP

    8  Pages

  5. 9QGPS4500

    8  Pages

  6. 9QGPS3100P

    12  Pages

  7. 9QGPS2960

    12  Pages

  8. 9QGPS2625

    8  Pages

  9. MB3600-HP10

    5  Pages

  10. MB3600-CH10

    8  Pages

  11. Magnos28

    10  Pages

  12. EL3060

    8  Pages

  13. HP30

    4  Pages

  14. Aztec ATS430

    12  Pages

  15. LMS200

    12  Pages

  16. PGC5000

    8  Pages

  17. LGR-ICOS 927S

    4  Pages

  18. MR Series

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.