SONIC DIGITAL MULTI- ultrasonic generators
SONIC DIGITAL MULTI MF - 22604 Error messages and troubleshootingHF - lights up but no powerPossible causes: No transducer connected Terminal box incorrectly connected www.weber-ultrasonics.com Cable bridge from PIN 2 to 3 in HF plug forgotten Transducer failure Remedy: Connect transducer Check terminal box contacts Insert a cable bridge into HF plug Fit a new transducer Generator turns on and off of its own accordPossible causes: Remote control has no contact Timer function is activated (Aux 2 is lit) Remedy: Check remote control Deactivate timer control Generator cannot be controlledPossible causes: Frequency window incorrectly adjusted External output setting activated/deactivated Remedy: Reset frequency window to Standard Activate or deactivate external output setting Generator outputs HF but a malfunction is indicatedPossible causes: Defective connection X1 PIN 7-13 Error signals not set correctly at factory Control lead failure Remedy: Check connection X1 PIN 7-13 Set error trap to open in I/O polarities menu Check control lead and replace if necessary Generator outputs HF, cleaning result unsatisfactoryPossible causes: Power set too low Transducer defective Transducer fitted incorrectly Remedy: Increase power Replace transducer Check fitting location of the transducer 19 |
|