SPIRO ANALYZER GAS BENCH USERS MANUAL Specification spiro analyzer gas bench Wuhan Cubic Optoelectronics Co.Ltd ADD: Fenghuang Yuan No.3 Road, Fenghuang Industrial Park, Eastlake Hi-tech Development Zone, Wuhan 430205, China TEL: +86-27-81628831 FAX: +86-27-87401159 Http://www.gassensor.com.cn E-mail: info@gassensor.com.cn

カタログの1ページ目を開く
Specification of Micro-flow multigas analyzer CO and CH4 3000ppm - 2

I Foreword 1.Pls reading and understanding the content completely the operation manual before installation, operation and repairing the instrument .There would be happen to human injury or analyzer damage if your operation unqualified. 2.Never remove the configuration and spare parts of the instrument unless approving by company. Our company will be derelict of duty due to the user make bold to remove the configuration and spare parts mistakenly. 3.The specification and guide content of this analyzer may be changed due to the technology improve of the product. Pls forgive us shall not give...

カタログの2ページ目を開く
Specification of Micro-flow multigas analyzer CO and CH4 3000ppm - 3

SPIRO ANALYZER GAS BENCH USERS MANUAL 1.Abstract Spiro analyzer gas bench / module uses state-of-the-art micro-flow infrared gas sensor , measuring CH4/CO concentration with fast response ,high accuracy and long stability .The gas analysis module is designed of simple structure, easy operation,long lifespan, with power supply modules, RS-232 data communication lines It’s an ideal gas module for PFT(Pulmonary Function Test) system . ①-- Infrared light source ②-- Chopper ③—Chopper motor ④—Measurement cell ⑤-- Detector ⑥—Micro-flow sensor ⑦、⑧—The 2nd measurement cell and detector ⑨-- Signal...

カタログの3ページ目を開く
Specification of Micro-flow multigas analyzer CO and CH4 3000ppm - 4

SPIRO ANALYZER GAS BENCH USERS MANUAL ※ NoteMeasure range can be customized, the output voltage can be adjusted. Gas flow Pressure of gas inlet Requirement of sample gas Response time(TD+T90) Warm-Up time Output interface Work Temperature Relative Humidity Atmospheric Pressure Dimension 0.7 1.2 L/min <10KPa Remove water vapor,dust(<1um) and oil etc. <1S (NDIR) >30min Voltage signal output 050℃ 585% 86108kPa 360mm×140mm×74mmwidth×length×height 1.4 Function module of gas sensor Sensor signal Signal amplifier Main board Function module picture of gas sensor

カタログの4ページ目を開く
Specification of Micro-flow multigas analyzer CO and CH4 3000ppm - 5

SPIRO ANALYZER GAS BENCH USERS MANUAL 1.6 Definition of interface 1.6.1 Definition of main board interface

カタログの5ページ目を開く
Specification of Micro-flow multigas analyzer CO and CH4 3000ppm - 6

1.6.2 Definition of wiring terminal SN During transfer the data by RS232 cable ,pls first disconnect the power supply of equipment to prevent equipment damage.

カタログの6ページ目を開く
Specification of Micro-flow multigas analyzer CO and CH4 3000ppm - 7

SPIRO ANALYZER GAS BENCH USERS MANUAL 2. Installation 2.1 Using RS232 Users can get the sensor measurement result from serial port, the communication protocol command according to Annex. Workflow is as follows: 1. Warm-up: need warm-up for some time.the signal is unstable during this period. 2. Zeroing: inlet zero gasO2/N2 more than 10s after the sensor stability, Send zero command "11 01 03 EB" and complete zeroing, adjust the sensor measurements result to 0. 3. Take measurements result: send "11 01 01 ED" order, read the measurement results. Send “11 01 03 EB” order complete zeroing send...

カタログの7ページ目を開く
Specification of Micro-flow multigas analyzer CO and CH4 3000ppm - 8

SPIRO ANALYZER GAS BENCH USERS MANUAL Warm up half an hour After signal stability, send zero adjustment order “4B”,complete zeroing. RS232 zero adjustment Warm up half an hour After signal stability, send zero adjustment order “4C”,complete zeroing. RS232 span adjustment 3. Operation and Maintenance 3.1 Operation After finishing circuit and pneumatic routine of gas sensor module, power it on, then the module start to work. We suggest you can do the test after 10 to 15 minutes for warming-up. Open the pump of gas sensor module for sampling in order to absorbing the sample gas, screw the flow...