Technical Note External Use! 5-May-06 Minimizing Darkcurrent and Clock Induced Charge in EMCCDs for Photon Counting Performance Since EMCCD technology effectively overcomes readout noise, the remaining instrumental detection limit is set by background electrons that are amplified by the EM gain register. These events are generally manifest as single pixel spikes, amplified clear of the baseline, and are derived from two sources: Darkcurrent and Clock Induced Charge (CIC), the latter sometimes known as Spurious Noise. In EMCCD’s we must pay particular care to virtually eliminate darkcurrent contribution – now that single thermally generated electrons will be amplified above the read noise floor! It is no longer sufficient to apply only enough cooling such that darkcurrent is buried in the read noise (as is historically the case in non-EMCCDs). Clock Induced Charge In order to keep CIC to a minimum, clock voltages need to be carefully controlled. Andor’s extensive experience in EM technology has resulted in the lowest CIC on the market! A detector that has not minimized EACH of these noise sources will NOT be able to successfully discriminate photon events from dark background events. The iXon+ DU-897-BV is the only camera on the market capable of true photon counting. Technical Note In fact, we have included a specification on the iXon+ spec sheet that reflects the extraordinarily low darkcurrent and CIC levels achieved – EMCCD Amplified Background Events – and is quoted alongside the traditional measurement of darkcurrent (listed still for historical reasons, but they cannot be used to extrapolate to short exposure times). For example, the iXon+ DU-897-BV yields a typical EMCCD-Amplified Background Events figure of 0.005 electrons/pixel for a 30 ms exposure @ -85oC, EM Gain x1000. We confidently challenge other camera manufacturers to match this specification!

カタログの1ページ目を開く
Darkcurrent - 2

There is misunderstanding and misinformation in the market place about the required cooling for optimum operation of EMCCDs. In particular the claim that “ –30oC is sufficient cooling because Spurious Charge (also known as Clock Induced Charge, CIC) increases with cooling”. THIS IS SIMPLY NOT THE CASE! Experimental Verification The Physics does get involved but you can sidestep it by doing a simple test, the results of which are shown below. Figure 1 (over) shows three dark images taken with a Back Illuminated CCD97 chip (from E2V) at –90oC, –50oC and – 30oC, all are in frame transfer mode...

カタログの2ページ目を開く
Darkcurrent - 3

Figure 1: Three dark images at varying temperatures Figure 2: Three traces at varying temperatures Notes: all images were taken with a Back Illuminated CCD97 chip at –90oC, –50oC and –30oC, all are in FT mode at maximum frame rate (34.5 frames per second, i.e. around 29 ms exposure) and an EM gain (Electron Multiplication gain) of X1000. Technical Note

カタログの3ページ目を開く

Andor Technologyのすべてのカタログと技術パンフレット

  1. Marana sCMOS

    9 ページ

  2. MicroPoint 4

    9 ページ

  3. ZL41 Cell sCMOS

    7 ページ

  4. Sona sCMOS

    9 ページ

  5. Optistat

    11 ページ

  6. BC43

    13 ページ

  7. Solis-Brochure

    6 ページ

  8. iKon-M/L SO Series

    10 ページ

  9. Neo 5.5 sCMOS

    6 ページ

  10. Mechelle 5000

    5 ページ

  11. Apogee Alta F9000

    5 ページ

  12. Apogee Aspen CG230

    5 ページ

  13. Apogee Aspen CG47

    5 ページ

  14. Apogee Aspen CG6

    5 ページ

  15. ApogeeAspen CG9000

    5 ページ

  16. iXon EMCCD

    13 ページ

  17. iVac OEM 2

    2 ページ

  18. iKon-M X-Ray 2

    2 ページ

  19. Newton EMCCD

    16 ページ

  20. iQ Software

    12 ページ

  21. sCMOS

    16 ページ

  22. Low Light Imaging

    8 ページ

  23. Revolution

    23 ページ

  24. Clara Flyer

    2 ページ

  25. Revolution DSD

    11 ページ

  26. Neo and Zyla sCMOS

    27 ページ

  27. iXon

    33 ページ

カタログアーカイブ

  1. Astronomy Brochure

    19 ページ