LMT200 external mount magnetostrictive level transmitter
1 / 20ページ

{{requestButtons}}

カタログの抜粋

LMT200 external mount magnetostrictive level transmitter - 1

ABB MEASUREMENT & ANALY TICS | DATA SHEET LMT200 external mount magnetostrictive level transmitter High accuracy non–intrusive liquid level and interface level detectio

カタログの1ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 2

— Measurement made easy K–TEK Level products — Features • Calibrated from the factory • High accuracy: .01 % of full scale or ±1.27 mm • No re–calibration needed: set it and forget it • Easy setup with waveform display • Not affected by agitation, foam or emulsion layers • No oscilloscope required • Designed to mount externally to K–TEK KM26 or other magnetic level gauge • Superior sensor (patent #5,473,245) • Local indication with HMI display • Dual compartment housing with separate field terminal compartment • Loop powered to 15.24 m (50 ft) probe length • Total and/or interface level...

カタログの2ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 3

— Principle of operation: The LMT200 is based upon the magnetostrictive principle. 1 1 The device electronics generates a low energy current pulse at fixed intervals. 2 The electrical pulses create a magnetic field which 2 travels down a specialized wire inside the sensor tube. 3 3 The interaction of the magnetic field around the wire and the magnetic float causes a torsional stress wave to be induced in the wire. This torsion propagates along the wire at a known velocity, from the position of the magnetic float and toward both ends of the wire. 4 4 A patented sensing element placed in the...

カタログの3ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 4

LMT200 EXTERNAL MOUNT MAGNETOSTRICTIVE LEVEL TRANSMITTER | DS/LMT200-EN REV.D Electronic transmitter Repeatability ±0.005 % Of full scale or 0.315 mm (0.012 in), whichever is greater Non-linearity ±0.01 % Of full scale or 0.86 mm (0.034 in), whichever is greater Measuring accuracy ±0.01 % Of full scale or 1.27 mm (0.050 in), whichever is greater1 Supply voltage 12 to 43 V DC for 4 to 20mA HART loop powered, 9.0 to 32 V DC for Foundation™ Fieldbus Output/Communications 4 to 20 mA HART7® or FOUNDATION Fieldbus ITK6.3.0 User interface Interactive display, DTM,...

カタログの4ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 5

— Float design for ABB K–TEK products – KM26S Every KM26 MLG float is precisely engineered to customer applica­ ion, ensuring optimal accuracy and performance. t Precisely spaced magnets create a 360° magnetic field coverage, safeguarding level transmitter and gauge performance, even the most challenging applications. Several materials of construction available including titanium, Monel®, Hastelloy® C, stainless steel, and plastics. Tefzel®, Halar®, TEFLON® S protective coatings are also available. Process pressures to 344 bar (5000 PSI) to full vacuum. A minimum of 75 grams of buoyancy are...

カタログの5ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 6

— Electrical connections Devices with HART communication Current output/HART output Terminal Power supply, current output/HART output PWR/COMM – EXT.METER Internal earthing terminal Power supply, current output/ 7 HART output 9 2 8 3 3 Load resistance 7 4 4 Power supply/supply isolator 5 5 PLC/DCS 6 HART handheld terminal 6 7 External indicator 7 8 External earthing terminal 8 9 Terminal for external indicator 9 0 Laptop 10 For connecting the signal voltage/supply voltage, twisted cables with a conductor cross–section of 0.8 to 0.35 mm2 (18 to 22 AWG )and a maximum length of 1500 m (4921...

カタログの6ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 7

— Power supply Devices with HART communication Terminals PWR/COMM +/PWR/COMM – Supply voltage Residual ripple Power consumption Current Output/HART output Only for devices with HART communication Terminals: PWR/COMM +/PWR/COMM – In HART communication, the smallest load is RB = 250 Ω. The load is RB is calculated as a function of the available supply voltage US and the selected, signal current IB as follows: US – min operating voltage (12.0) V DC RB Load resistance US Supply voltage IB Singal current Internal ground termination point Fieldbus lines Devices with Foundation Fieldbus...

