2461-EC Graphical Potentiostat
1 / 16ページ

{{requestButtons}}

カタログの抜粋

2461-EC Graphical Potentiostat - 1

2461-EC Graphical PotentiostatDatasheet KEITH LEY Key Features • Perform Cyclic, Squarewave, Galvanic Voltammetry, Chronoamperometry, and Chronopotentiometry • Simplified user interface for faster test setup and analysis of results • Real-time plotting of voltammograms on the front panel • Analytical graph cursors for immediate analysis of results current without the need for a PC • Create libraries of reusable, customizable experimental software with built-in scripting The 2461-EC Graphical Potentiostat brings speed, flexibility, and simplicity right to your fingertips. Its innovative graphical user interface (GUI) and advanced, capacitive touchscreen technology allow intuitive usage and minimize the learning curve to enable researchers, scientists, and students to learn faster, work smarter, and invent easier. The 2461-EC is a versatile instrument, particularly well-suited for research and development in fundamental electrochemical lab research, characterizing the next generation of materials and electrolytes, new energy storage devices, and faster, smaller sensors. • Screen capture function allows copying test results from the display to reports • Wide coverage up to 105 V, 7 A DC/7 A pulse, 100 W max. • Front panel input banana jacks; rear panel input mass termination screw connections • Context-sensitive help function minimizes learning • Front-panel USB 2.0 memory I/O port for transferring data, test scripts, or test configurations While potentiostats are excellent instruments for electrochemistry applications, they typically lack any front panel display and control knobs, often are 2-quadrant systems only, and must be completely controlled by a computer with software that is not always open for users to customize tests beyond what the software can do. The 2461-EC includes a wide range of voltages and currents for sourcing or measuring, pV/pA sensitivities, and high impedance sense leads with a typical input resistance of 50G ohms and only 1pA of input bias current, typically acceptable with a wide variety of reference electrodes. The 2461-EC can run internal application test scripts so electrochemistry measurements can be run without the use of an external computer. Results are immediately displayed right on the instrument front panel touchscreen. Connecting the 2461-EC to a 2-, 3-, or 4-electrode cell is simple with the included translation cable.

カタログの1ページ目を開く
2461-EC Graphical Potentiostat - 2

Electrochemical Cell 2461 SourceMeter® Working Electrode Reference Electrode Counter Electrode The 2461-EC can be easily connected to a 3-electrode cell. Learn Faster; Work Smarter; Invent Easier Unlike traditional potentiostats that lack a user-interface front panel to interact with, the 2461-EC features a fiveinch, full-color, high resolution touchscreen that facilitates ease of use, and optimizes overall speed and productivity. Built-in, context-sensitive help enables intuitive operation and minimizes the need to review a separate manual. These capabilities combined with its application...

カタログの2ページ目を開く
2461-EC Graphical Potentiostat - 3

Built-in real-time graphing, charting, scope-like cursors, and data display spreadsheet for export simplifies converting test results into useful information. Test Applications The 2461-EC’s built-in scripting language enables electrochemists, chemists, and materials scientists to create libraries of reusable, customizable experimental software for running tests including cyclic voltammetry, chronoamperometry, chronopotentiometry, and more. The following electrochemistry test scripts are loaded in the internal memory of the 2461-EC. • Cyclic Voltammetry: Potential is swept at a...

カタログの3ページ目を開く
2461-EC Graphical Potentiostat - 4

All-in-One Instrument The 2461-EC offers a highly flexible, four-quadrant voltage and current source/load coupled with precision voltage and current meters. When not used in potentiostat type applications, this all-in one instrument can be repurposed as a general lab instrument, including use as a: • Precision power supply with V and I readback • True current source • Digital multimeter (DCV, DCI, ohms, and power with 6^-digit resolution) • Precision electronic load • Trigger controller Maximum Voltage and Current In addition to its five-inch, color touchscreen, the 2461-EC front panel has...

カタログの4ページ目を開く
2461-EC Graphical Potentiostat - 5

Mass screw terminal connections Rear panel connections are optimized for signal integrity. Free Instrument Control Start-up Software The 2461-EC can be repurposed for applications beyond electrochemistry as a general purpose lab tool, e.g. I-V testing, leakage testing, battery charge/discharge profiling, etc. KickStart, Keithley’s instrument control non-programming start-up software, lets users start taking measurements in minutes for typical current versus voltage applications. In most cases, users merely need to make quick measurements, graph the data, and store the data to disk to...

カタログの5ページ目を開く
2461-EC Graphical Potentiostat - 6

Current Measure Ranges 1 mA, 10 mA, 100 mA, 1 A, 4 A, 5 A, 7 A Source Limit (Compliance) 100% of selected current range Potential Step Size During Scanning 100 |jV (0.1 mV/s < scan rate < 35 mV/s) 1 mV (35 mV/s < scan rate < 350 mV/s ) 10 mV (350 mV/s < scan rate < 3500 mV/s) Number of Cycles 1 to 100 User Selectable Sampling Intervals Points/Test (10 to 10000) Points/Cycle (10 to 10000) Seconds/Point (0.01 to 100) Points/Second (0.01 to 100) Maximum Total Number of Samples 100,000 Potential Measure Ranges 200 mV, 2 V, 7 V, 10 V, 20 V Sample Interval 0.75 s < sample interval...

カタログの6ページ目を開く
2461-EC Graphical Potentiostat - 7

2461-EC Current Pulse and Square Wave Specifications Peak Current -7.35 A < Ipeak < +7.35 A Base Current -7.35A < Ipeak < +7.35A Potential Measure Ranges 200 mV, 2 V, 7 V, 10 V, 20 V Source Limit (Compliance) 105% of selected potential range Pulse Period and Width ilpeaki and ilbasei < 1.05 pA 80 ms < period < 3600 s 40 ms < pulse width < (0.99 x period) ilpeaki or ilbasei > 1.05 pA 10 ms < period < 3600 s 5 ms < pulse width < (0.99 x period) Sample Time Minimum is 166.667 gs. Maximum is lesser of 166.667 ms and (pulse width - 1 ms) Number of Cycles 1 to 100,000 Program...

カタログの7ページ目を開く

Keithley Instruments/ケースレーのすべてのカタログと技術パンフレット

  1. 6220-6221

    5 ページ

  2. 6482

    3 ページ

  3. 2520

    8 ページ

  4. 2606B

    13 ページ

  5. 2601B

    20 ページ

  6. AFG1000 Series

    13 ページ

  7. 6 Series B MSO

    69 ページ

  8. TBS1000B-EDU Series

    14 ページ

  9. 3 Series MDO

    36 ページ

  10. 4 Series MSO

    40 ページ

  11. TSG4100A Series

    24 ページ

  12. RTPA2A

    6 ページ

  13. 10G-KR Datasheet

    8 ページ

  14. 4200-SCS

    16 ページ

カタログアーカイブ