カタログの抜粋

MULTICOMPONENT ANALYZER SYSTEMS CEMS Solution Sensor Intelligence.
カタログの1ページ目を開く
CONTINUOUS, EXTRACTIVE FLUE GAS MONITORING On MCS100E HW, from sampling to the cell, all components that are in contact with the sample gas are heated to above dew point and thus protected from corrosion. On MCS100E CD/PD, gas drying is performed via a cooler/permeation dryer. The sample gas pump is located in the MCS100E system cabinet. Fast sample gas exchange minimizes adsorption or desorption effects, especially of HCl and NH3. In case of a malfunction, the system is purged with zero gas and thus protected from corrosion. During span gas feeding on the sampling probe, the complete...
カタログの2ページ目を開く
Heated sample gas line Bifrequency and gas filter correlation principle The single beam infrared filter photometer of the analyzer allows the simultaneous use of bifrequency and gas filter correlation methods. The cell is optimized for fast gas exchange and thermostatically controlled to high temperatures. A sintered metal protective filter is fitted in the sample gas inlet. An integrated flow meter triggers an alarm when the value is below the set limit value. Optionally, the analyzer can contain an oxygen measurement. The optional use of an internal calibration check allows fast checking...
カタログの3ページ目を開く
EMISSION AND RAW GAS MONITORING WITH HOT MEASUREMENT Product description The MCS100E HW is an analyzer system for extractive measurement of up to 8 IR-active gas compounds. It can be supplemented with oxygen and total hydrocarbon analyzers. From probe sampling to cell, all components that are in contact with the sample gas are heated above dew point and thus protected from corrosion. This hot measurement technique allows the measurement of water soluble components HCl and NH3. At a glance • Extractive measurement of up to 8 IR-active gas compounds • Additional oxygen and total hydro carbon...
カタログの4ページ目を開く
CEMS solutions MCS100E HW Fields of application • Emission monitoring of waste incineration plants • Monitoring of combustion plants with secondary fuels • Flue gas monitoring in cement plants • Emission monitoring of refineries • Emission monitoring of heavy oil fired diesel engines Detailed technical data The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications. MCS100E HW system Measured values Performance tested measurands...
カタログの5ページ目を開く
MCS100E HW CEMS solutions Conformities Electrical safety Enclosure rating Analog outputs 4 outputs: 0/4 ... 20 mA, 500 Ω Max. 64 outputs with opto box via optical fibre possible Analog inputs 4 inputs: 0/4 ... 20 mA Max. 64 inputs with opto box via optical fibre possible Digital outputs 12 outputs: 40 V, 1 A Max. 64 outputs with opto box via optical fibre possible Digital inputs 4 inputs: 50 V AC, 4 A / 24 V DC, 4 A Max. 64 inputs with opto box via optical fibre possible RS-232 RS-485 Modem Additional interfaces on request MODBUS Additional buses on request Via integrated operating unit Two...
カタログの6ページ目を開く
CEMS solutions MCS100E HW Ordering information Our regional sales organization will help you to select the optimum device configuration. Dimensional drawings (Dimensions in mm (inch)) MCS100E HW 806 (31.73) 735 (28.94) 8010576/2013-12 Subject to change without notice
カタログの7ページ目を開く
EMISSION MONITORING OF VERY SMALL MEASURING RANGES INCLUDING HCL Product description The MCS100E PD is an analyzer system for extractive measurement of up to 8 IR-active gas compounds. It can be supplemented with an oxygen sensor. The gas is dried by a permeation dryer, whereby HCl remains in the gas. Very small measuring ranges are possible, especially for SO2, NO, NO2 and HCl. At a glance • Extractive measurement of up to 8 IR-active gas compounds • Additional oxygen sensor as an option • Integrated permeation dryer • Test gas supply at the gas sampling probe or at the analyzer •...
