Product Data Sheet Opacity/Dust Density Monitor The OPM 4001 Opacity/Dust Density Monitor is a high performance opacity monitoring system with double-pass transmissometer that meets or exceeds ■ Digital display instant percent opacity, average percent opacity and time to next calibration ■ Alphanumeric display for system set-up and fault diagnostics display ■ 3 User selectable displays (instant opacity, average opacity and optical density) ■ U.S. Environmental Protection Agency (EPA) approved approved zero/span calibration ■ Stack exit correlation permanent factory setting ■ Sensor location service module digital display for percent correlated opacity, T2, manual zero and span initiate and transceiver current loop test ■ Insensitive to ambient light ■ Double pass optical system ■ Light source rated for greater than 5 years ■ Electronic modulated light source ■ Two 4-20mA signals for instantaneous opacity, ■ Calibration cycle control remote, manual and ■ System and fault diagnostics display ■ Path length of 3-50 feet Transmissometer/Cover assembly Standard Controller Process Management
カタログの1ページ目を開く
Product Data Sheet OPM 4001 Opacity/Dust Density Monitor Emerson’s Rosemount Analytical OPM 4001 high performance opacity monitoring system with double-pass transmissometer uses field-proven and time-tested optics and circuitry that is simple, yet accurate. Each monitor and all its components are of the highest quality; pre-tested at the factory to your site specifications. The simple instructions will have your monitor up and running in just a few hours. Enclosure Panel mounted IP65/NEMA4X Dimensions 96x96x64 mm (3.8˝x3.8´x2.5˝). Power 20.4 to 28.8VDC < 10% ripple, 400mA. Emerson has taken...
カタログの2ページ目を開く
Product Data Sheet Specifications (cont) Transceiver/Reflector Enclosure NEMA 4 watertight enclosure power Path Length 3-50 feet, 0.9 to 12 meters standard Optical System Double pass Calibration Drift < 0.5% Zero/Span Calibration Manual or automatic with zero mirror and neutral density filter Light Source Aging Compensation Automatic Light Source Life > 5 years Alignment Verification Passive built-in, through-the lens system Standard Mounting Flanges 3 inch IPS, 150 lb. flange, standard Ambient Light Immunity Solid-state electronic light modulation Design and Performance Ambient temperature...
カタログの3ページ目を開く
Product Data Sheet PDS 105-4001.A01 Electronic Display Options Both are EPA 40 CFR B PS-1 and ASTM D 6216 compliant OPM 4001 n Dual beam measurement Automatic or manual online calibration n User friendly microprocessor controller n 3 user selectable numercial displays Sensor located service module with a digital display for % correlated Opacity and %T2 Additional features with Enhanced Touch screen n Intuitively designed 5.7˝ color touch screen user interface with expanded diagnostics n 5 numerical and 2 trend display screens Standard SD card for program and data backup Automatic, manual or...
カタログの4ページ目を開く
Product Data Sheet Electronic Display Comparsion Chart w/Touch Screen Option ASTMK D 6216 and PS-1 compliant Communications via 4-20 mA and RS MODBUS Automatic or manual online calibration Microprocessor controller 5 numercial display screens Intuitively designed icon driven navigation Standard 4 analog outputs 5.7˝ color touch screen user interface Standard program includes % opacity, mg/m3 and O.D. Standard 2GB SD card for program and data backup Real-time diagnostics for testing outputs and relays Two selectable trend screen displays Easy to read color coded fault screen User display...
カタログの5ページ目を開く
Product Data Sheet mechanical installation SIDE VIEW Mrrum Swing ;tS?vt rrrr.) Oplcnitl WMthw *Note 7: The top view represents the transceiver and retro reflector assemblies with their swing clearances. Optional weather covers are not shown. * *Note 2: The side view represents the installation and swing clearance dimensions for the optional weather covers.
カタログの6ページ目を開く
Product Data Sheet Flange to stack installation Clearance Clearance MASONRY STACK METAL WALL STACK WALL INSTALLATION OR DUCT INSTALLATION
カタログの7ページ目を開く
Product Data Sheet Installation Dimensions OPM 4001 Standard Control Unit OPM 4001 Enhanced Control Unit
カタログの8ページ目を開く
Product Data Sheet Ordering information OPM 4001 Opacity/Dust Density Monitor - High performance opacity monitoring system with double-pass transmissometer. Model Product Description Opacity monitor Basic Unit - Digital Display, (2) 4-20mA Outputs, (6) Alarm Relays, RS 232/485, Modbus Enhanced Control Unit with 5.7” LCD graphic color touch screen, icon-driven menu with SD memory card Transceiver and Path Length 11 Weather Cover and Blower 00 Weather covers only Weather covers and single blower/Tee Weather covers and dual blowers Zero Jig and Neutral Density filters Opacity calculation Dust...
カタログの9ページ目を開く
Product Data Sheet PDS 105-4001.A01
カタログの10ページ目を開く
Product Data Sheet Ordering Information (con't) Choose the drawing below that best fits the installation. Mounting Flange
カタログの11ページ目を開く
Product Data Sheet Rosemount Analytical Headquarters & Liquid Analytical 2400 Barranca Parkway Irvine, CA 92606 United States Phone: 1.800.854.8257 Fax: 1.949.474.7250 Rosemount Analytical Gas Chromatographs & Service Center 10241 West Little York, Suite 200 Houston, Texas 77040 United States Phone: 1.866.422.3683 Fax: 1.713.827.3865 Rosemount Analytical Process Gas & Combustion Center 6565-P Davis Industrial Pkwy Solon, OH 44139 United States Phone: 1.800.433.6076 Fax: 1.440.914.1262 Rosemount Analytical – Latin America Emerson Process Management 10241 West Little York, Suite 200 Houston,...
カタログの12ページ目を開くEmerson Automation Solutions - ROSEMOUNTのすべてのカタログと技術パンフレット
-
FRL Accessories
14 ページ
-
TankRadar Pro
32 ページ
-
Roxar subsea PIG
2 ページ
-
Roxar subsea CM10K
2 ページ
-
Roxar PTPT15K
2 ページ
-
SW-300
4 ページ
-
DL8000
12 ページ
-
DeltaV Analyze
8 ページ
-
AMS Machine Works
4 ページ
-
Rosemount 975UR
4 ページ
-
Rosemount 975HR
4 ページ
-
Rosemount 975UF
4 ページ
-
Rosemount 975MR
4 ページ
-
Safety Monitoring
6 ページ
-
AMS Suite
16 ページ
-
Line Card
2 ページ
-
Drinking Water Industry
16 ページ
カタログアーカイブ
-
UVS Flame Detector
4 ページ