1. Catalogs
  2. Aanderaa
  3. Oxygen Optode 4835
video corpo

Oxygen Optode 4835

Oxygen Optode 4835

Oxygen Optode 4835

Product catalog summary
Overview: The Oxygen Optode 4835 is a compact sensor designed for measuring oxygen concentration in shallow water environments. It utilizes an optical measurement principle based on lifetime-based luminescence quenching, offering long-term stability and low maintenance.
Key Features:
  • Optical measurement with luminescence quenching.
  • Stable for over a year without recalibration.
  • Low maintenance and user-friendly.
  • Compatible with Aanderaa SmartGuard/SeaGuard systems.
  • Can function as a stand-alone sensor.
  • Output formats include CANbus AiCaP and RS232.
  • Operational depth range of 0-300 meters.
Importance of Oxygen Measurement: Oxygen is crucial for biological and chemical processes in aquatic environments and serves as a tracer in oceanographic studies. Monitoring is essential in areas with limited oxygen supply, such as coastal areas with algae blooms, fjords, fish farms, and waste dumping sites.
Technical Specifications:
  • Measurement Range: 0-500 µM for O2 concentration, 0-150% for air saturation.
  • Resolution: <1 µM for O2, 0.4% for air saturation.
  • Accuracy: <8 µM or 5% for O2, <5% for air saturation.
  • Temperature Range: -5 to +40°C with 0.01°C resolution.
  • Response Time: <25 seconds for O2, <10 seconds for temperature.
  • Sampling Interval: 2 seconds to 255 minutes.
  • Supply Voltage: 5 to 14Vdc.
  • Dimensions: Ø36 x 86mm, Weight: 118g.
  • Materials: Titanium, Hostaform (POM).
Operating Principle: The sensor uses modulated blue light to excite the sensing foil, measuring the phase of returned red light. A reference phase reading is performed using a red LED. A thermistor provides temperature compensation for accurate O2 concentration readings.
Connectivity and Accessories: The sensor connects via a 10-pin receptacle and offers various cable options for different setups. Accessories include a standard foil service kit and various cables for PC and data logger connections.
Additional Information: For the latest updates and more details, visit the Aanderaa website. Aanderaa is a trademark of Xylem Inc.
See more

Catalog excerpts

Oxygen Optode 4835-1

Oxygen Optode 4835 is a compact fully integrated sensor for measuring the O2-concentration in shallow water. Advantages: • Optical measurement principle • Lifetime-based luminescence quenching principle • Long time stability • More than one year without recalibration • Low maintenance needs • User friendly • Use with Aanderaa SmartGuard/SeaGuard • Automatically detected and recognized • Use as stand-alone sensor • Output format: CANbus AiCaP, RS232 • Operating range: 0-300 meters Since oxygen is involved in most of the biological and chemical processes in aquatic environments, it is one of the most important parameters needed to be measured. Oxygen can also be used as a tracer in oceanographic studies. For environmental reasons it is critical to monitor oxygen in areas where the supply of oxygen is limited compared to demand, e.g. • In shallow coastal areas with significant algae blooms • In fjords or other areas with limited exchange of water • Around fish farms • Areas of interest for dumping of mine or dredging waste The Aanderaa oxygen optodes are based on the ability of selected substances to act as dynamic fluorescence quenchers. The fluorescent indicator is a special platinum porphyrin complex embedded in a gas permeable foil that is exposed to the surrounding water. A black optical isolation coating protects the complex from sunlight and fluorescent particles in the water. This sensing foil is attached to a sapphire window providing optical access for the measuring system from inside a watertight housing. The lifetime-based luminescence quenching principle offers the following advantages over electro-chemical sensors: • Not stirring sensitive (it consumes no oxygen) • Less affected by fouling • Measures absolute oxygen concentrations without repeated calibrations • Better long-term stability • Less affected by pressure • Pressure behaviour is predictable • Faster response time The oxygen optode outputs data in AiCaP CANbus and RS-232. The sensor can present the O2 concentration in µM, the Air Saturation in % and the Temperature in °C. The SmartGuard and SeaGuard data logger and the smart sensor are interfaced by means of a reliable CANbus interface (AiCaP), using XML for plug and play capabilities.

 Open the catalog to page 1
Oxygen Optode 4835-2

Securing plate Sensing foil Sapphire window Analog board Digital board Oxygen: O2-Concentration Air Saturation Measurement Range: 0 – 500 µM1) 0 - 150% Resolution: < 1 µM 0.4 % Accuracy: <8 µM or 5%2) <5 %3) whichever is greater Response Time (63%): <25 sec Temperature: Range: -5 to +40°C (23 - 104˚F) Resolution: 0.01°C (0.018°F) Accuracy: ±0.1°C (0.18°F) 4) Response Time (63%): <10 sec Output format: AiCaP CANbus, RS-232 Output parameters: O-ring 13.1x16 Sensor foot PIN CONFIGURATION Receptacle, exterior view; pin = • bushing = ° CAN_H 5 4 NCE 3 6 NCG Do not use 9 10 NCR CAN_L 2 7 Gnd RS232...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.