カタログの7ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 8

— ...Probe selection guide Probe selection chart - standard 85 (185) Directions: 1. determine the maximum ambient tempearture and the maximum process temperature of the installation. 2. Plot the intersection between the two temperatures on the chart. 3. Choose the appropiate probe by matching the zone number with the probe type. *Ambient temperature is the air temperature in the work area surrounding the device. It takes into account the influence of heat created by equipment surrounding the area personnel will be working. It is not the skin temperature of the chamber or insulation blanket....

カタログの8ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 9

LMT200 EXTERNAL MOUNT MAGNETOSTRICTIVE LEVEL TRANSMITTER | DS/LMT200-EN REV.D Approvals General purpose INMETRO, ATEX/IECEx flameproof, intrinsically safe, no-sparking (protection type marked by customer) NEPSI (China), intrinsically safe NEPSI (China), flameproof housing NEPSI (China), non-sparking NEPSI (China), ATEX/IECEx flameproof, intrinsically safe, non-sparking (protection type marked by customer) ATEX/IECEx intrinsic safety ATEX/IECEx flameproof housing ATEX/IECEx flameproof non-sparking (Ex nA) EAC, intrinsically safe EAC, flameproof housing EAC, non-sparking Multi-approval -...

カタログの9ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 10

LMT200 EXTERNAL MOUNT MAGNETOSTRICTIVE LEVEL TRANSMITTER | DS/LMT200-EN REV.D Mounting orientation Bottom left connected electronic housing Bottom right connected electronic housing Top left connected electronic housing Top right connected electronic housing Housing Aluminum with 2 x M20 x 1.5 Aluminum with 2 x NPT V2 in 316L stainless steel with 2 x M20 x 1.5 316L stainless steel with 2 x NPT V in Remote/aluminum/2 x M20 x 1.5* Remote/aluminum/2 x NPT V in* Remote/stainless steel/2 x M20 x 1.5* Remote/stainless steel/2 x NPT V in* Special Through the glass (TTG) push buttons, display and...

カタログの10ページ目を開く
LMT200 external mount magnetostrictive level transmitter - 11

Sensor probe options 90 degree bend housing extension (maximum probe length 7.62 m/25 ft) Add nitrogen purged vapor seal to standard probe Sensor special Device identification plate Add stainless steel hang tag, custom markings 4 lines, 22 characters per line Other tagging special Signal cable length (for remote transmitter only) 5 m (approx. 16 ft)* Surge protector Surge/transient protector Special other Special paint or treatment on housing Nuclear use, device to be used in a nuclear facility (application must be reviewed by ABB) Mounted accessories Mounted to chamber with vibration...

カタログの11ページ目を開く

ABB Measurement & Analyticsのすべてのカタログと技術パンフレット

  1. ProKiln GAC400

    24 ページ

  2. AV1 and AV2

    24 ページ

  3. OriMaster FPD500

    24 ページ

  4. AquaProbe FEA200

    16 ページ

  5. FTPA2000-HP260

    4 ページ

  6. 266HSH and 266NSH

    40 ページ

  7. SCK

    20 ページ

  8. INDI-BOY DSP

    8 ページ

  9. 9QGPS4500

    8 ページ

  10. 9QGPS3100P

    12 ページ

  11. 9QGPS2960

    12 ページ

  12. 9QGPS2625

    8 ページ

  13. MB3600-HP10

    5 ページ

  14. MB3600-CH10

    8 ページ

  15. ACF5000

    16 ページ

  16. Magnos28

    10 ページ

  17. Dynamic QR Code

    4 ページ

  18. EL3060

    8 ページ

  19. HP30

    4 ページ

  20. Aztec ATS430

    12 ページ

  21. LMS200

    12 ページ

  22. PGC5000

    8 ページ

  23. LGR-ICOS 927S

    4 ページ

  24. ALSCAN Brochure

    6 ページ

  25. MR Series

    2 ページ

  26. ABB product catalog

    32 ページ

  27. process photometer

    6 ページ