Te c h n i c a l D a t a N i t r a Ly t ®P l u s
Appropriate Electrodes Reference electrode VARiON®Plus Ref, Measuring electrode VARiON®Plus NO3, Compensation electrode VARiON®Plus Cl,
VARiON®Plus NO3-HS
Measuring Ranges/ Resolution NO3-N: 1 ... 1000 mg/l / 1 mg/l; 0.1 ... 100.0 mg/l / 0.1 mg/l
NO3
-: 5 ... 4500 mg/l / 5 mg/l; 0.5 ... 450.0 mg/l / 0.5 mg/l
Cl-: 1 ...1000 mg/l / 1 mg/l
Temp. Measurement
and Compensation
Integrated NTC thermistor
Range: 32 … 104 °F (0 °C … +40 °C), Accuracy ± 0.5 K, resolution 0.1 K, t95 < 20 s
Calibration Procedures Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution
Ambient Conditions Operating temperature: 32 °F … 104 °F (0 °C ... + 40 °C),
storing temperature: 32 °F … 104 °F (0 °C ... + 40 °C)
pH Range pH 4 ... pH 11
Measuring Accuracy
in laboratory standard solutions
± 5 % of measured value ± 0.2 mg/l in standard solutions
Working Life (typically) All electrodes: 18 months (in typical application - sewage plants)
Mechanical Sensor body: V4A Stainless Steel 1.4571 Protective cap: POM,
Temp. sensor: V4A Stainless Steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes)
Electrode connector: POM
Max. Pressure Maximum 0.2 bar (incl. SACIQ sensor connection cable; with installed electrodes)
Power Consumption 0.2 Watt
Dimensions 15.43 x 1.57 in. (392 x 40 mm, length x diameter), incl. sensor connection cable SACIQ
Weight Approx. 1.48 lb (670 g, without electrode, without sensor connection cable)
Guaranty NitraLyt®Plus 700 IQ: 2 years for defects of quality
O r d e r i n g I n f o r m a t i o n
NitraLyt®Plus-System Order No.
NitraLyt®Plus 700 IQ Robust digital armature for ionselective electrodes (VARiON®Plus Ref/VARiON®Plus NO3/VARiON®Plus Cl/
VARiON®Plus NO3-HS; not included in the delivery scope)
107 080
VARiON®Plus Ref Reference electrode for VARiON®Plus 700 IQ/AmmoLyt®Plus 700 IQ/NitraLyt®Plus 700 IQ 107 042
VARiON®Plus NO3 Nitrate electrode for measuring nitrate with dynamic CI compensation 107 045
VARiON®Plus NO3-HS Nitrate electrode for measuring nitrate with automatic CI compensation 107 049
VARiON®Plus Cl 107 047
CH Cleaning head 900 107
MIQ/CHV PLUS Valve module for automatic cleaning by compressed air controlled directly via the IQ Sensor Net bus 480 018
DIQ/CHV Valve module for automatic compressed air cleaning for System 182; accessible by means of an DIQ/S 182 relay 472 007
IP 68 cETLus 2 Year Standard solutions see brochure “Product Details”
Warranty
*
* on armatures
I Q - L a b L i n k
With the initial installation of NitraLyt®Plus the dependency
of the used electrode reference values for nitrate and chloride
is determined by a photometric system and can be
adapted with NitraLyt®Plus.
The measuring data is required for guaranteeing a precise
matrix adaption.
For enhancing the data transfer between the laboratory
spectrophotometers photoLab® 6100 / 6600 and
NitraLyt®Plus, a USB memory stick in combination with the
IQ-LabLink function can be used for the MIQ/TC 2020 XT
to automatically read the data and store it on NitraLyt®Plus.
• Safe, comfortable and fast data transfer
• Automatic plausibility check of data
50
|