Residual Chlorine Monitor AW400
1 / 8ページ

カタログの抜粋

Residual Chlorine Monitor AW400 - 1

Data Sheet SS/AW400 Issue 3 Residual Chlorine Monitor AW400 Total and Free Residual Chlorine Measurement – add potassium iodide to reagent to convert free chlorine measurement to total chlorine Reagentless Operation – reduced operating costs with reagentless operation at a pH of 7.5 or less Multiple Sensor Input Capability – reduced installation costs for multiple point monitoring operations Self Cleaning Sensor Assembly – prolongs sensor life and maintains accuracy Fast Response Time – instant response to changes of residual chlorine, unlike batch/DPD methods Pressure Regulated Sample Flow – no additional plumbing required to ensure optimum sample delivery rate and pressure For the measurement of Free and Total Residual Chlorine, Chlorine Dioxide or Ozone in water

カタログの1ページ目を開く
Residual Chlorine Monitor AW400 - 2

Residual Chlorine Monitor AW400 SS/AW400 Issue 3 2 Introduction Residual chlorine is one of the most important measurements for the effective monitoring of cooling, potable and industrial water treatment processes. ABB's new AW400 Residual Chlorine Monitoring System incorporates the features our customers need to monitor and control residual chlorine levels accurately and affordably. The design enables users to either connect up to three residual chlorine monitoring sensors to the system or to combine pH, ORP and residual chlorine sensors; thus reducing the cost of each measurement...

カタログの2ページ目を開く
Residual Chlorine Monitor AW400 - 3

Residual Chlorine Monitor AW400 SS/AW400 Issue 3 3 Specification Measurement Range Chlorine, Chlorine Dioxide, Ozone 0 to 20 ppm (mg/l) Measurement Mode Measurement technology Amperometric cell Measurement interval Continuous Measurement Performance Accuracy Unbuffered: 2% or ±0.01 10ppb, whichever is the greater at pH <7.5 Buffered: 2% or ±10ppb, whichever is the greater at Sensitivity 0.001 ppm Minimum Detection Limit 0.005 ppm Drift Zero drift: ±1% over 30 days Max. Span drift: -5% over 30 days Max. Response Time 90% in less than 90 seconds Interferences Samples containing high...

カタログの3ページ目を開く
Residual Chlorine Monitor AW400 - 4

Residual Chlorine Monitor AW400 SS/AW400 Issue 3 4 Dimensions Mounting Dimensions – AW400 Cell Mounting Dimensions – Reagent Feed Pump

カタログの4ページ目を開く
Residual Chlorine Monitor AW400 - 5

Residual Chlorine Monitor AW400 SS/AW400 Issue 3 5 Mounting Dimensions – Transmitter

カタログの5ページ目を開く
Residual Chlorine Monitor AW400 - 6

Residual Chlorine Monitor AW400 SS/AW400 Issue 3 6 Ordering Information Accessories Residual Chorine Reagent Feeder Assemblies* Reagent Feeder Asembly (Free Chlorine 115V AC) AW401 011 Reagent Feeder Asembly (Free Chlorine 230V AC) AW401 012 Reagent Feeder Asembly (Total Chlorine 115V AC) AW401 013 Reagent Feeder Asembly (Total Chlorine 230V AC) AW401 014 *Comes complete with tubing and reagent bottles Residual Chlorine Monitor AW4 XX/ X X X X Transmitter Type Transmitter Transmitter with PID Control (Channel 1 only) 01 02 Sensor Type Channel 1 Chlorine Cell 1 Sensor Type Channel 2 No...

カタログの6ページ目を開く
Residual Chlorine Monitor AW400 - 7

Residual Chlorine Monitor AW400 SS/AW400 Issue 3 7 Benefits Summary Total and Free Residual Chlorine Measurement – add potassium iodide to reagent to convert free chlorine measurement to total chlorine Reagentless Operation – reduced operating costs with reagentless operation at a pH of 7.5 or less Multiple Sensor Input Capability – reduced installation costs for multiple point monitoring operations Self Cleaning Sensor Assembly – prolongs sensor life and maintains accuracy Fast Response Time – instant response to changes of residual chlorine, unlike batch/DPD methods Pressure Regulated...

カタログの7ページ目を開く
Residual Chlorine Monitor AW400 - 8

SS/AW400 Issue 3 Residual Chlorine Monitor AW400 ABB has Sales & Customer Support expertise in over 100 countries worldwide www.abb.com The Company’s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice. Printed in UK (07.09) © ABB 2009 ABB Limited 8 Hawksworth Southmead Industrial Park Didcot, Oxfordshire OX11 7HR UK Tel: +44 (0)1235 512000 Fax: +44 (0)1235 512020

カタログの8ページ目を開く

ABB Measurement & Analyticsのすべてのカタログと技術パンフレット

  1. ProKiln GAC400

    24 ページ

  2. AV1 and AV2

    24 ページ

  3. OriMaster FPD500

    24 ページ

  4. AquaProbe FEA200

    16 ページ

  5. FTPA2000-HP260

    4 ページ

  6. 266HSH and 266NSH

    40 ページ

  7. SCK

    20 ページ

  8. INDI-BOY DSP

    8 ページ

  9. 9QGPS4500

    8 ページ

  10. 9QGPS3100P

    12 ページ

  11. 9QGPS2960

    12 ページ

  12. 9QGPS2625

    8 ページ

  13. MB3600-HP10

    5 ページ

  14. MB3600-CH10

    8 ページ

  15. ACF5000

    16 ページ

  16. Magnos28

    10 ページ

  17. Dynamic QR Code

    4 ページ

  18. EL3060

    8 ページ

  19. HP30

    4 ページ

  20. Aztec ATS430

    12 ページ

  21. LMS200

    12 ページ

  22. PGC5000

    8 ページ

  23. LGR-ICOS 927S

    4 ページ

  24. ALSCAN Brochure

    6 ページ

  25. MR Series

    2 ページ

  26. ABB product catalog

    32 ページ

  27. process photometer

    6 ページ