カタログの抜粋

CozIR®-Blink Ultra-Low Power CO2 Sensor DESCRIPTION The CozIR®-Blink is an ultra-low power NDIR CO2 sensor using state-of-the-art solid-state LED optical technology. The low power LEDs are manufactured in-house, giving GSS complete control of the CO2 sensor signal chain. The CozIR®-Blink uniquely allows users to reduce power consumption of the CO2 measurement to unprecedented levels. The CO2 module is designed to be power cycled when not in use. The sensor is designed to allow the user to wake up the unit from a dormant state and take a reading, and then power down it completely. This reduces the power consumption of the sensor to zero when dormant. The sensor is designed to be power cycled, where the sensor is only powered-up for the duration of the measurement, and then powered down again to save power for battery powered or wireless interface applications. The CozIR®-Blink also features a built-in autozero function that maintains CO2 measurement accuracy over the lifetime of the product. BLOCK DIAGRAM FEATURES • Ultra-low power CO2 sensor (60uW per reading, 60s, nPulse = 1) • 30ppm typical measurement accuracy • Solid state NDIR LED optical technology • UART or I2C control and data interface • Designed for power-cycling • Built-in auto-zeroing APPLICATIONS • Air Quality and HVAC • Building management systems (BMS) • Demand-Controlled Ventilation (DCV) systems • IoT and Smart Technology wireless equipment • Indoor Air Quality (IAQ) equipment Production Data, Revision 4.19, 10 May 2021 Page | 1 For regular updates, go to https://www.gassensing.co.uk Copyright © 2021 Gas Sensing Solutions Ltd.
カタログの1ページ目を開く
For regular updates, go to https://www.gassensing.co.uk Production Data, Revision 4.19, 10 May 2021 Copyright © 2021 Gas Sensing Sol
カタログの2ページ目を開く
Gas Sensing Solutions Ltd. Page | 3 For regular updates, go to https://www.gassensing.co.uk Production Data, Revision 4.19, 10 May 2021 Copyright © 2021 Gas Sensing
カタログの3ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor ORDERING INFORMATION COZIR-BLINK-X-X Notes: 1. Sensors are shipped individually or in trays 2. Tray quantity = 50 See separate data sheet for CozIR®-Blink evaluation kit options. Production Data, Revision 4.19, 10 May 2021 Page | 4 For regular updates, go to https://www.gassensing.co.uk Copyright © 2021 Gas Sensing Solutions Ltd.
カタログの4ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor PACKAGE DRAWING – WITH PINS Bottom View Gas Sensing Solutions Ltd. Page | 5 For regular updates, go to https://www.gassensing.co.uk Production Data, Revision 4.19, 10 May 2021 Copyright © 2021 Gas Sensing Solutions Ltd
カタログの5ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor VDD NAME TYPE GND Supply VDD Supply Rx_In Digital Input Tx_Out Digital Output NC VDD Unused VDD1 READY Digital Output NC Unused I2C_SDA NC Unused CozIR-Blink NC Unused No Pins I2C_SCL I2C_ENABLE Digital Input READY Digital Input I2C_ENABLE Digital Input/Output DESCRIPTION 4.7kΩ Sensor ground Sensor supply voltage Sensor Power Control UART Receive Input UART Transmit Output Do not connect Data ready pin. Pulsed high when data ready I2C MASTER Do not connect SDA Do not connect Do not connect SCL Set low for I2C interface mode. Leave floating to select...
カタログの6ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor PACKAGE DRAWING – NO PINS Top View Gas Sensing Solutions Ltd. Page | 7 For regular updates, go to https://www.gassensing.co.uk Production Data, Revision 4.19, 10 May 2021 Copyright © 2021 Gas Sensing Solutions Ltd
カタログの7ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor PIN-OUT DESCRIPTION – NO PINS PIN 1 2 3 4 5 6 7 8 TYPE Supply Supply Unused Supply Unused Digital Output Unused Digital Input Digital Input Tx_Out Rx_In Digital Input/Output Digital Output Digital Input DESCRIPTION Sensor ground Sensor supply voltage Do not connect (For internal use only) Sensor supply voltage Do not connect Data ready pin. Pulsed high when data ready Do not connect Set low for I2C interface mode. Leave floating to select UART interface mode. Pin status detected at power on. I2C serial clock input. Open drain, external 4.7kΩ resistor...
カタログの8ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor ABSOLUTE MAXIMUM RATINGS Absolute Maximum Ratings are stress ratings only. Permanent damage to the CozIR®-Blink may be caused by continuously operating at or beyond these limits. The CozIR®-Blink functional operating limits and guaranteed performance specifications are given at the test conditions specified. ESD Sensitive Device. This sensor uses ESD sensitive components. It is therefore generically susceptible to damage from excessive static voltages. Proper ESD precautions must be taken during handling and storage of this device. Production Data,...
カタログの9ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor PERFORMANCE CHARACTERISTICS Test Conditions Unless Otherwise Specified VDD/VDD1 = 3.3V, GND = 0V. CO2 = 450ppm, RH = 0% non-condensing, T= 25°C, Pressure = 1013mbar, nPulse = 16 Production Data, Revision 4.19, 10 May 2021 Page | 10 For regular updates, go to https://www.gassensing.co.uk Copyright © 2021 Gas Sensing Solutions Ltd.
カタログの10ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor POWER CONSUMPTION Test Conditions Unless Otherwise Specified VDD/VDD1 = 3.3V, GND = 0V. CO2 = 450ppm, RH = 0% non-condensing, T= 25°C, Pressure = 1013mbar SYMBOL TEST CONDITIONS OFF Active, SLEEP mode, no measurement The energy per reading is approximately 25mJ using default settings. Power consumption is strongly dependent on digital filter settings (# of pulses) and the measurement period. ްݟ. ϰݟӰݟݟаݟ Ӣȗ ݒϰݒְݒΰݒðݒưݒ Ӱݒаݒ ǰݒѰݒְݒͰݒݒưݒ ݒѰݒưݒ ӰݒӰݒưݒ°ݒŰݒʰݒϰݒ Ȣȗ ݟϰݟΰݟΰݟ ΰݑݒưݒ°ݒݒְݒӰݒưݒΰݒưݒϰݒ հݒѰݒưݒӰݒʰݒаݒ Űݒʰݒ ϰݑڰݒưݒİݒаݒϰݒŰݒ In a typical application...
カタログの11ページ目を開く
CozIR®-Blink Ultra-Low Power CO2 Sensor The graph below shows the trade-off between nPulse, the measurement period and the impact on average power consumption. Average Power Consumption Per Reading vs. nPulse Setting vs. Measurement Period ™ n=16 © n=8 © n=4 an n=2 ™ n=1 Production Data, Revision 4.19, 10 May 2021 Page | 12 For regular updates, go to https://www.gassensing.co.uk Copyright © 2021 Gas Sensing Solutions Ltd.
カタログの12ページ目を開くGas Sensing Solutions (GSS)のすべてのカタログと技術パンフレット
-
CO2 Sensor Response Time
12 ページ
-
Zeroing a CO2 Sensor
11 ページ
-
SprintIR-W Data Sheet
36 ページ
-
SprintIR-R Data Sheet
31 ページ
-
SprintIR-6S Data Sheet
31 ページ
-
ExplorIR-W Data Sheet
33 ページ
-
What is fresh Air
7 ページ
-
ExplorIR-M Data Sheet
28 ページ
-
CozIR-LP3 Data Sheet
53 ページ
-
CozIR-LP2 Data Sheet
40 ページ
-
CozIR-LP Data Sheet
29 ページ
-
CozIR-A Data Sheet
32 ページ