Oxygen Only or Optional Oxygen andCombustibles Ą Rugged Industrial Design Easy Installation and Start-up Ą Highly Accurate Sensors Simplified In-house Maintenance Ą Diagnostics and Alarming Patented Close Coupled Sample System Ą Process Temperature Measurement Standard Combustion Efficiency Calculation Ą Automatic Sensor Calibration-Standard Calibration Verification Ą Distinctive Operator Interface Optional Filter Blowback > Ą High Temperature Operating Range Optional Dual Filtration System Ą Probe & Cable Length Options Inherent Flashback Protection >
Open the catalog to page 1
Analytical Instruments Smart Analyzer 90 - Type SMAD-NO-SMA_2 The Type SMA Smart Analyzer 90 is a combustionanalyzer. It continuously samples and analyzes industrial flue gases. The Type SMA1 Smart Analyzer 90 monitors oxygen O The sensor assembly flange mounts to a duct or process wall so that the probe with filter assembly extends into the flue gas stream. The probe with filter assembly continuously draws in the sample which is then analyzed by the sensor assembly. Electrical outputs from the sensors are fed through the interconnecting flexible cableto the electronics assembly for...
Open the catalog to page 2
Analytical Instruments Smart Analyzer 90 - Type SMAD-NO-SMA_2 After the process gas is analyzed by the sensors, it iscombined with the aspirating air and is sent back into the duct through the aspirator exhaust filter. Electrical outputs from the sensors are fed through the intercon- necting flexible cable to the electronics assembly forinterpretation. > The NEMA 4 electronics assembly provides the micro-processor based architecture and overall intelligence necessary to monitor sensor calibrations, field input/ output (I/O) signals and the operator interface. The electronics assembly...
Open the catalog to page 3
5 12.00304.80 MIN CLEARANCE REQUIREDFOR SENSOR ASSEMBLYINSTALLATION/REMOVAL > 8 ASPIRATORSUCTIONPRESSUREPORT 5 TEST GAS PORT 4.50114.301.3133.381.6642.06 > PROBE 7 FILTER 13.13333.50 AIR SUPPLYPORT 207 kPa(30 PSIG) 5.00127.00 > smartanalyzer HOTSURFACE CAUTION EXHAUST FILTER 2.9474.63 ASPIRATOR 9.10231.14 > NOTES: 1ALL TUBE CONNECTIONS REQUIRE1/4 in. DIA. COPPER OR STAINLESSSTEEL TUBING.3 in. AND 4 in. OPTIONAL MOUNTINGREQUIRES ADAPTER FLANGE KIT PT. NO. 4.60116.80 7.00177.80 > MINIMUMCABLEBEND 2 > 2 in. MOUNTING IS STANDARD.THE OPTIONAL HIGH TEMPERATUREPROBE AND FILTER REQUIRE A3 in. OR 4...
Open the catalog to page 4
12.53318.26 1.9148.54 3.2582.552.9474.556.50165.10 7.14 mm (0.281 in.) WIDEX19.05 mm (0.750 in.)LNGMOUNTING SLOTS ZEROGAS INSPANGAS IN > VAR SELECT SETUP PREVSCREENNEXTSCREEN SYSTEMFAULTALARMACK CAL CONFIGENTER 17.45443.1018.00457.2018.75476.25 TESTGAS OUT TERMINATIONCHAMBER DOOR 4.38111.120.4310.79 NOMINAL VIEWING HEIGHT1500 mm (59 in.) FROM FLOORTO CENTERLINE OFDISPLAY 0.6516.56 2.1353.98 7.00177.80 MINIMUMCABLEBEND CABLE ASSEMBLY6638420_1_ 16.15410.08CLEARANCE REQDTO OPEN DOORCUSTOMER SIGNAL WIRINGCONDUIT 3/4 NPT FITTING(REMOVE PLUG BUTTONAS REQҒD), 3 PLACESPOWER WIRINGCONDUIT 1/2 NPT...
Open the catalog to page 6
Analytical Instruments Smart Analyzer 90 - Type SMAD-NO-SMA_2 > DXBX 222222 22
Open the catalog to page 7
Analytical Instruments Smart Analyzer 90 - Type SMAD-NO-SMA_2 Variable Analog Output Ranges O > 2 spanMinimum 0 to 5%Maximum 0 to 25%CO Measurement Errors Due to Flue Gas Variables CO > e spanMinimum 0 to 200 ppm(0.00 to 0.02%) Maximum 0 to 20000 ppm (0.00 to 2.00%)Temperature Zero-46 to 1371ںC(-50 to 2500ںF)Temperature spanMinimum 260C (500ںF)Maximum 1649C(3000ںF) NoneWater VaporNoneParticulateNone Display Screen Accuracy > 2 O > 2 2.5% or reading (0.1 to 25.0%)CO Sensor Response Time to 63% of Span O > 2 <3.5 secsCO > e ѱ 20 ppm (from 200 to 1000 ppm) 2% of span (from 1000 to 20000...
Open the catalog to page 8
Cable Standard length15 m (50 ft)Optional lengths30 m (100 ft) or46 m (150 ft)Minimum bend radius17.8 cm (7 in.)Maximum diameter1.9 cm (0.75 in.)(outside diameter) > Analytical Instruments Smart Analyzer 90 - Type SMAD-NO-SMA_2 Output Signals Four Analog > 1 1 to 5 VDC or4 to 20 mAAO1: Process O Environmental Sensor assemblyNEMA 4(indoor/outdoor)Electronics assemblyNEMA 4 (indoor) > 2 2 (isolated or non-isolated)AO2: Process CO > e AO3: Inlet/outlet temperature AO4: Combustion efficiencySix isolated digital Maximum Probe w/Filter Assembly Temperature Standard probe with filter649C...