PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters
1 / 44ページ

カタログの抜粋

PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 1

ABB MEASUREMENT & ANALY TICS | DATA SHEET PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters

カタログの1ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 2

— Measurement made easy Engineered solutions for all applications Standard overload resistance Base accuracy • From ± 0.25 % up to ± 0.075 % of calibrated span Span limits • 0,3 – 10000 kPa; 16 inH2O up to 1450 psi • 0,3 – 10000 kPa abs; 30 mmHg up to 1450 psia Proven sensor technology together with state-of-the-art digital technology • Valuable turn down ratio of up to 100:1 Stainless steel housing • Optimized for use in industrial process environments • Extremely robust New touch keypad technology • allows quick and easy local configuration without opening the cover, even in hazardous...

カタログの2ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 3

— General Description PxS100 Model PxS100 is the compact, reliable, essential pressure transmitter suitable for liquid, gas or steam applications. It is a straightforward and robust device designed for rapid delivery and easy installation and use. Thanks to its rugged design offering high resistance in even the most demanding environments, it is a competitive solution for most applications, including ones in hazardous areas. PxD100 Model PxD100 is the pressure transmitter from the Pxx100 series featuring the S26 diaphragm seal line accessories. While having a compact, sturdy stainless-steel...

カタログの3ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 4

— Specification - Physical(1) Materials Process isolating diaphragms (2) AISI 316 L ss, Hastelloy C276, Diaflex (antiabrasion), H-Shield (against hydrogen permeation) Process connection (2) AISI 316 L ss Hastelloy C276 (Available on request) Filling fluids (3) Silicone oil, inert fill (fluorocarbon), Mineral oil (FDA approved), Vegetal oil (FDA approved) Mounting bracket AISI 304 ss Sensor housing, electronics housing and cover AISI 316 L ss Filter for atmospheric ventilation • Filter material: polyamide (PA) • Filter housing: plastic (general purpose version), stainless steel (Ex certified...

カタログの4ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 5

— Functional – specification Range and span limits Sensor code Upper range limit (URL) Lower range limit (LRL) Minimum measuring span Model PGx100 Model PAx100 Gauge Absolute pressure pressure IMPORTANT (NOTE) The lower range limits (LRL) for models PAx100 are 0 absolute for all measuring ranges. Span limits Maximum span = upper range limit (URL) To optimize performance characteristics, it is recommended that you select the transmitter sensor code providing the lowest turn down ratio. TURNDOWN = Upper range limit / set span Zero suppression and elevation Zero and span can be adjusted to any...

カタログの5ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 6

— Specification – Operative limits Pressure limits Overpressure limits Without damage to the transmitter No damage will occur during transmitter operation, if they are used within the specifications and subject to the following limits: Model PxP100 Seal type Sensors The below table shows the maximum pressure limits of the connection when coupled with customer process interface. It also shows the compatibility of the sensor ranges. Pressure limit Universal G1" hygienic - AISI Universal G1" hygienic - HC Universal G1" hygienic Diaflex on AISI Seal model (Standard) Sensors L to R Diary 11851...

カタログの6ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 7

Temperature limits °C (°F) Ambient Seal model (Standard) Seal diaphragm size (thickness) / flange type PN 16 "Max. pressure Carbon steel flange @ 120 °C" "Max. pressure AISI 316 ss flange @ 20 °C" Ambient temperature limits Operating temperature range Process temperature limits Process temperature range Seal model (Standard) Seal diaphragm size (thickness) / flange type "Max. pressure AISI 316 L ss flange @ 20 °C" 1.6 Mpa, 16 bar, 230 psi Please refer to process temperature limits driven by S26 remote seals Process temperature limits Process temperature range EPDM gasket for universal...

カタログの7ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 8

— ...Specification – Operative limits Refer to the following FILL FLUID CHARACTERISTICS table detailing characteristics of fill fluids when used in transmitters. Operational limits may vary depending on the materials, gasket and configuration selected. Models Fill fluid (application) Process temperature and pressure limits Pmin mbar abs (mmHg) Inert oil Galden G5 (oxygen service) Silicone polymer Syltherm XLT (cryogenic service) Silicone oil for high temperature (for REMOTE SEAL) Silicone oil for high temperature (for DIRECT MOUNT SEAL) Vegetable oil Neobee M-20 (food - sanitary) FDA...

カタログの8ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 9

Model PGD100, PAD100 Seal system selection criteria The following table shows the types of seals available as direct mount and remote. According to the combination SEAL/TRANSMITTER SENSOR the table details the compatibility for one direct mount seal construction and the MAXIMUM CAPILLARY LENGTH when a seal is selected as remote. The mnemonics will be used as shortest cross references in the S26 data sheet. Seal model Seal type "Seal diaphragm size (thickness) [flange type]" Direct mount seal Remote seal Flanged flush diaphragm (ASME and EN standards) 1.5 in /DN 40 Wafer (ASME and EN...

カタログの9ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 10

— ...Specification – Operative limits Environmental limits Electromagnetic compatibility (EMC) The devices comply with the requirements and tests for EMC Directive 2014/30/EU to standards EN 61326-1 Table2 (industrial electromagnetic environment) concerning both emitted interference (CISPR11) and interference immunity. • Burst test: up to 2 kV at 5 kHz and 100 kHz • Surge test: up to 1 kV line-line and 2 kV line-ground Radiated emission: Group 1 - class B according to CISPR11 • FCC 47 CFR part 18.305 • ICES 005 - Issue 4 Conducted emission: group 1 - class A according to CISPR11 and to...

カタログの10ページ目を開く
PxS100, PxF100, PxP100, PxD100 Gauge and absolute pressure transmitters - 11

— Specification - Performance Reference conditions according to IEC 62828-1 • Ambient temperature constant in the range 15 – 25 °C (59 – 77 °F) • Relative humidity constant in the range 50 – 70 % • Ambient pressure constant in the range 860 – 1060 mbar • Measuring span based on zero position • Transmitter with AISI 316 L or Hastelloy C276 process isolating diaphragm • Filling fluid: Silicone oil • Supply voltage: 24 V DC ± 1 % • Load with HART: 250 Ω • Transmitter not grounded • Characteristic setting: linear, 4 – 20 mA. Unless otherwise specified, errors are quoted as % of calibrated span....

カタログの11ページ目を開く

ABB Measurement & Analyticsのすべてのカタログと技術パンフレット

  1. ProKiln GAC400

    24 ページ

  2. AV1 and AV2

    24 ページ

  3. OriMaster FPD500

    24 ページ

  4. AquaProbe FEA200

    16 ページ

  5. FTPA2000-HP260

    4 ページ

  6. 266HSH and 266NSH

    40 ページ

  7. SCK

    20 ページ

  8. INDI-BOY DSP

    8 ページ

  9. 9QGPS4500

    8 ページ

  10. 9QGPS3100P

    12 ページ

  11. 9QGPS2960

    12 ページ

  12. 9QGPS2625

    8 ページ

  13. MB3600-HP10

    5 ページ

  14. MB3600-CH10

    8 ページ

  15. ACF5000

    16 ページ

  16. Magnos28

    10 ページ

  17. Dynamic QR Code

    4 ページ

  18. EL3060

    8 ページ

  19. HP30

    4 ページ

  20. Aztec ATS430

    12 ページ

  21. LMS200

    12 ページ

  22. PGC5000

    8 ページ

  23. LGR-ICOS 927S

    4 ページ

  24. ALSCAN Brochure

    6 ページ

  25. MR Series

    2 ページ

  26. ABB product catalog

    32 ページ

  27. process photometer

    6 ページ