1. Catalogs
  2. Aanderaa
  3. Oxygen Optode 4831/4831F
video corpo

Oxygen Optode 4831/4831F

Oxygen Optode 4831/4831F

Oxygen Optode 4831/4831F

Product catalog summary
Introduction
Oxygen is a critical parameter in aquatic environments due to its involvement in biological and chemical processes. Monitoring oxygen levels is essential, especially in areas with limited oxygen supply compared to demand, such as shallow coastal areas with algae blooms, fjords, fish farms, and areas of waste dumping.

Oxygen Optode Technology
The Aanderaa oxygen optodes utilize dynamic fluorescence quenching, with a platinum porphyrin complex in a gas-permeable foil. This technology offers advantages over electro-chemical sensors, including no stirring sensitivity, reduced fouling effects, absolute oxygen concentration measurement without frequent calibration, better long-term stability, predictable pressure behavior, and faster response times.

Specifications
The Oxygen Optode 4831/4831F is a compact sensor for measuring O2 concentration. It features:
  • Optical measurement principle
  • Lifetime-based luminescence quenching
  • Long-term stability with over a year without recalibration
  • Low maintenance
  • Small size and weight
  • Stand-alone operation
  • Output formats: RS-232, 0-5V
  • Three depth ranges up to 6000 meters

Operating Principle
The sensor uses modulated blue light to excite the sensing foil and measures the phase of returned red light. A reference phase reading is performed using a red LED. A temperature thermistor allows for linearization and temperature compensation.

Measurement Capabilities
  • O2 Concentration: 0 – 500 µM
  • Air Saturation: 0 - 150%
  • Temperature Range: -5 to +40°C
  • Resolution and Accuracy: High precision with specific values for each parameter
  • Response Time: Fast response foil <8 sec, standard foil <25 sec

Output and Power
  • Output Parameters: O2 Concentration, Air Saturation, Temperature
  • Sampling Interval: 2s – 255 minutes
  • Supply Voltage: 5 to 14Vdc
  • Current Drain: Average and maximum values provided

Design and Materials
  • Dimensions: Ø36 x 111.5mm
  • Weight: 217g
  • Materials: Epoxy coated Titanium, PA

Sensing Foil Considerations
The standard sensing foil is rugged and sunlight-insensitive, while the fast response foil lacks this protection and requires careful handling to avoid erroneous readings due to ambient light.

Contact Information
Aanderaa Data Instruments AS, Bergen, Norway. Visit their website for more information.
See more

Catalog excerpts

Oxygen Optode 4831/4831F-1

Oxygen Optode 4831/4831F is a compact fully integrated sensor for measuring the O2 concentration. Fast Response Foil (4831F, refer overleaf) Advantages: • Optical measurement principle • Lifetime-based luminescence quenching principle • Long time stability • More than one year without recalibration • Low maintenance needs • Small size and weight • Stand-alone sensor • Output format: RS232, 0-5V • Three depth ranges maximum 6000 meter Since oxygen is involved in most of the biological and chemical processes in aquatic environments, it is one of the most important parameters to be measured. Oxygen can also be used as a tracer in oceanographic studies. For environmental reasons it is critical to monitor oxygen in areas where the supply of oxygen is limited compared to demand e.g. • In shallow coastal areas with significant algae blooms • In fjords or other areas with limited exchange of water • Around fish farms • Areas of interest for dumping of mine or dredging waste The Aanderaa oxygen optodes are based on the ability of selected substances to act as dynamic fluorescence quenchers. The fluorescent indicator is a special platinum porphyrin complex embedded in a gas permeable foil that is exposed to the surrounding water. For the standard version 4831 a black optical isolation coating protects the complex from sunlight and fluorescent particles in the water. This sensing foil is mounted on a sapphire window providing optical sampling from inside a watertight housing. The lifetime-based luminescence quenching principle offers the following advantages over electro-chemical sensors: • Not stirring sensitive (it consumes no oxygen) • Less affected by fouling • Measures absolute oxygen concentration without repeated calibrations • Better long-term stability • Less affected by pressure • Pressure behaviour is predictable • Faster response time The oxygen optode outputs data in RS-232 and analog 0-5V. The sensor can present the O2 concentration in µM, Air Saturation in % and Temperature in °C.

 Open the catalog to page 1
Oxygen Optode 4831/4831F-2

Securing plate Sapphire window Thermistor LED Analog board Digital board Interface board 111.5mm PIN CONFIGURATION SUBCONN MCBH8 V+ Gnd Analog SGnd 1 Oxygen: O2 Concentration Air Saturation 1) Measurement Range: 0 – 500 µM 0 - 150% Resolution: <1 µM 0.4 % 2) 3) Accuracy: <8 µM or 5% <5 % whichever is greater Response Time (63%): 4831F (with fast response foil) <8 sec 4831 (with standard foil) <25 sec Temperature: Range: -5 to +40°C (23 - 104˚F) Resolution: 0.01°C (0.018°F) 4) Accuracy: ±0.03°C (0.18°F) Response Time (63%): <2 sec Output format: RS-232, 0-5V Output Parameters: RS-232: O2 Concentration...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.