カタログの8ページ目を開く
CEMS solutions MCS100E PD Fields of application • Flue gas monitoring in waste incineration plants • Clean gas monitoring in industrial plants • Emission monitoring in refineries Detailed technical data The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications. MCS100E PD system Measured values Performance tested measurands Measurement principles Interference filter correlation, Gas filter correlation, Zirconium dioxide sensor...
カタログの9ページ目を開く
MCS100E PD CEMS solutions Enclosure rating Analog outputs 4 outputs: 0/4 ... 20 mA, 500 Ω Max. 64 outputs with opto box via optical fibre possible Analog inputs 4 inputs: 0/4 ... 20 mA Max. 64 inputs with opto box via optical fibre possible Digital outputs 12 outputs: 40 V, 1 A Max. 64 outputs with opto box via optical fibre possible Digital inputs 4 inputs: 50 V AC, 4 A / 24 V DC, 4 A Max. 64 inputs with opto box via optical fibre possible RS-232 RS-485 Modem Additional interfaces on request MODBUS Additional buses on request Via integrated operating unit Two operating levels, one...
カタログの10ページ目を開く
CEMS solutions MCS100E PD Ordering information Our regional sales organization will help you to select the optimum device configuration. Dimensional drawings (Dimensions in mm (inch)) MCS100E PD 806 (31.73) 735 (28.94) 8010576/2013-12 Subject to change without notice
カタログの11ページ目を開く
EMISSION MONITORING OF VERY SMALL MEASURING RANGES Product description The MCS100E CD is an analyzer system for extractive measurement of up to 8 IR-active gas compounds. It can be supplemented with an oxygen sensor. The gas is dried by an integrated gas cooler allowing measurement of very small measuring ranges, especially for SO2, NO and NO2. At a glance • Extractive measurement of up to 8 IR-active gas compounds • Optional oxygen sensor available • Integrated gas cooler • Test gas supply at the gas sampling probe or at the analyzer • Back-purging of gas sampling probe for cleaning of...
カタログの12ページ目を開くSICKのすべてのカタログと技術パンフレット
-
W4F en
17 ページ
-
SLG-2 en
14 ページ
-
IMS en
12 ページ
-
IO-Link Supplement
5 ページ
-
T-Easic FTS
56 ページ
-
T-Easic® FTS
12 ページ
-
T-Easic® FTS
4 ページ
-
FLOWSIC100 PROCESS
134 ページ
-
SPEETEC 1D
6 ページ
-
SPEETEC
2 ページ
-
midiCam
5 ページ
-
W4-3 PTFE
4 ページ
-
SMART SENSORS
48 ページ
-
AHS/AHM36 SSI Inox
29 ページ
-
STEEL INDUSTRY
8 ページ
-
MultiPulse
37 ページ
-
EDS/EDM35
12 ページ
-
TR110 Lock
20 ページ
-
24-hour helpdesk
8 ページ
-
outdoorScan3
20 ページ
-
EnforceKey Single Door
84 ページ
-
WTB16 Bluetooth®
135 ページ
-
PET Pressure transmitter
44 ページ
-
LFV 310/330
48 ページ
-
WT18X-3P930
2 ページ
-
Proximity Sensors
20 ページ
-
Registration Sensors
24 ページ
-
QAL3 Master
4 ページ
-
Dust measuring devices
171 ページ
-
Automation Light Grids
228 ページ
-
Raw Gas Measurement
2 ページ
-
GR18 product information
56 ページ
-
Top Products from SICK
486 ページ
-
SICK Essentials
118 ページ
-
Magnetic Cylinder Sensors
164 ページ
-
Presence Detection
170 ページ
-
Industrial Safety Systems
1054 ページ
-
Fluid Sensors
174 ページ
-
Encoders
388 ページ
-
Distance Sensors
404 ページ
-
Identification Solutions
308 ページ
-
Zone control
5 ページ
-
fiber-optic sensors
5 ページ
-
PACKAGING2009/10
20 ページ
カタログアーカイブ
-
3D cameras
6 ページ
-
FLOWSIC 600
4 ページ