Overview: The Turbidity Sensor 4296 is a compact, fully integrated sensor designed for measuring optical backscatter in water. It is compatible with SeaGuard or SmartGuard data loggers using AiCaP CANbus or can function as a stand-alone sensor using RS-232.
Advantages:
  • Smart Sensor for easy integration with SeaGuard and SmartGuard.
  • Direct readout of engineering data, multipoint factory calibrated.
  • Capability to enter site-specific reference data for absolute values in mg/l.
  • Low power consumption and robust design with minimal maintenance needs.
  • Output formats include AiCaP CANbus and RS-232.
  • Available in three depth ranges: 300m, 3000m, and 6000m.
Technical Specifications:
  • Turbidity: Ranges from 0-25 FTU to 0-2500 FTU, with a resolution of 0.1% of reading or 0.025 FTU and accuracy of ±3% of range. Wavelength is 880 nm with a scattering angle of 150° and aperture of 15°.
  • Temperature: Range from -5 to 40°C, resolution of 0.001°C, accuracy of ±0.15°C, and response time of less than 8 seconds.
  • Output Parameters: Turbidity, Temperature, Raw data.
  • Sampling Interval: 2 seconds to 255 minutes.
  • Supply Voltage: 5 to 14VDC.
  • Current Drain: Average: 0.22 + 40mA/S, Maximum: 100mA, Quiescent: 0.22mA.
  • Operating Depth: Shallow Water: 0-300m, Intermediate Water: 0-3000m, Deep Water: 0-6000m.
  • Dimensions and Weight: 36 x 40.5 x 86mm, 185g.
  • Materials: Epoxy coated Titanium, Sapphire glass window, POM.
Applications: Suitable for coastal and deep water monitoring, aquaculture, sediment transport, dredging, river input, coastal erosion, offshore gas/oil and mining, seismic events, and more.
Additional Information: Specifications are subject to change without prior notice. Accessories such as sensor cables and setup cables are not included.
See more

Catalog excerpts

4296-1

Turbidity Sensor 4296 is a compact fully integrated sensor for measuring optical backscatter in water. It is designed to be used with SeaGuard or SmartGuard data logger using AiCaP CANbus or as stand-alone sensor using RS-232 Advantages: • • • • • • • Smart Sensor for easy integration with SeaGuard and SmartGuard Direct read out of engineering data, multipoint factory calibrated Enter site specific reference data to receive absolute values in mg/l Low power consumption Rugged and robust with low maintenance needs Output format AiCaP CANbus, RS-232 3 depth ranges available, 300, 3000 and 6000 meters Turbidity is the optical property of the water that causes light to be scattered by suspended particles. High particle concentration causes high turbidity. By establishing the relationship between turbidity and the local suspended matter this measurement can be used for assessing the concentration of suspended matter. The Turbidity Sensor 4296 measures the backscatter of infrared light. A high efficiency light emitting diode is used to transmit modulated light into the water. The backscattered light is picked up by a sensitive photodiode and this signal is then conditioned, linearized and converted to data in engineering unit (FTU). In addition to turbidity, the sensor also measures water t

 Open the catalog to page 1
4296-2

Range: Technical 4296,4296R 4296A,4296RA Turbidity: 4296B,4296RB Range: 4296C,4296RC Scattering angle: Resolution: Aperture: Temperature: Wavelength: Range: Scattering angle: Resolution: Aperture: Accuracy: 0.1% of reading or 0.025 FTU ±3% of range. Multipoint calibrated for each range 880 nm Output Parameters: Turbidity, Temperature, Sampling interval: Output format: Raw data 4296, 4296A, 4296B, 4296C AiCaP CANbus and RS-232 Output Parameters: Supply voltage: Current drain: interval: Sampling Average: Supply voltage: Current Maximum: drain: Quiescent: Average: Maximum: Operating depth: Quiescent:...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.