Sodium Monitor Model 8036
1 / 8ページ

{{requestButtons}}

カタログの抜粋

Sodium Monitor Model 8036 - 1

Wide measuring range (0.01 gkg > 1 to 10mgkg > ֖1 ) covers all power station applications Full sample temperature compensation ֖maximises measurement accuracy 'Pumpless' liquid handling section low maintenance Automatic range changing ֖switches the current output in steps covering twodecades Simple calibration sequence automatically controlled by the microprocessor >

カタログの1ページ目を開く
Sodium Monitor Model 8036 - 2

Sodium Monitor Model 8036 > SS/8036_5 The prevention of damage to the boilers on modern powerstations becomes more and more critical as the cost of repairsand out of service plant continually rises. To reduce the effect ofboiler tube corrosion and the formation of scale on the inside of the tubes, the levels of impurities in the various parts of thesteam/water cycle must be carefully monitored.Sodium ions are normally the most abundant species in solutionfound in boiler plant, therefore the measurement of sodiumprovides a valuable indication of the overall purity of the solution.Whilst...

カタログの2ページ目を開く
Sodium Monitor Model 8036 - 3

Heat exchanger this is used to bring the standard solution close to the temperatureof the sample during calibration. Thermal differences therefore are minimised and calibration times reduced. Pressure relief valve ֖ during a calibration sequence the sample flow to the constanthead unit is stopped. The valve releases the pressure to prevent pressurising the sample line. Solenoid valve > under the control of the microprocessor the valve is used to switchbetween sample in normal running mode and standard solution during a calibration sequence. Constant head unit ֖ > 1 this removes the effect...

カタログの3ページ目を開く
Sodium Monitor Model 8036 - 5

Sodium concentration gkg > 1 or mgkg > ֖1 Sample mode indicates sample sodium concentration. During calibration indicates either of the two calibration sequences (CA1 or CA2). Calibration concentration displays the values of CA1 (the primary calibration solution) or CA2(the secondary calibration solution) ֖ also indicates when a calibration is in progress and shows if a single or two-point calibration has been initiated.An instantaneous display of concentration during calibration canalso be obtained. Alarm settings displays high or low alarm setting > Calibration fail displays CF following...

カタログの5ページ目を開く
Sodium Monitor Model 8036 - 6

Ranges Connections to sensor unit 0.01 gkg > 1 to 1mgkg > ֖1 or 0.1 gkg > 1 to 10mgkg > ֖1 internallyselectable Sample inlet:6mm O.D. compression fittingSample drain:10mm flexible, atmospheric drain Sample line material:Stainless steel Electrical:Via gland cable size 7 to 10.5mmMax. core size:Mains 32/0.2mmSignal 24/0.2mm Accuracy Dimensions of transmitter unit 10% of concentration or 0.02 gkg > ֖1 whichever is the greater 300mm wide x 300mm high x 200mm deep Reproducibility Mounting for transmitter unit 5% of concentration at constant temperature 230mm horizontal230mm vertical Response...

カタログの6ページ目を開く

ABB Measurement & Analyticsのすべてのカタログと技術パンフレット

  1. ProKiln GAC400

    24 ページ

  2. AV1 and AV2

    24 ページ

  3. OriMaster FPD500

    24 ページ

  4. AquaProbe FEA200

    16 ページ

  5. FTPA2000-HP260

    4 ページ

  6. 266HSH and 266NSH

    40 ページ

  7. SCK

    20 ページ

  8. INDI-BOY DSP

    8 ページ

  9. 9QGPS4500

    8 ページ

  10. 9QGPS3100P

    12 ページ

  11. 9QGPS2960

    12 ページ

  12. 9QGPS2625

    8 ページ

  13. MB3600-HP10

    5 ページ

  14. MB3600-CH10

    8 ページ

  15. ACF5000

    16 ページ

  16. Magnos28

    10 ページ

  17. Dynamic QR Code

    4 ページ

  18. EL3060

    8 ページ

  19. HP30

    4 ページ

  20. Aztec ATS430

    12 ページ

  21. LMS200

    12 ページ

  22. PGC5000

    8 ページ

  23. LGR-ICOS 927S

    4 ページ

  24. ALSCAN Brochure

    6 ページ

  25. MR Series

    2 ページ

  26. ABB product catalog

    32 ページ

  27. process photometer

    6 ページ